Thermoluminescent dosimetry of the Symmetra 125I model I25.S06 interstitial brachytherapy seed.

N S Patel1, S T Chiu-Tsao, J F Williamson, P Fan, T Duckworth, D Shasha, L B Harrison.   

Abstract

As the efficacy of brachytherapy prostate treatment is becoming realized, new models of 125I seeds are being introduced. In this article we present thermoluminescent dosimetry (TLD) in a solid water phantom for a new design of 125I seed (UroMed/Bebig Symmetra, Model I25.S06). TLD cubes, LiF TLD-100, from Bicron (Solon, OH) with dimension 1 x 1 x 1 mm3 were irradiated at various distances from the seed at angles ranging from 0 degrees to 90 degrees in 10 degrees increments. The TLD detectors were calibrated by irradiation in a 60Co teletherapy beam. Monte Carlo simulation was used to account for TLD energy dependence and the deviation of solid water composition (as determined by chemical analysis of a sample) from liquid water. Dose rates per unit air kerma strength were determined based on calibrations traceable to the 1999 NIST standard (corrected for NIST measurement errors made in 1999) for the Symmetra seed. Dose data is presented in TG-43 format as a function of distance and angle. Values for lambda, F(r, theta), g(r), and the anisotropy constant are obtained for use in radiation treatment planning (RTP) software. The dose rate constant was determined to be 1.033+/-6.4% cGy h(-1) U(-1), which is comparable to model 6702 and higher than model 6711. We find the relative dose distributions of the Symmetra seed are similar to model 6702, and less anisotropic than model 6711. After accounting for deviation of measured solid water composition from the manufacturer's specification, good agreement between TLD results and Monte-Carlo-aided values was found.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11548947     DOI: 10.1118/1.1388218

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO.

Authors:  Larry A DeWerd; Geoffrey S Ibbott; Ali S Meigooni; Michael G Mitch; Mark J Rivard; Kurt E Stump; Bruce R Thomadsen; Jack L M Venselaar
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

2.  Dosimetric characterization and output verification for conical brachytherapy surface applicators. Part I. Electronic brachytherapy source.

Authors:  Regina K Fulkerson; John A Micka; Larry A DeWerd
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

3.  Determination of dosimetric parameters for shielded 153Gd source in prostate cancer brachytherapy.

Authors:  Mahdi Ghorbani; Benyamin Khajetash; Najmeh Ghatei; Mohammad Mehrpouyan; Ali S Meigooni; Ramin Shahraini
Journal:  Radiol Oncol       Date:  2017-02-22       Impact factor: 2.991

4.  Measured transverse-axis dosimetric parameters of the model STM1251 125I interstitial source.

Authors:  Z Li; J F Williamson
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.