| Literature DB >> 18429016 |
André de Oliveira1, Jaane Rauschenberg, Dirk Beyersdorff, Wolfhard Semmler, Michael Bock.
Abstract
MR-guided transrectal prostate biopsy is currently a time-consuming procedure because the imaging slice is often manually realigned with the biopsy needle during lesion targeting. In this work a pulse sequence is presented that automatically follows a passive marker attached to a dedicated MR biopsy device holder, thus providing an alternative to existing active tracking methods. In two orthogonal tracking FLASH images of the marker the position of the needle axis is automatically identified using a phase-only cross-correlation (POCC) algorithm. The position information is then used to realign a trueFISP imaging slice in real time. In phantom experiments the sensitivity of this technique to initial misalignments of the marker and to the signal-to-noise ratio was evaluated. In several puncture experiments the precision of the needle placement was analyzed. The POCC algorithm allowed for a precise identification of the marker in the images even under severe initial misalignments of up to 45 degrees. At a frame rate 1 image/s a precision of the needle placement of 1.5 +/- 1.1 mm could be achieved. (c) 2008 Wiley-Liss, Inc.Mesh:
Year: 2008 PMID: 18429016 DOI: 10.1002/mrm.21430
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668