| Literature DB >> 24489560 |
Manal Tadros1, Astrid Petrich2.
Abstract
Matrix-assisted laser desorption/ionization time of flight mass spectrometry (MALDI-TOF MS) can be used to identify bacteria directly from positive blood and sterile fluid cultures. The authors evaluated a commercially available kit - the Sepsityper Kit (Bruker Daltonik, Germany) - and MALDI-TOF MS for the rapid identification of organisms from 80 flagged positive blood culture broths, of which 73 (91.2%) were blood culture specimens and seven (8.7%) were cerebrospinal fluid specimens, in comparison with conventional identification methods. Correct identification to the genus and species levels was obtained in 75 of 80 (93.8%) and 39 of 50 (78%) blood culture broths, respectively. Applying the blood culture analysis module, a newly developed software tool, improved the species identification of Gram-negative organisms from 94.7% to 100% and of Gram-positive organisms from 66.7% to 70%. MALDI-TOF MS is a promising tool for the direct identification of organisms cultured from sterile sites.Keywords: Blood culture analysis module; Blood cultures; Direct identification; MALDI-TOF MS
Year: 2013 PMID: 24489560 PMCID: PMC3905001 DOI: 10.1155/2013/701093
Source DB: PubMed Journal: Can J Infect Dis Med Microbiol ISSN: 1712-9532 Impact factor: 2.471