Literature DB >> 24501713

The deep inspiration breath hold technique using Abches reduces cardiac dose in patients undergoing left-sided breast irradiation.

Ha Yoon Lee1, Jee Suk Chang1, Ik Jae Lee1, Kwangwoo Park1, Yong Bae Kim1, Chang Ok Suh1, Jun Won Kim1, Ki Chang Keum1.   

Abstract

PURPOSE: We explored whether the deep inspiration breath hold (DIBH) technique using Abches during left-sided breast irradiation was effective for minimizing the amount of radiation to the heart and lung compared to free breathing (FB).
MATERIALS AND METHODS: Between February and July 2012, a total of 25 patients with left-sided breast cancer underwent two computed tomography scans each with the DIBH using Abches and using FB after breast-conserving surgery. The scans were retrospectively replanned using standardized criteria for the purpose of this study. The DIBH plans for each patient were compared with FB plans using dosimetric parameters.
RESULTS: All patients were successfully treated with the DIBH technique using Abches. Significant differences were found between the DIBH and FB plans for mean heart dose (2.52 vs. 4.53 Gy), heart V30 (16.48 vs. 45.13 cm(3)), V20 (21.35 vs. 54.55 cm(3)), mean left anterior descending coronary artery (LAD) dose (16.01 vs. 26.26 Gy, all p < 0.001), and maximal dose to 0.2 cm(3) of the LAD (41.65 vs. 47.27 Gy, p = 0.017). The mean left lung dose (7.53 vs. 8.03 Gy, p = 0.073) and lung V20 (14.63% vs. 15.72%, p = 0.060) of DIBH using Abches were not different significantly compared with FB.
CONCLUSION: We report that the use of a DIBH technique using Abches in breathing adapted radiotherapy for left-sided breast cancer is easily feasible in daily practice and significantly reduces the radiation doses to the heart and LAD, therefore potentially reducing cardiac risk.

Entities:  

Keywords:  Abches; Cardiac toxicity; Deep inspiration breath hold

Year:  2013        PMID: 24501713      PMCID: PMC3912239          DOI: 10.3857/roj.2013.31.4.239

Source DB:  PubMed          Journal:  Radiat Oncol J        ISSN: 2234-1900


Introduction

Postoperative radiotherapy for breast cancer patients has been shown to significantly reduce the risk of local recurrence as well as improving long-term survival [1,2]. However, many studies have demonstrated increased cardiac mortality and morbidity after breast radiotherapy [3-5]. Although many of the studies included in the this review involved old treatment techniques, which probably delivered a higher dose to the heart than seen in modern radiotherapy clinics, the issue of cardiac morbidity and mortality after breast cancer treatment is still relevant [6,7]. Therefore, dose-volume reduction of irradiation to the heart has become a common treatment goal in order to avoid increased risk of cardiovascular morbidity and mortality. A promising method for minimizing the volume of the heart included in the treatment field involves increasing the distance between the target and heart through deep inspiration. In the deep inspiration breath hold (DIBH) technique, the goal is to achieve the maximum separation between the target area and heart, with irradiation therefore taking place only at or near maximum inspiration. This allows high doses of radiation to be delivered to the chest wall and to the breast tissue while reducing the high dose area of the heart. The DIBH technique has been recommended to reduce radiation to the heart by displacing the breast and chest wall away from the heart during treatment [8-11]. Different methods, such as the use of real-time position management systems [12], active breathing control [13], and magnetic sensors [14] have been used to monitor or control respiratory motion. However, these methods have not been routinely implemented by many breast cancer treatment centers. The failure to adopt DIBH into routine practice is due to concern about increases in the required resources and workloads. We explored whether the DIBH technique using Abches during left-sided breast irradiation was effective for minimizing the amount of radiation to the heart and lung compared to free breathing (FB). This is the first study to evaluate the feasibility of DIBH using the commercially available Abches system and to incorporate this method into routine clinical practice. We also performed dosimetric evaluations to assess the extent of dose reduction to the heart, left anterior descending coronary artery (LAD), and ipsilateral lung.

