Alex P Lange1, Reza Sadjadi2, Fiona Costello3, Ivo Guber4, Anthony L Traboulsee2.
Abstract
Objective. The aim of this study was to test the reproducibility of the Heidelberg Spectralis SD-OCT and to determine if provided software retest function for follow-up exam is superior to manual centration. Design. Prospective, cross-sectional study. Participants. 20 healthy subjects. Methods. All subjects underwent SD-OCT testing to determine retinal nerve fiber layer (RNFL) measurements sequentially on two different days and with two different centration techniques. Within-subject standard deviation, coefficient of variation, and intraclass correlation coefficient were used to assess reproducibility. Results. RNFL measurements showed high reproducibility, low within-subject standard deviation (1.3), low coefficient of variation (0.63%), and low intra-class correlation coefficient (0.98 (95% CI 0.97-0.99)) in the automated centration and manual centration groups for average RNFL Thickness. Quadrants showed slightly higher variability in the manual group compared to the automated group (within-subject standard deviation 2.5-5.3 versus 1.1-2.4, resp.). Conclusions. SD-OCT provides high-resolution RNFL measurements with high reproducibility and low variability. The re-test function allows for easier recentration for longitudinal examinations with similar results in average RNFL, but less variability in quadrant RNFL. SD-OCT high reproducibility and low variability is a promising fact and should be further evaluated in longitudinal studies of RNFL.Entities:
Year: 2012 PMID: 24558593 PMCID: PMC3914277 DOI: 10.5402/2012/860819
Source DB: PubMed Journal: ISRN Ophthalmol ISSN: 2090-5688
Figure 1OCT algorithm for every included subject.
Descriptive data of study subjects.
| Number of subjects | 20 |
| Males | 11 |
| Females | 9 |
| Number of eyes included | 39 |
| Mean age (years, ± SD) | 31.51 (8.04) |
| Number of emmetropic eyes | 24 |
| Number of myopic eyes | 15 |
| Mean RNFL in | 98.46 (97.57–99.35) |
SD-OCT RNFL thickness reproducibility.
| Manual centration | Automatic centration | |||||||
|---|---|---|---|---|---|---|---|---|
| Mean RNFL in | Sw | CV % | ICC (95% CI) | Mean RNFL in | Sw | CV % | ICC (95% CI) | |
| Average | 98.35 (97.07–99.62) | 1.6 | 0.65 | 0.999 (0.998–0.999) | 98.57 (97.33–99.81) | 1.3 | 0.63 | 0.983 (0.974–0.990) |
| Temporal | 76.06 (74.59–77.53) | 2.8 | 0.97 | 0.980 (0.970–0.988) | 74.83 (73.54–76.11) | 1.1 | 0.86 | 0.998 (0.998–0.999) |
| Superior | 118.35 (116.36–120.31) | 5.3 | 0.83 | 0.997 (0.995–0.998) | 118.84 (116.95–120.73) | 2.1 | 0.80 | 0.999 (0.983–0.999) |
| Nasal | 71.80 (69.75–73.85) | 3.2 | 1.43 | 0.989 (0.984–0.994) | 72.86 (70.98–74.75) | 2.4 | 1.29 | 0.998 (0.998–0.999) |
| Inferior | 126.87 (124.83–128.91) | 2.5 | 0.80 | 0.997 (0.995–0.998) | 127.31 (125.35–129.27) | 1.9 | 0.77 | 0.999 (0.999–0.999) |
SD-OCT: spectral-domain optical coherence tomography; RNFL: retinal nerve fiber layer; CI: confidence intervals; Sw: within subject standard deviation; CV: coefficient of variation; ICC: intraclass correlation.
Overview published studies comparing RNFLT in SD-OCT.
| Author | SD-OCT used | Study population (eyes) | Results |
|---|---|---|---|
| Leung et al. [ | Cirrus and stratus | 83 glaucoma | Cirrus: (95% CI) |
|
| |||
| Vizzeri et al. [ | Cirrus | 78 glaucoma | Cirrus (95% CI) |
|
| |||
| Kim et al. [ | Cirrus and stratus | 27 controls | Cirrus (95% CI) |
|
| |||
| Gonzalez-Garcia et al. [ | RTVue-100 | 76 glaucoma | RTVue (95% CI) |
|
| |||
| Seibold et al. [ | Cirrus, Spectralis, RTVue-100, stratus | 80 controls | Cirrus (95% CI) |
|
| |||
| Arthur et al. [ | Spectralis | 30 controls | Spectralis (95% CI) |
|
| |||
| Langenegger et al. [ | Spectralis | 56 controls | Spectralis (95% CI) |
|
| |||
| Serbecic et al. [ | Spectralis | 62 Controls | Spectralis (95% CI) |
|
| |||
| Current study | Spectralis | 39 controls | Spectralis (95% CI) |