Dose calculation and treatment plan optimization including imaging dose from kilovoltage cone beam computed tomography.

Parham Alaei1, Emiliano Spezi, Margaret Reynolds.   

Abstract

BACKGROUND: With the increasing use of cone beam computed tomography (CBCT) for patient position verification and radiotherapy treatment adaptation, there is an increasing need to develop techniques that can take into account concomitant dose using a personalized approach.
MATERIAL AND METHODS: A total of 20 patients (10 pelvis and 10 head and neck) who had undergone radiation therapy using intensity modulated radiation therapy (IMRT) were selected and the dose from kV CBCT was retrospectively calculated using a treatment planning system previously commissioned for this purpose. The imaging dose was added to the CT images used for treatment planning and the difference in its addition prior to and after the planning was assessed.
RESULTS: The additional isocenter dose as a result of daily CBCT is in the order of 3-4 cGy for 35-fraction head and neck and 23-47 cGy for 25-fraction pelvis cases using the standard head and neck and pelvis image acquisition protocols. The pelvic dose is especially dependent on patient size and body mass index (BMI), being higher for patients with lower BMI. Due to the low energy of the kV CBCT beam, the maximum energy deposition is at or near the surface with the highest dose being on the patient's left side for the head and neck (∼7 cGy) and on the posterior for the pelvic cases (∼80 cGy). Addition of imaging dose prior to plan optimization resulted in an average reduction of 4% in the plan monitor units and 5% in the number of control points.
CONCLUSION: Dose from daily kV CBCT has been added to patient treatment plans using previously commissioned kV CBCT beams in a treatment planning system. Addition of imaging dose can be included in IMRT treatment plan optimization and would facilitate customization of imaging protocol based on patient anatomy and location of isocenter.

Entities:  

Mesh:

Year:  2014        PMID: 24438661     DOI: 10.3109/0284186X.2013.875626

Source DB:  PubMed          Journal:  Acta Oncol        ISSN: 0284-186X            Impact factor:   4.089


  9 in total

1.  Technical Note: Imaging dose resulting from optimized procedures with limited-angle intrafractional verification system during stereotactic body radiation therapy lung treatment.

Authors:  George X Ding; Yawei Zhang; Lei Ren
Journal:  Med Phys       Date:  2019-04-29       Impact factor: 4.071

2.  Use of a laser-guided collimation system to perform direct kilovoltage x-ray spectra measurements on a linear accelerator onboard imager.

Authors:  Zachary Grelewicz; Andrew H Belcher; Rodney D Wiersma
Journal:  Med Phys       Date:  2018-10-09       Impact factor: 4.071

3.  Dosimetric impact of gastrointestinal air column in radiation treatment of pancreatic cancer.

Authors:  Neil C Estabrook; Jonathan B Corn; Marvene M Ewing; Higinia R Cardenes; Indra J Das
Journal:  Br J Radiol       Date:  2017-12-22       Impact factor: 3.039

4.  Evaluation of Dose Distribution and Normal Tissue Complication Probability of a Combined Dose of Cone-Beam Computed Tomography Imaging with Treatment in Prostate Intensity-Modulated Radiation Therapy.

Authors:  Tetsuya Tomita; Tomonori Isobe; Yoshinobu Furuyama; Hideyuki Takei; Daisuke Kobayashi; Yutaro Mori; Toshiyuki Terunuma; Eisuke Sato; Hiroshi Yokota; Takeji Sakae
Journal:  J Med Phys       Date:  2020-07-20

5.  Influence of daily imaging on plan quality and normal tissue toxicity for prostate cancer radiotherapy.

Authors:  Katharina Bell; Marina Heitfeld; Norbert Licht; Christian Rübe; Yvonne Dzierma
Journal:  Radiat Oncol       Date:  2017-01-10       Impact factor: 3.481

Review 6.  Cancer risk assessment in modern radiotherapy workflow with medical big data.

Authors:  Fu Jin; Huan-Li Luo; Juan Zhou; Ya-Nan He; Xian-Feng Liu; Ming-Song Zhong; Han Yang; Chao Li; Qi-Cheng Li; Xia Huang; Xiu-Mei Tian; Da Qiu; Guang-Lei He; Li Yin; Ying Wang
Journal:  Cancer Manag Res       Date:  2018-06-22       Impact factor: 3.989

7.  How low can you go? A CBCT dose reduction study.

Authors:  Arthur J Olch; Parham Alaei
Journal:  J Appl Clin Med Phys       Date:  2021-01-15       Impact factor: 2.102

8.  Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment.

Authors:  Katharina Bell; Norbert Licht; Christian Rübe; Yvonne Dzierma
Journal:  Radiat Oncol       Date:  2018-10-01       Impact factor: 3.481

9.  CBCT imaging: a simple approach for optimising and evaluating concomitant imaging doses, based on patient-specific attenuation, during radiotherapy pelvis treatment.

Authors:  Caroline Ordóñez-Sanz; Mark Cowen; Neda Shiravand; Niall D MacDougall
Journal:  Br J Radiol       Date:  2021-08-01       Impact factor: 3.039

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.