P Morelli1, E Porazzi2, M Ruspini3, U Restelli2, G Banfi4.
Abstract
INTRODUCTION: The study objective is to evaluate critical points in the process of pre-analytical histology in an Anatomic Pathology laboratory. Errors are an integral part of human systems, including the complex system of Anatomic Pathology. Previous studies focused on errors committed in diagnosis and did not consider the issues related to the histology preparation of routine processes.Entities:
Mesh:
Year: 2013 PMID: 24396989 PMCID: PMC4718384
Source DB: PubMed Journal: J Prev Med Hyg ISSN: 1121-2233
Errors that may occur in the processes
| Accessioning | Gross dissecting | Processing | Embedding | Tissue cutting and slide mounting | Coloring slides | Labeling and releasing slides |
|---|---|---|---|---|---|---|
| The specimen was not in the container | Incorrect numbering of the slides containers | Unfinished program | Loss of specimen | Loss / exhaustion of specimen | Insufficient time for dewaxing | Error of number reported on the slide labels |
| Inconsistency between the specimen and request | Incorrect numbering of the containers of biological material | Instrumental error / error in temperature | Contamination | Number was reported incorrectly | Exhaustion of reactive | Exchange of slides |
| Request or supply not received | Error in checking of the request | Exchange of reagents | Exchange of specimen | Collection of incorrect section | Wrong coloring | Broken slide |
| Specimen wrongly accessioned | Incorrect choice of the containers | Mistaken choice of the program | Incorrect selection of paraffin | Contamination | Detachment of section from the slide | Slides were not delivered |
| Incomplete request | Loss of specimen | Excessive number of containers | The specimen was badly positioned | Thickness selection error | Breakage of slide | Specimen to be decalcified not reported |
| Allocation number error | Mistaken specimen | Loss of specimen | Damaged sample | Misuse of coloring solutions | Mistaken requests | |
| Worksheet wrongly attached | Contaminated specimen | Error in identification of block to be cut | Error in assembling | |||
| Registration error | The specimen was not decalcified | Error in coloring or lack of coloring | Error due to automatic coloring | |||
| Incorrect type of fixative | The specimen was not loaded or Incorrectly loaded | Collection of slides in incorrect dyes | Error in the choice of the program | |||
| Incorrectly mounted slide |
Severity rate scale adopted in the study.
| Index | Consequence |
|---|---|
| 1 | No effect |
| 2 | Repetition of a process |
| 3 | Repetition of multiple processes |
| 4 | Lengthening of response time |
| 5 | Re-sampling for sampling error |
| 6 | Partial loss of material - re-sampling |
| 7 | Loss of material: repeat sampling (non invasive) |
| 8 | Loss of material: repeat sampling (invasive) |
| 9 | Loss of material: it is not possible to repeat sampling |
| 10 | Incorrect diagnosis |
Fig. 1.Incorrect transcription of the number of acceptance 'fishbone diagram'.
Fig. 2.Flow chart of the histologies processing process.
Type and number of errors identified.
| Type of error | Errors, No. (%) |
|---|---|
| accessioning | 9 (6.5) |
| inconsistency between the specimen and request | 3 |
| requests or supply not received | 3 |
| allocation number error | 1 |
| worksheets wrongly attached to the request | 2 |
| gross dissecting | 37 (28) |
| incorrect numbering of the slides container | 26 |
| errors in checking of the request | 3 |
| incorrect choice of the containers | 3 |
| specimen not decalcified | 3 |
| specimen incorrectly loaded | 2 |
| processing | 2 (1.5) |
| exchange of reagents | 1 |
| instrumental error | 1 |
| embedding | 6 (4.5) |
| specimen was badly positioned | 5 |
| contamination | 1 |
| tissue cutting and slide mounting | 30 (23) |
| error in identification of block to be cut | 7 |
| the number was reported incorrectly | 15 |
| lack of coloring information | 6 |
| collection of incorrect section | 1 |
| thickness selection error | 1 |
| coloring slides | 2 (1.5) |
| misuse of coloring solutions | 1 |
| detachment of the section from the slide | 1 |
| labeling and releasing slides | 46 (35) |
| errors of number reported on the slide label | 44 |
| slides were not delivered | 2 |
| Total | 132 (100) |
Fig. 3.Incorrect transcription of the specimen identification number 'Five whys' technique results.