OBJECTIVE: To review functional results and quality of life of the first patients implanted with a newly introduced bone conduction implant system. STUDY
DESIGN: Retrospective chart analysis of 6 patients (6 ears) implanted for conductive hearing loss (CHL) and mixed hearing loss (MHL) in 1 tertiary referral center between July 2012 and February 2013.
METHODS: Implantation of a new bone conduction hearing device. Pure tone audiometry (air conduction and bone conduction thresholds, pure tone average, air-bone gap, and functional gain), speech audiometry (Freiburg Monosyllabic Test), intraoperative and postoperative complication rate, and patient satisfaction (Glasgow benefit inventory [GBI]) were assessed.
RESULTS: Air-conduction pure tone average (PTA) was 58.8 ± 8.2 dB HL. Unaided average air-bone gap (ABG) was 33.3 ± 6.2 dB. Aided air-conduction PTA in sound field was 25.2 ± 5.1 dB HL. Aided average ABG was -0.3 ± 7.3 dB. Average functional gain was 33.6 ± 7.2 dB. Mean improvement of GBI was +36.1. No intraoperative complications occurred. During a follow-up period of 8.5 ± 2.2 months, no device failure and no need for revision surgery occurred.
CONCLUSION: Audiometric results of the new bone conduction hearing system are satisfying and comparable to the results of devices that have been applied previously for CHL and MHL. Intraoperatively and postoperatively, no complications were noted.

Entities:  

Mesh:

Year:  2014        PMID: 24448279     DOI: 10.1097/MAO.0000000000000208

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  21 in total

1.  Safety and effectiveness of the Bonebridge transcutaneous active direct-drive bone-conduction hearing implant at 1-year device use.

Authors:  Sébastien Schmerber; O Deguine; M Marx; P Van de Heyning; O Sterkers; I Mosnier; P Garin; B Godey; C Vincent; F Venail; M Mondain; A Deveze; J P Lavieille; A Karkas
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-07-30       Impact factor: 2.503

Review 2.  [Active hearing implants in chronic otitis media].

Authors:  S Lailach; C Müller; N Lasurashvili; H Seidler; T Zahnert
Journal:  HNO       Date:  2021-06       Impact factor: 1.284

3.  Bone-anchored hearing aids in conductive and mixed hearing losses: why do patients reject them?

Authors:  Richard T K Siau; Baljeet Dhillon; Derrick Siau; Kevin M J Green
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-02-19       Impact factor: 2.503

4.  Functional results after Bonebridge implantation in adults and children with conductive and mixed hearing loss.

Authors:  Torsten Rahne; Ingmar Seiwerth; Gerrit Götze; Cornelia Heider; Florian Radetzki; Michael Herzog; Stefan K Plontke
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-11-26       Impact factor: 2.503

5.  Long-term audiological benefit with an active transcutaneous bone-conduction device: a retrospective cohort analysis.

Authors:  Julia Hundertpfund; Jens Eduard Meyer; Attila Ovari
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-08-23       Impact factor: 2.503

6.  Three-Year Follow-Up with the Bone Conduction Implant.

Authors:  Ann-Charlotte Persson; Sabine Reinfeldt; Bo Håkansson; Cristina Rigato; Karl-Johan Fredén Jansson; Måns Eeg-Olofsson
Journal:  Audiol Neurootol       Date:  2020-04-08       Impact factor: 1.854

7.  The bone conduction implant: Clinical results of the first six patients.

Authors:  Sabine Reinfeldt; Bo Håkansson; Hamidreza Taghavi; Karl-Johan Fredén Jansson; Måns Eeg-Olofsson
Journal:  Int J Audiol       Date:  2015-02-23       Impact factor: 2.117

Review 8.  New developments in bone-conduction hearing implants: a review.

Authors:  Sabine Reinfeldt; Bo Håkansson; Hamidreza Taghavi; Måns Eeg-Olofsson
Journal:  Med Devices (Auckl)       Date:  2015-01-16

Review 9.  Implantable hearing devices.

Authors:  Matthias Tisch
Journal:  GMS Curr Top Otorhinolaryngol Head Neck Surg       Date:  2017-12-18

10.  The Prediction of Speech Recognition in Noise With a Semi-Implantable Bone Conduction Hearing System by External Bone Conduction Stimulation With Headband: A Prospective Study.

Authors:  Friedrich Ihler; Jenny Blum; Max-Ulrich Berger; Bernhard G Weiss; Christian Welz; Martin Canis
Journal:  Trends Hear       Date:  2016-10-03       Impact factor: 3.293

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.