| Literature DB >> 24465663 |
Ji-Guo Yu1, Yi-Fan Feng2, Yi Xiang1, Jin-Hai Huang3, Giacomo Savini4, Vincenzo Parisi4, Wan-Ju Yang1, Xun-An Fu1.
Abstract
BACKGROUND: Parkinson disease (PD) is a neurodegenerative process that leads to a selective loss of dopaminergic neurons, mainly in the basal ganglia of the brain. Numerous studies have analyzed the ability of optical coherence tomography (OCT) to detect retinal nerve fiber layer (RNFL) thickness abnormalities and changes in PD, but the results have not always been consistent. Therefore, we carried out a meta-analysis to evaluate the RNFL thickness measured with OCT in PD. METHODS ANDEntities:
Mesh:
Year: 2014 PMID: 24465663 PMCID: PMC3897496 DOI: 10.1371/journal.pone.0085718
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1Search and study selection process.
Characteristics and quality of studies included in the Meta-analysis.
| OCT | No. of eyes | Mean Age (years) | Gender (Male/Female) | NOS Scale | |||||||||
| Author(s) Year | Country | Manufacturer | Model | PD | HC | PD | HC | PD | HC | Selection | Comparability | Expose | Total Score |
| Morgia/2013 | Italy | Carl Zeiss | Stratus | 81 | 86 | 65.6 | 65.5 | 24/19 | 34/52 | *** | * | ** | 6 |
| Aaker/2010 | USA | Heidelberg | Spectralis | 18 | 19 | 64 | 67 | 13/5 | 9/10 | **** | * | ** | 7 |
| Moschos/2011 | Greece | Carl Zeiss | Stratus Model 3000 | 32 | 40 | 57 | 52 | 9/7 | 10/10 | *** | * | ** | 6 |
| Albrecht/2012 | Germany | Heidelberg | Spectralis | 80 | 70 | 61.2 | NR | 30/10 | NR | *** | * | ** | 6 |
| Tsironi/2012 | Greece | Carl Zeiss | Stratus | 24 | 24 | 66.6 | 64.3 | 14/10 | 11/13 | *** | ** | ** | 7 |
| Rohani/2012 | Iran | Topcon | 3D-OCT 1000 Mark II | 54 | 50 | 54.6 | 55.0 | 40/14 | 34/16 | **** | * | ** | 7 |
| Kirbas/2013 | Turkey | Carl Zeiss | Cirrus HD | 84 | 80 | 59.3 | 57.0 | 24/28 | 24/16 | *** | * | ** | 6 |
| Inzelberg/2004 | Israel | NR | NR | 10 | 10 | 57 | 52 | 5/5 | NR | *** | * | ** | 6 |
| Archibald/2011 | UK | Carl Zeiss | Stratus Model 3000 | 34 | 17 | 71.6 | 71.3 | 22/15 | 9/10 | **** | ** | ** | 8 |
| Satue/2013 | Spain | Carl Zeiss Heidelberg | Cirrus HD Spectralis | 100 | 100 | 64 | 64 | 82/18 | NR | ** | * | ** | 5 |
| Altintas/2008 | Turkey | Carl Zeiss | Stratus Model 3000 | 34 | 22 | 59.3 | 58.1 | 9/8 | 6/5 | ** | * | ** | 5 |
| Garcia-Martin/2012 | Spain | Carl Zeiss Heidelberg | Cirrus HD Spectralis | 75 | 75 | 64.4 | 64.2 | 45/30 | 45/30 | *** | * | ** | 6 |
| Sen/2013 | Turkey | Optovue | RTVue-100 | 18 | 11 | 63.56 | 61.18 | 8/10 | 5/6 | **** | ** | ** | 8 |
PD = Parkinson's disease; HC = healthy controls; RNFL = retinal nerve fiber layer; NOS = Newcastle-Ottawa Scale; NR = not reported.
Figure 2Meta-analysis of average RNFL thickness between PD patients and healthy controls.
Figure 3Meta-analysis of RNFL thickness between PD patients and healthy controls: (A) superior, (B) inferior, (C) nasal and (D) temporal.
Subgroup analysis according to types of Optical Coherence Tomography.
| No. of studies | WMD (95% CI) of RNFL thickness |
| |
| Stratus OCT | 5 | ||
| average | −5.53 [−7.82, −3.24] | <0.00001 | |
| superior | −3.94 [−7.53, −0.34] | 0.03 | |
| inferior | −6.51 [−9.99, −3.02] | 0.0003 | |
| nasal | −4.38 [−7.76, −1.00] | 0.01 | |
| temporal | −4.99 [−7.57, −2.42] | 0.0001 | |
| Spectralis OCT | 4 | ||
| average | −3.05 [−4.62, −1.47] | 0.0002 | |
| superior | −5.31 [−7.87, −2.75] | <0.0001 | |
| inferior | −5.54 [−11.07, −0.02] | 0.05 | |
| nasal | −3.19 [−8.64, 2.27] | 0.25 | |
| temporal | −4.16 [−6.23, −2.10] | <0.0001 | |
| Cirrus HD OCT | 3 | ||
| average | −8.37 [−11.75, −5.00] | <0.00001 | |
| superior | −5.15 [−9.52, −0.79] | 0.02 | |
| inferior | −6.29 [−11.92, −0.67] | 0.03 | |
| nasal | −0.91 [−1.85, 0.02] | 0.06 | |
| temporal | −6.21 [−7.56, −4.87] | <0.00001 | |
| 3D-OCT | 1 | ||
| average | −11.39 [−16.06, −6.72] | <0.00001 | |
| superior | −15.28 [−23.12, −7.44] | 0.0001 | |
| inferior | −14.95 [−22.64, −7.26] | 0.0001 | |
| nasal | −8.30 [−14.84, −1.76] | 0.01 | |
| temporal | −7.05 [−11.72, −2.38] | 0.003 | |
| RTVue-100 | 1 | ||
| average | −8.84 [−16.10, −1.58] | 0.02 | |
| superior | −3.35 [−11.05, 4.35] | 0.39 | |
| inferior | −13.72 [−22.44, −5.00] | 0.002 |
Figure 4A funnel plot for evaluating the publication bias: (A) average, (B) superior, (C) inferior, (D) nasal and (E) temporal.