Materials and Methods

1. Patient selection

Beginning in February 2012, we implemented the DIBH technique using Abches for selected patients undergoing adjuvant radiotherapy after breast-conserving surgery for left-sided invasive or in situ breast carcinoma at our institution. Patient selection criteria for DIBH irradiation were early stage (≤T2 and ≤N1a stage) breast cancer for which whole breast irradiation (without regional nodal irradiation) was indicated. Between February and July 2012, the feasibility of 30-second DIBH was evaluated among 26 selected patients by a dedicated technician. Only 1 patient was excluded from use of the DIBH technique due to frequent cough. The remaining 25 patients underwent computed tomography (CT) simulation scans (SOMATOM sensation; Siemens, Erlangen, Germany) with both FB and DIBH without intravenous contrast. These scans were retrospectively replanned using standardized criteria for the purpose of this study. The median age of these patients was 29 years (range, 23 to 76 years).

2. Abches system

Abches is a respiration-monitoring device that was developed by Onishi et al. at the University of Yamanashi, Japan [15] (Fig. 1). The main unit of the Abches system is composed primarily of plastic, with the use of metal kept to an absolute minimum. Fulcrum A and B are placed on the abdomen and breast of the patient, respectively. During respiration, pointer C of the device moves along with the fulcrum. This device therefore serves as an external device to monitor respiration. The Abches system is not used for gating irradiation but for facilitating precise irradiation of a target by assuring a stable breath hold.
Fig. 1

Abches system.

3. Respiratory control using the Abches system

All patients were advised that the treatment was to be administered during respiratory standstill and were informed about the importance of reproducing the tumor position at the outpatient clinic by the treating physician. Next, each patient was trained how to hold her breath in the supine position with the left arm above the head in a customized arm support using a breast board to ensure that the sternum was in a horizontal position. Fig. 2 shows the procedure implemented for respiratory control using the Abches system. For all patients, the positions of fulcrum A and B were correctly reproduced every time using anatomical landmarks on the patient as indices, the contralateral breast and just below the xiphoid process, respectively. Patients were instructed to hold their breath at the end of inspiration and asked to practice breath holding by repeating the instructions "breathe in, out, and stop" with a mirror in front of the forehead (Fig. 2A). During each respiration session, the position of pointer C shifted with the fulcrum and stopped at some other fixed point. The red and blue marks indicated the ends of expiration and inspiration during regulatory respiration for each patient. The yellow mark indicated the deep inspiration breath hold, which is different for each patient. It was important that each patient maintained her breath hold well at the yellow mark (Fig. 2B-D). The monitor was continuously observed during actual CT scanning and irradiation in the operating room. If the position pointer C varied considerably or if the patient could hold her breath for only a few seconds even after repeated training, the use of the Abches system was discontinued for that patient.
Fig. 2

Procedure for respiratory control using the Abches system. (A) Patient position using Abches. (B) The red mark indicates the end of expiration during regulatory respiration. (C) The blue mark indicates the end of inspiration during regulatory respiration. (D) The yellow mark indicates the deep inspiration breath hold.

4. Delineation of target volumes and organs at risk

For consistency, the same physician performed all of the delineation procedures according to the Radiation Therapy Oncology Group guidelines using MIMvista ver. 5.4 (MIM Software, Cleveland, OH, USA), and the same CT threshold values, window levels and widths were used for all scans of all patients. The breast clinical target volume (CTV) included visible glandular breast tissue and the surgical cavity visualized on the CT scan, excluding 5 mm from the skin; contouring was assisted by a lead wire placed along the palpated breast tissue prior to the CT scan. Medially and laterally, the breast planning target volume (PTV) was defined by the angle of the tangential beams and extended from the midline to the midaxilla. Superiorly and inferiorly, the breast PTV was extended to include visible glandular breast tissue with a margin of 1.5 cm. If the boundary of the breast tissue could not be determined, the superior border was placed at the level of the suprasternal notch. The breast PTV was not adjusted to minimize the volume of the heart included within the tangential fields. The heart contour included the ventricles and atria, including the root of the aorta, the pulmonary artery and veins, and venae cavae. Since cardiac vessels run in the fatty tissue within the pericardium, they were included in the contours. The LAD was delineated from the root of the aorta to the apex of the heart with a standardized 5-mm diameter to account for uncertainties associated with cardiac motion and respiration. Intravenous contrast was not used for the CT scans. However, in most patients, the LAD could be visualized directly, and its position could also be identified according to its location overlying the interventricular septum [16]. The ipsilateral lung was delineated using the CT lung window. MIMvista ver. 5.4 (MIM Software) is an intensity-based free-form deformable registration algorithm with essentially limitless degrees of freedom designed to expedite the delineation of target volumes and organs at risk. Rigid registration allows linear or uniform transformation of all voxels in the image set within six degrees of freedom (translations in three orthogonal directions and yaw, pitch, and roll rotations). It is of limited utility in the DIBH setting due to significant changes in anatomy. On the other hand, deformable registration algorithms compute the deformation maps of vector fields that can connect the voxels in FB CT images to DIBH CT images, a process that accounts for nonlinear and nonuniform relationships between the image sets. Thus, MIMvista is a useful tool for target volume delineation in breast cancer patients undergoing radiotherapy using the DIBH technique. Deformable registrations of images at our institution were performed with MIMvista. This algorithm was evaluated for clinical use by Piper [17].

