| Literature DB >> 24505781 |
Junichi Tokuda1, Sang-Eun Song2, Kemal Tuncali2, Clare Tempany2, Nobuhiko Hata2.
Abstract
We propose a novel automatic fiducial frame detection and registration method for device-to-image registration in MRI-guided prostate interventions. The proposed method does not require any manual selection of markers, and can be applied to a variety of fiducial frames, which consist of multiple cylindrical MR-visible markers placed in different orientations. The key idea is that automatic extraction of linear features using a line filter is more robust than that of bright spots by thresholding; by applying a line set registration algorithm to the detected markers, the frame can be registered to the MRI. The method was capable of registering the fiducial frame to the MRI with an accuracy of 1.00 +/- 0.73 mm and 1.41 +/- 1.06 degrees in a phantom study, and was sufficiently robust to detect the fiducial frame in 98% of images acquired in clinical cases despite the existence of anatomical structures in the field of view.Entities:
Mesh:
Year: 2013 PMID: 24505781 PMCID: PMC4009697 DOI: 10.1007/978-3-642-40760-4_45
Source DB: PubMed Journal: Med Image Comput Comput Assist Interv