Dose reduction using a dynamic, piecewise-linear attenuator.

Scott S Hsieh1, Dominik Fleischmann2, Norbert J Pelc3.   

Abstract

PURPOSE: The authors recently proposed a dynamic, prepatient x-ray attenuator capable of producing a piecewise-linear attenuation profile customized to each patient and viewing angle. This attenuator was intended to reduce scatter-to-primary ratio (SPR), dynamic range, and dose by redistributing flux. In this work the authors tested the ability of the attenuator to reduce dose and SPR in simulations.
METHODS: The authors selected four clinical applications, including routine full field-of-view scans of the thorax and abdomen, and targeted reconstruction tasks for an abdominal aortic aneurysm and the pancreas. Raw data were estimated by forward projection of the image volume datasets. The dynamic attenuator was controlled to reduce dose while maintaining peak variance by solving a convex optimization problem, assuminga priori knowledge of the patient anatomy. In targeted reconstruction tasks, the noise in specific regions was given increased weighting. A system with a standard attenuator (or "bowtie filter") was used as a reference, and used either convex optimized tube current modulation (TCM) or a standard TCM heuristic. The noise of the scan was determined analytically while the dose was estimated using Monte Carlo simulations. Scatter was also estimated using Monte Carlo simulations. The sensitivity of the dynamic attenuator to patient centering was also examined by shifting the abdomen in 2 cm intervals.
RESULTS: Compared to a reference system with optimized TCM, use of the dynamic attenuator reduced dose by about 30% in routine scans and 50% in targeted scans. Compared to the TCM heuristics which are typically used withouta priori knowledge, the dose reduction is about 50% for routine scans. The dynamic attenuator gives the ability to redistribute noise and variance and produces more uniform noise profiles than systems with a conventional bowtie filter. The SPR was also modestly reduced by 10% in the thorax and 24% in the abdomen. Imaging with the dynamic attenuator was relatively insensitive to patient centering, showing a 17% increase in peak variance for a 6 cm shift of the abdomen, instead of an 82% increase in peak variance for a fixed bowtie filter.
CONCLUSIONS: A dynamic prepatient x-ray attenuator consisting of multiple wedges is capable of achieving substantial dose reductions and modest SPR reductions.

Entities:  

Mesh:

Year:  2014        PMID: 24506631      PMCID: PMC3987627          DOI: 10.1118/1.4862079

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies.

Authors:  M Gies; W A Kalender; H Wolf; C Suess
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

2.  An inverse-geometry volumetric CT system with a large-area scanned source: a feasibility study.

Authors:  Taly Gilat Schmidt; Rebecca Fahrig; Norbert J Pelc; Edward G Solomon
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

3.  A prototype table-top inverse-geometry volumetric CT system.

Authors:  Taly Gilat Schmidt; Josh Star-Lack; N Robert Bennett; Samuel R Mazin; Edward G Solomon; Rebecca Fahrig; Norbert J Pelc
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

4.  A three-dimensional-weighted cone beam filtered backprojection (CB-FBP) algorithm for image reconstruction in volumetric CT-helical scanning.

Authors:  Xiangyang Tang; Jiang Hsieh; Roy A Nilsen; Sandeep Dutta; Dmitry Samsonov; Akira Hagiwara
Journal:  Phys Med Biol       Date:  2006-01-25       Impact factor: 3.609

5.  Inverse-geometry volumetric CT system with multiple detector arrays for wide field-of-view imaging.

Authors:  Samuel R Mazin; Josh Star-Lack; N Robert Bennett; Norbert J Pelc
Journal:  Med Phys       Date:  2007-06       Impact factor: 4.071

6.  Simulated dose reduction in conventional chest CT: validation study.

Authors:  J R Mayo; K P Whittall; A N Leung; T E Hartman; C S Park; S L Primack; G K Chambers; M K Limkeman; T L Toth; S H Fox
Journal:  Radiology       Date:  1997-02       Impact factor: 11.105

7.  The feasibility of a piecewise-linear dynamic bowtie filter.

Authors:  Scott S Hsieh; Norbert J Pelc
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

8.  Automatic patient centering for MDCT: effect on radiation dose.

Authors:  Jianhai Li; Unni K Udayasankar; Thomas L Toth; John Seamans; William C Small; Mannudeep K Kalra
Journal:  AJR Am J Roentgenol       Date:  2007-02       Impact factor: 3.959

9.  A method for correcting bone induced artifacts in computed tomography scanners.

Authors:  P M Joseph; R D Spital
Journal:  J Comput Assist Tomogr       Date:  1978-01       Impact factor: 1.826

10.  Noise due to photon counting statistics in computed X-ray tomography.

Authors:  D A Chesler; S J Riederer; N J Pelc
Journal:  J Comput Assist Tomogr       Date:  1977-01       Impact factor: 1.826

View more
  6 in total

1.  Dynamic fluence field modulation for miscentered patients in computed tomography.

Authors:  Andrew Mao; Grace J Gang; William Shyr; Reuven Levinson; Jeffrey H Siewerdsen; Satomi Kawamoto; J Webster Stayman
Journal:  J Med Imaging (Bellingham)       Date:  2018-10-24

2.  Dynamic fluence field modulation in computed tomography using multiple aperture devices.

Authors:  Grace J Gang; Andrew Mao; Wenying Wang; Jeffrey H Siewerdsen; Aswin Mathews; Satomi Kawamoto; Reuven Levinson; J Webster Stayman
Journal:  Phys Med Biol       Date:  2019-05-21       Impact factor: 3.609

3.  Control algorithms for dynamic attenuators.

Authors:  Scott S Hsieh; Norbert J Pelc
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

4.  Reconstruction of three-dimensional tomographic patient models for radiation dose modulation in CT from two scout views using deep learning.

Authors:  Juan C Montoya; Chengzhu Zhang; Yinsheng Li; Ke Li; Guang-Hong Chen
Journal:  Med Phys       Date:  2022-01-06       Impact factor: 4.506

5.  The piecewise-linear dynamic attenuator reduces the impact of count rate loss with photon-counting detectors.

Authors:  Scott S Hsieh; Norbert J Pelc
Journal:  Phys Med Biol       Date:  2014-05-13       Impact factor: 3.609

6.  A prototype piecewise-linear dynamic attenuator.

Authors:  Scott S Hsieh; Mark V Peng; Christopher A May; Picha Shunhavanich; Dominik Fleischmann; Norbert J Pelc
Journal:  Phys Med Biol       Date:  2016-06-10       Impact factor: 3.609

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.