5. Treatment planning and dosimetric evaluation

For each patient, both FB and DIBH CT scans were planned by a single dosimetrist to ensure plan uniformity. For all CT scans, computerized treatment planning was performed using Pinnacle ver. 9.4 (Philips, Andover, MA, USA). Two opposing 6- or 10-MV tangential conformal fields with multileaf collimators were used. The prescription doses of all plans were 50.4 Gy in 1.8 Gy per fraction for the whole breast. Comparisons of the following dosimetric values for FB and DIBH plans were performed using MIMvista: mean heart dose, the volume of the heart receiving 20 Gy and 30 Gy (heart V20 and V30), mean LAD dose, the maximal dose to 0.2 cm3 of the LAD, the mean left lung dose and the volume of the left lung receiving 20 Gy (left lung V20). Because the LAD is a serial structure, the maximal dose to 0.2 cm3 of the LAD, which is approximately 1 cm in length using the standardized 5-mm diameter and may be more biologically relevant than the point maximal dose, was evaluated. Paired t-tests were performed using SPSS ver. 20.0 (SPSS Inc., Chicago, IL, USA) to compare dosimetric parameters between treatment techniques. A two-tailed p < 0.05 was considered statistically significant.

Results

1. Patient compliance

After excluding only 1 patient from the DIBH technique because of frequent cough, the remaining 25 patients were successfully treated with the DIBH technique using Abches. The procedure was easily comprehended by patients. As noted earlier, because the patients were able to watch their own breathing by looking directly into a mirror, it was simple to educate the patients and the patients learned quickly. The mean monitor unit (MU) for two opposing 6- or 10-MV tangential plan was about 213 MU/min (range, 196 to 265 MU/min) and the beam-on time for a dose rate of 400 MU/min was about 15 seconds. Thus, the breath hold was well accepted with a duration of about 15 seconds. All 25 patients were able to produce at least three consecutive breath holds without problems.

2. Dosimetric comparison between FB and DIBH plans

As shown in Table 1, significant differences were found between the DIBH and FB plans for the mean heart dose (2.52 vs. 4.53 Gy), heart V30 (16.48 vs. 45.13 cm3), V20 (21.35 vs. 54.55 cm3), mean LAD dose (16.01 vs. 26.26 Gy, all p < 0.001), and maximal dose to 0.2 cm3 of the LAD (41.65 vs. 47.27 Gy, p = 0.017). In Fig. 3, beams-eye views are shown for the medial tangential fields of left-sided breast cancer patients for both FB and DIBH. Cardiac volume within the radiation field was minimized as much as possible, primarily by conforming the posterior borders of the tangential fields to the CTV using multileaf collimators. The use of DIBH allowed the left-sided breast and heart to be separated during radiation treatment, thus excluding substantial heart volumes from the high dose area. This is demonstrated by the calculated dose volume histogram (Fig. 4). There were significantly lower doses delivered to the heart and LAD with the DIBH plan compared to the FB plan. Although lung inflation during DIBH increased the absolute lung volume within the field on beams-eye view (Fig. 3), the mean left lung dose (7.53 vs. 8.03 Gy, p = 0.073) and lung V20 (14.63% vs. 15.72%, p = 0.060) were reduced with DIBH compared to the FB plan (Table 1), although the difference was not significant.
Table 1

Dosimetric comparison between FB and DIBH plans for all 25 patients

Values are presented as mean (range).

FB, free breathing; DIBH, deep inspiration breath hold; LAD, left anterior descending coronary artery.

a)Paired t-test.

Fig. 3

Beams-eye views of the medial tangential fields for free breathing (A) and deep inspiration breath hold (B) plans. Heart outline in pink, left anterior descending coronary artery in red, and surgical cavity in sky blue.

Fig. 4

Dose volume histograms comparing free breathing (A) and deep inspiration breath hold (B) plans for the same patient. Whole breast line in green, left anterior descending coronary artery in red, heart in pink, and ipsilateral lung in blue.

