Literature DB >> 24448180

Robot-assisted stereotactic laser ablation in medically intractable epilepsy: operative technique.

Jorge Gonzalez-Martinez1, Sumeet Vadera, Jeffrey Mullin, Rei Enatsu, Andreas V Alexopoulos, Ravish Patwardhan, William Bingaman, Imad Najm.   

Abstract

BACKGROUND: Stereotactic laser ablation offers an advantage over open surgical procedures for treatment of epileptic foci, tumors, and other brain pathology. Robot-assisted stereotactic laser ablation could offer an accurate, efficient, minimally invasive, and safe method for placement of an ablation catheter into the target.
OBJECTIVE: To determine the feasibility of placement of a stereotactic laser ablation catheter into a brain lesion with the use of robotic assistance, via a safe, accurate, efficient, and minimally invasive manner.
METHODS: A laser ablation catheter (Visualase, Inc) was placed by using robotic guidance (ROSA, Medtech Surgical, Inc) under general anesthesia into a localized epileptogenic periventricular heterotopic lesion in a 19-year-old woman with 10-year refractory focal seizure history. The laser applicator (1.65 mm diameter) position was confirmed by using magnetic resonance imaging (MRI). Ablation using the Visualase system was performed under multiplanar imaging with real-time thermal imaging and treatment estimates in each plane. A postablation MRI sequence (T1 postgadolinium contrast injection) was used to immediately confirm the ablation.
RESULTS: MRI showed accurate skin entry point and trajectory, with the applicator advanced to the lesion's distal boundary. Ablation was accomplished in less than 3 minutes of heating. The overall procedure, from time of skin incision to end of last ablation, was approximately 90 minutes. After confirmation of proper lesioning by using a T1 contrast-enhanced MRI, the applicator was removed, and the incision was closed using a single stitch. No hemorrhage or other untoward complication was visualized. The patient awoke without any complication, was observed overnight after admitting to a regular floor bed, and was discharged to home the following day.
CONCLUSION: This technique, using a combination of Visualase laser ablation, ROSA robot, and intraoperative MRI, facilitated a safe, efficacious, efficient, and minimally invasive approach that could be used for placement of 1 or multiple electrodes in the future.

Entities:  

Mesh:

Year:  2014        PMID: 24448180     DOI: 10.1227/NEU.0000000000000286

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  21 in total

1.  An Integrated Robotic System for MRI-Guided Neuroablation: Preclinical Evaluation.

Authors:  Niravkumar A Patel; Christopher J Nycz; Paulo A Carvalho; Katie Y Gandomi; Radian Gondokaryono; Gang Li; Tamas Heffter; Everette Clif Burdette; Julie G Pilitsis; Gregory S Fischer
Journal:  IEEE Trans Biomed Eng       Date:  2020-02-17       Impact factor: 4.538

2.  The comparative accuracy of the ROSA stereotactic robot across a wide range of clinical applications and registration techniques.

Authors:  Nicholas J Brandmeir; Sandip Savaliya; Pratik Rohatgi; Michael Sather
Journal:  J Robot Surg       Date:  2017-05-08

Review 3.  [Stereotactic laser thermocoagulation in epilepsy surgery].

Authors:  C Hoppe; J-A Witt; C Helmstaedter; T Gasser; H Vatter; C E Elger
Journal:  Nervenarzt       Date:  2017-04       Impact factor: 1.214

4.  Epilepsy: preserving memory in temporal lobectomy-are networks the key?

Authors:  Michael Duchowny; Sanjiv Bhatia
Journal:  Nat Rev Neurol       Date:  2014-04-22       Impact factor: 42.937

Review 5.  Laser Interstitial Thermal Therapy for Epilepsy.

Authors:  Eric Prince; Shahin Hakimian; Andrew L Ko; Jeffrey G Ojemann; Michelle S Kim; John W Miller
Journal:  Curr Neurol Neurosci Rep       Date:  2017-09       Impact factor: 5.081

6.  Robot-assisted laser tissue soldering system.

Authors:  Svetlana Basov; Amit Milstein; Erez Sulimani; Max Platkov; Eli Peretz; Marcel Rattunde; Joachim Wagner; Uri Netz; Abraham Katzir; Ilana Nisky
Journal:  Biomed Opt Express       Date:  2018-10-19       Impact factor: 3.732

7.  Stereo-Electro-Encephalo-Graphy (SEEG) With Robotic Assistance in the Presurgical Evaluation of Medical Refractory Epilepsy: A Technical Note.

Authors:  Jeffrey P Mullin; Saksith Smithason; Jorge Gonzalez-Martinez
Journal:  J Vis Exp       Date:  2016-06-13       Impact factor: 1.355

Review 8.  Seizure outcomes in nonresective epilepsy surgery: an update.

Authors:  Dario J Englot; Harjus Birk; Edward F Chang
Journal:  Neurosurg Rev       Date:  2016-05-21       Impact factor: 3.042

Review 9.  Emerging surgical therapies in the treatment of pediatric epilepsy.

Authors:  Michael Karsy; Jian Guan; Katrina Ducis; Robert J Bollo
Journal:  Transl Pediatr       Date:  2016-04

Review 10.  Multimodal imaging for improved diagnosis and treatment of cancers.

Authors:  Clare M C Tempany; Jagadeesan Jayender; Tina Kapur; Raphael Bueno; Alexandra Golby; Nathalie Agar; Ferenc A Jolesz
Journal:  Cancer       Date:  2014-09-09       Impact factor: 6.860

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.