Asystole detected by implantable loop recorders: true or false?

Hussam Ali1, Antonio Sorgente, Elisabetta Daleffe, Riccardo Cappato.   

Abstract

We report a case of a false asystole detected by an implantable loop recorder a few days after its implantation. In the discussion section we try to give some hints to help cardiac electrophysiologists in distinguish true from false asystoles, in order to avoid unuseful and potentially dangerous implantations of pacemakers.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  asystole; implantable loop recorder; pacemaker

Mesh:

Year:  2014        PMID: 24533648      PMCID: PMC6932227          DOI: 10.1111/anec.12135

Source DB:  PubMed          Journal:  Ann Noninvasive Electrocardiol        ISSN: 1082-720X            Impact factor:   1.468


  4 in total

1.  Electromagnetic interference of an implantable loop recorder by commonly encountered electronic devices.

Authors:  C C de Cock; H J Spruijt; L M van Campen; A W Plu; C A Visser
Journal:  Pacing Clin Electrophysiol       Date:  2000-10       Impact factor: 1.976

2.  Implantable loop recorder undersensing mimicking complete heart block.

Authors:  S I Chrysostomakis; E N Simantirakis; M E Marketou; P E Vardas
Journal:  Europace       Date:  2002-04       Impact factor: 5.214

3.  Electromagnetic interference with pacemakers caused by portable media players.

Authors:  Jay P Thaker; Mehul B Patel; Krit Jongnarangsin; Valdis V Liepa; Ranjan K Thakur
Journal:  Heart Rhythm       Date:  2008-01-29       Impact factor: 6.343

4.  Inappropriate asystole detection in early postoperative phase after loop recorder implantation.

Authors:  Miriam Bortnik; Eraldo Occhetta; Andrea Magnani; Anna Degiovanni; Paolo Marino
Journal:  ISRN Cardiol       Date:  2011-04-17
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.