3. Workload requirements and patient and staff acceptance

The implementation of the DIBH technique required minimal additional resources and only a moderate increase in workload. We assessed the overall treatment time for each patient. The mean durations of FB and DIBH were about 5 minutes 10 seconds and 10 minutes 30 seconds, respectively. All planning staff became familiar with the DIBH planning and Abches education processes. Staff responsible for operating breast cancer treatment machines became proficient with patient setup and treatment delivery using Abches in about 6 weeks.

Discussion and Conclusion

Delivering treatment to the left breast using the DIBH technique has been shown to reduce both the irradiated heart volume and dose to the heart. Early studies indicated that DIBH could reduce the volume of the heart within tangential fields. Remouchamps et al. [8] reported a mean absolute reduction in heart V30 of 3.6% in a preliminary study including five patients. Korreman et al. [18] reported a reduction in the median heart volume receiving more than 50% of the prescription dose from 19.2% to 1.9% and a reduction in the median volume of the LAD receiving the same dose from 88.9% to 3.6% in a sample of nine patients. In a subsequent publication, the same group used the relative seriality model to calculate the expected reduction in cardiac mortality from the use of DIBH in 16 cases. The cardiac mortality probabilities for FB and DIBH were 4.8% and 0.1% [12]. More recent reports support these findings. At Princess Margaret Hospital, Toronto, Canada, 20 patients were planned for both FB and DIBH treatments. Statistically significant reductions in the volume of the heart receiving more than 50% of the prescription dose, mean heart dose, mean LAD dose, and the maximum dose to 0.2 cm3 of the LAD were noted after applying the DIBH technique [11]. In a similar study of 19 unselected patients from the Netherlands Cancer Institute the maximum and mean doses and the V30 for the heart, the left ventricle and the LAD were all significantly reduced by DIBH [10]. Several reports have documented the tolerability and interfraction reproducibility of the DIBH technique [13,19,20]. Our data are consistent with those of other published reports demonstrating significant reductions in the doses to the heart and LAD with DIBH technique using Abches. However, several factors may limit comparisons between studies. The reported heart doses may be influenced by variability in the method used to contour the heart between studies. We employed non-contrast CT scans to outline the atria and ventricles. Previous studies excluded the epicardial fat and pericardium, the roots of the great vessels and/or the venae cavae, and were performed with or without the use of intravenous contrast. Contouring the coronary arteries is particularly difficult, as they are often poorly visualized due to non-contrast CT scans and often seem noncontiguous throughout the data set due to cardiac and respiratory motion. To account for these uncertainties, we used a standardized contouring protocol [16]. We attempted to maintain consistency of volume measurements by employing the same radiation oncologist to perform all contouring. Even so, some minor variations in all contours may have occurred due to variation in the anatomy of the heart and lung between FB and DIBH scans. Although the absolute left lung volume appeared to have increased with DIBH, proportionately less of the left lung was irradiated. While lung volume may increase with deep inspiration, the lung density may decrease, resulting in irradiation of a reduced fraction of normal lung mass [21]. In our study, the mean left lung dose (7.53 vs. 8.03 Gy, p = 0.073) and percentage of the left lung (14.63% vs. 15.7%, p = 0.060) receiving 20 Gy were both smaller with DIBH (borderline significance). Therefore, the risk of radiation-induced pneumonitis is unlikely to increase using the DIBH technique. Implementing the DIBH technique using Abches for all patients receiving left-sided breast radiotherapy was neither desirable nor feasible, because the benefits were questionable for most patients with minimal or no heart included within the radiation field. Therefore, it is important to establish criteria to select patients with unfavorable cardiac anatomy, who would benefit most from using Abches. Our results demonstrate that DIBH technique using Abches resulted in statistically significant reductions in the mean heart dose, heart V20, heart V30, mean LAD dose, and maximal dose to 0.2 cm3 of the LAD. These parameters may be of assistance in identifying patients with unfavorable cardiac anatomy who would most benefit from reductions in cardiac dose during radiotherapy to the left breast. Further research is needed to determine which patients are likely to derive the most benefit from this technique before DIBH CT scans are performed. It has been demonstrated in several studies that breathing adaptation techniques can be used to reduce the irradiated heart and lung volumes, primarily by utilizing lung inflation which dilutes the amount of lung tissue in the radiation fields and spatially separates the heart from target. An active breathing control device has been used to freeze the lung volume at 75% of maximum inspiration capacity, providing significant lung and heart dose reductions [22]. The device consists of a digital spirometer to record the patients' real-time breathing trace and a balloon valve that is triggered to automatically inflate when the patient inspire to their preset DIBH level. This allows for a predictable, consistent level of chest wall expansion with each DIBH. Another one of breathing adaptation techniques is the Real-time Position Management (RPM) system (Varian Medical Systems, Palo Alto, CA, USA). The device permits breathing-synchronized CT scanning and radiotherapy during certain time intervals on the linear accelerator [18]. The position and chest wall excursion are assessed individually for each breath-hold mode, and the patient is then instructed to repeat each mode to improve reproducibility. We think that Abches would probably be less precise than other breathing adaptation techniques in terms of consistent level of chest wall expansion and reproducibility of breath-hold level. However, because the patient can monitor her own breathing and breath holds with durations of about 15 seconds were well accepted, the breathing procedure was easily comprehensible to the patients and there was good compliance using Abches. In addition, the main unit of the Abches system is composed primarily of plastic and is therefore quite light. Thus, it is easy to perform patient setup in the radiation treatment room without affecting dose calculation using the radiation treatment planning system (Pinnacle ver. 9.4, Philips). All planning staff easily became familiar with the DIBH planning and Abches education processes. The staff who operated breast cancer treatment machines became proficient with patient setup and treatment delivery using Abches after about 6 weeks. The additional workload per patient treated with DIBH using Abches was modest, and the use of MIMvista dramatically reduced the breast radiation treatment planning time, allowing rapid delineation and dose-volume assessments of organs at risk in real-time using a FB plan. This procedure can be implemented for routine use in busy clinics. In conclusions, we report that the application of the DIBH technique using Abches in breathing adapted radiotherapy for left-sided breast cancer is easily feasible in daily practice and significantly reduces the doses to the heart and LAD, therefore potentially reducing the cardiac risk. Based on the results of this study, at our institution we now treat left-sided breast cancer patients with Abches.
  21 in total

