Validation of MR thermometry: method for temperature probe sensor registration accuracy in head and neck phantoms.

Matthew R Tarasek1, Ruben Pellicer, Lorne W Hofstetter, Wouter C M Numan, Jurriaan F Bakker, Gyula Kotek, Paolo Togni, René F Verhaart, Eric W Fiveland, Gavin C Houston, Gerard C van Rhoon, Margarethus Marius Paulides, Desmond Teck Beng Yeo.   

Abstract

PURPOSE: Magnetic resonance thermometry (MRT) is an attractive means to non-invasively monitor in vivo temperature during head and neck hyperthermia treatments because it can provide multi-dimensional temperature information with high spatial resolution over large regions of interest. However, validation of MRT measurements in a head and neck clinical set-up is crucial to ensure the temperature maps are accurate. Here we demonstrate a unique approach for temperature probe sensor localisation in head and neck hyperthermia test phantoms.
METHODS: We characterise the proton resonance frequency shift temperature coefficient and validate MRT measurements in an oil-gel phantom by applying a combination of MR imaging and 3D spline fitting for accurate probe localisation. We also investigate how uncertainties in both the probe localisation and the proton resonance frequency shift (PRFS) thermal coefficient affect the registration of fibre-optic reference temperature probe and MRT readings.
RESULTS: The method provides a two-fold advantage of sensor localisation and PRFS thermal coefficient calibration. We provide experimental data for two distinct head and neck phantoms showing the significance of this method as it mitigates temperature probe localisation errors and thereby increases accuracy of MRT validation results.
CONCLUSIONS: The techniques presented here may be used to simplify calibration experiments that use an interstitial heating device, or any heating method that provides rapid and spatially localised heat distributions. Overall, the experimental verification of the data registration and PRFS thermal coefficient calibration technique provides a useful benchmarking method to maximise MRT accuracy in any similar context.

Entities:  

Mesh:

Year:  2014        PMID: 24571177     DOI: 10.3109/02656736.2014.887794

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  8 in total

1.  Average SAR prediction, validation, and evaluation for a compact MR scanner head-sized RF coil.

Authors:  M R Tarasek; Y Shu; D Kang; S Tao; E Gray; J Huston; Y Hua; D T B Yeo; M A Bernstein; T K Foo
Journal:  Magn Reson Imaging       Date:  2021-10-16       Impact factor: 2.546

2.  Optimization of Single Voxel MR Spectroscopy Sequence Parameters and Data Analysis Methods for Thermometry in Deep Hyperthermia Treatments.

Authors:  J Hartmann; J Gellermann; T Brandt; M Schmidt; S Pyatykh; J Hesser; O Ott; R Fietkau; C Bert
Journal:  Technol Cancer Res Treat       Date:  2016-07-14

3.  Open Source 3D Multipurpose Measurement System with Submillimetre Fidelity and First Application in Magnetic Resonance.

Authors:  Haopeng Han; Raphael Moritz; Eva Oberacker; Helmar Waiczies; Thoralf Niendorf; Lukas Winter
Journal:  Sci Rep       Date:  2017-10-18       Impact factor: 4.379

4.  Validation of hybrid angular spectrum acoustic and thermal modelling in phantoms.

Authors:  Sara L Johnson; Douglas A Christensen; Christopher R Dillon; Allison Payne
Journal:  Int J Hyperthermia       Date:  2018-10-15       Impact factor: 3.914

5.  Quantitative, Multi-institutional Evaluation of MR Thermometry Accuracy for Deep-Pelvic MR-Hyperthermia Systems Operating in Multi-vendor MR-systems Using a New Anthropomorphic Phantom.

Authors:  Sergio Curto; Bassim Aklan; Tim Mulder; Oliver Mils; Manfred Schmidt; Ulf Lamprecht; Michael Peller; Ruediger Wessalowski; Lars H Lindner; Rainer Fietkau; Daniel Zips; Gennaro G Bellizzi; Netteke van Holthe; Martine Franckena; Margarethus M Paulides; Gerard C van Rhoon
Journal:  Cancers (Basel)       Date:  2019-11-02       Impact factor: 6.639

6.  Predicting ablation zones with multislice volumetric 2-D magnetic resonance thermal imaging.

Authors:  Zahabiya Campwala; Benjamin Szewczyk; Teresa Maietta; Rachel Trowbridge; Matthew Tarasek; Chitresh Bhushan; Eric Fiveland; Goutam Ghoshal; Tamas Heffter; Katie Gandomi; Paulo Alberto Carvalho; Christopher Nycz; Erin Jeannotte; Michael Staudt; Julia Nalwalk; Abigail Hellman; Zhanyue Zhao; E Clif Burdette; Gregory Fischer; Desmond Yeo; Julie G Pilitsis
Journal:  Int J Hyperthermia       Date:  2021       Impact factor: 3.753

7.  Modeling of Interstitial Ultrasound Ablation for Continuous Applicator Rotation With MR Validation.

Authors:  Katie Y Gandomi; Paulo A W G Carvalho; Matthew Tarasek; Eric W Fiveland; Chitresh Bhushan; Emery Williams; Paul Neubauer; Zhanyue Zhao; Julie Pilitsis; Desmond Yeo; Christopher J Nycz; Everette Burdette; Gregory S Fischer
Journal:  IEEE Trans Biomed Eng       Date:  2021-05-21       Impact factor: 4.756

8.  Heat Modulation of Intrinsic MR Contrasts for Tumor Characterization.

Authors:  Matthew Tarasek; Oguz Akin; Jeannette Roberts; Thomas Foo; Desmond Yeo
Journal:  Cancers (Basel)       Date:  2022-01-14       Impact factor: 6.639

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.