1.  Clinical results of image-guided deep inspiration breath hold breast irradiation.

Authors:  Gerben R Borst; Jan-Jakob Sonke; Suzanne den Hollander; Anja Betgen; Peter Remeijer; Aline van Giersbergen; Nicola S Russell; Paula H M Elkhuizen; Harry Bartelink; Corine van Vliet-Vroegindeweij
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-06       Impact factor: 7.038

2.  The management of tumor motions in the stereotactic irradiation to lung cancer under the use of Abches to control active breathing.

Authors:  Tohru I Tarohda; Mitsuru Ishiguro; Kouhei Hasegawa; Yukihiko Kohda; Hiroaki Onishi; Tetsuya Aoki; Tsuyoshi Takanaka
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

3.  Radiation-induced heart morbidity after adjuvant radiotherapy of early breast cancer - Is it still an issue?

Authors:  Birgitte Offersen; Inger Højris; Marie Overgaard
Journal:  Radiother Oncol       Date:  2011-08       Impact factor: 6.280

4.  Short-term displacement and reproducibility of the breast and nodal targets under active breathing control.

Authors:  Jean M Moran; James M Balter; Merav A Ben-David; Robin B Marsh; Marcel Van Herk; Lori J Pierce
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-06-01       Impact factor: 7.038

5.  Deep inspiration breath hold to reduce irradiated heart volume in breast cancer patients.

Authors:  K E Sixel; M C Aznar; Y C Ung
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-01-01       Impact factor: 7.038

6.  Three-dimensional evaluation of intra- and interfraction immobilization of lung and chest wall using active breathing control: a reproducibility study with breast cancer patients.

Authors:  Vincent M Remouchamps; Nicola Letts; Di Yan; Frank A Vicini; Michel Moreau; Julie A Zielinski; Jian Liang; Larry L Kestin; Alvaro A Martinez; John W Wong
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-11-15       Impact factor: 7.038

7.  Significant reductions in heart and lung doses using deep inspiration breath hold with active breathing control and intensity-modulated radiation therapy for patients treated with locoregional breast irradiation.

Authors:  Vincent M Remouchamps; Frank A Vicini; Michael B Sharpe; Larry L Kestin; Alvaro A Martinez; John W Wong
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-02-01       Impact factor: 7.038

Review 8.  Radiation-induced cardiac toxicity after therapy for breast cancer: interaction between treatment era and follow-up duration.

Authors:  Senem Demirci; Jiho Nam; Jessica L Hubbs; Thu Nguyen; Lawrence B Marks
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-03-15       Impact factor: 7.038

9.  Improved heart, lung and target dose with deep inspiration breath hold in a large clinical series of breast cancer patients.

Authors:  Henrik D Nissen; Ane L Appelt
Journal:  Radiother Oncol       Date:  2012-11-28       Impact factor: 6.280

10.  Long-term risk of cardiovascular disease in 10-year survivors of breast cancer.

Authors:  Maartje J Hooning; Akke Botma; Berthe M P Aleman; Margreet H A Baaijens; Harry Bartelink; Jan G M Klijn; Carolyn W Taylor; Flora E van Leeuwen
Journal:  J Natl Cancer Inst       Date:  2007-03-07       Impact factor: 13.506

View more
  21 in total

1.  Accuracy of Breath-hold CT in Treatment Planning for Lung Stereotactic Ablative Radiotherapy.

Authors:  Dominique Mathieu; Charles Martel; Marie-Pierre Campeau; Édith Filion; David Roberge; Houda Bahig; Toni Vu; Louise Lambert; Karim Boudam; Jean-Francois Carrier
Journal:  Cureus       Date:  2014-12-22

2.  An offline technique to evaluate residual motion of the diaphragm during deep inspiratory breath-hold from cone-beam CT datasets.

Authors:  Manuel Blessing; Julian Hofmann; Lena Vogel; Judit Boda-Heggemann; Frank Lohr; Frederik Wenz; Florian Stieler; Anna Simeonova-Chergou
Journal:  Strahlenther Onkol       Date:  2018-05-22       Impact factor: 3.621

3.  Left-sided Breast Cancer Irradiation With Deep Inspiration Breath-hold: Changes in Heart and Lung Dose in Two Periods.

Authors:  Anne Caroline Knöchelmann; Nese Ceylan; Michael Bremer
Journal:  In Vivo       Date:  2022 Jan-Feb       Impact factor: 2.155

4.  Usefulness of a thermoplastic breast bra for breast cancer radiotherapy : A prospective analysis.

Authors:  Marc D Piroth; Dalma Petz; Michael Pinkawa; Richard Holy; Michael J Eble
Journal:  Strahlenther Onkol       Date:  2016-06-10       Impact factor: 3.621

5.  Cardiac dose reduction with breathing adapted radiotherapy using self respiration monitoring system for left-sided breast cancer.

Authors:  KiHoon Sung; Kyu Chan Lee; Seung Heon Lee; So Hyun Ahn; Seok Ho Lee; Jinho Choi
Journal:  Radiat Oncol J       Date:  2014-06-30

6.  Evaluation of the Flash effect in breast irradiation using TomoDirect: an investigational study.

Authors:  Dae Gyu Kang; Sung Ill Park; Sung Hwan Kim; Mi Joo Chung; Kwang-Man Lee; Jong Hoon Lee
Journal:  J Radiat Res       Date:  2015-02-10       Impact factor: 2.724

Review 7.  Deep Inspiration Breath Hold: Techniques and Advantages for Cardiac Sparing During Breast Cancer Irradiation.

Authors:  Carmen Bergom; Adam Currey; Nina Desai; An Tai; Jonathan B Strauss
Journal:  Front Oncol       Date:  2018-04-04       Impact factor: 6.244

8.  Radiation Pneumonitis in Association with Internal Mammary Node Irradiation in Breast Cancer Patients: An Ancillary Result from the KROG 08-06 Study.

Authors:  Jinhyun Choi; Yong Bae Kim; Kyung Hwan Shin; Sung-Ja Ahn; Hyung-Sik Lee; Won Park; Su Ssan Kim; Jin Hee Kim; Kyu Chan Lee; Dong Won Kim; Hyun Suk Suh; Kyung Ran Park; Hyun Soo Shin; Chang-Ok Suh
Journal:  J Breast Cancer       Date:  2016-09-23       Impact factor: 3.588

9.  Is the Deep Inspiration Breath-Hold Technique Superior to the Free Breathing Technique in Cardiac and Lung Sparing while Treating both Left-Sided Post-Mastectomy Chest Wall and Supraclavicular Regions?

Authors:  Anupama Darapu; Rajesh Balakrishnan; Patricia Sebastian; Mohamathu Rafic Kather Hussain; Paul Ravindran; Subhashini John
Journal:  Case Rep Oncol       Date:  2017-01-16

Review 10.  The cardiac dose-sparing benefits of deep inspiration breath-hold in left breast irradiation: a systematic review.

Authors:  Lloyd M Smyth; Kellie A Knight; Yolanda K Aarons; Jason Wasiak
Journal:  J Med Radiat Sci       Date:  2015-01-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.