Literature DB >> 24386522

Investigation of Water Safety in Non-treated Drinking Water with Trace Toxic Metals.

Suw Young Ly1, Dae Hong Kim2, Ga Eun Lee2.   

Abstract

The trace toxic metal copper was assayed using mercury immobilized on a carbon nanotube electrode (MCW), with a graphite counter and a reference electrode. In this study, a macro-scale convection motor was interfaced with a MCW three-electrode system, in which a handmade MCW was optimized using cyclic- and square-wave stripping voltammetry. An analytical electrolyte for tap water was used instead of an expensive acid or base ionic solution. Under these conditions, optimum parameters were 0.09 V amplitude, 40 Hz frequency, 0.01 V incremental potential, and a 60-s accumulation time. A diagnostic working curve was obtained from 50.0 to 350 μg/L. At a constant Cu(II) concentration of 10.0 μg/L, the statistical relative standard deviation was 1.78% (RSD, n = 15), the analytical accumulation time was only 60 s, and the analytical detection limit approached 4.6 μg/L (signal/noise = 3). The results were applied to nontreated drinking water. The content of the analyzed copper using 9.0 and 4.0 μg/L standards were 8.68 μg/L and 3.96 μg/L; statistical values R(2) = 0.9987 and R(2) = 0.9534, respectively. This method is applicable to biological diagnostics or food surveys.

Entities:  

Keywords:  Copper; Diagnosis; Toxic metal; Trace; Voltammetry

Year:  2013        PMID: 24386522      PMCID: PMC3878001          DOI: 10.5487/TR.2013.29.3.211

Source DB:  PubMed          Journal:  Toxicol Res        ISSN: 1976-8257


INTRODUCTION

Trace copper is an essential metal for biological components of many enzymatic reactions. Nevertheless, concentrated in vivo, a high level of trace metals is associated with neurological disorders such as Alzheimer’s disease, Parkinson’s disease, prion diseases (1), and ischemic heart disease (2). Here metals were found in drinking water (3), food systems of agricultural soil water (4) and in vivo fluids. For in vivo or in vitro diagnostics, very low analytical detection limits (DLs) are needed. Recently, various analytical methods have been sought such as X-ray fluorescence detection (5), X-ray microprobe assay (6), atomic absorption spectrometry (7) and the energy-dispersive X-ray fluorescence method (8). Some of these spectrics, however, require high energy amplification, pretreated sample preparation, and other detection systems. They are complicated, time-consuming, used only in laboratory conditions, and not usable for in vivo or ex vivo direct assay. On the other hand, voltammetric stripping techniques are very simple and compact. Recently, specified trace sensors were used such as infrared photo diode electrodes (9), solid paraffin-based carbon paste electrodes (10) and others. However in this study, more sensitive mercury (11) immobilized on a carbon nanotube (9) paste (12) probe was developed, the catalytic effects of which were more sensitive and with which optimum results were attained that considerably lowered the detection limit from that of previous methods. Moreover, though expensive acid or base electrolyte solutions were not used in this study, the results of this study were applied to a directly non-treated raw fluid for ex vivo diagnostics.

MATERIALS AND METHODS

The voltammetric experiments were performed with the second version of Bioelectronics- 2 from the authors’ institution. It was combined with three electrode systems: mercury-immobilized MCW and two graphite pencils as the platinum counter and the Ag/AgCl reference, the circuit of which was a computerized, handheld compact system similar to a cellular phone and had a ± 3.0 V potential range, a 2 mA current range, and a 10 pA detection current. An automic absorption standard 1,000 mg/L Cu(II) solution was obtained from Aldrich. All the experiment solutions were prepared with high-purity distilled water (18 MΩcm−1) from a millipore water purification system. Modified MCW was prepared by mixing for 30 min weighed amounts of paste, highly purified CNT, mineral oil (Aldrich), and 1,000 ppm-standard mercury, with weight percentages of 40, 40, and 20%, respectively. Then portions of the paste were packed at the end of a handmade cylindrical tube with a 0.3 mm diameter, and a copper wire was connected to the handheld voltammetric system. All the experiments were performed at room temperature, without removing the oxygen; and under these conditions, cyclic and stripping voltammetric optimizations were performed.

RESULTS AND DISCUSSION

In vivo or ex vivo surveys are dependent on the analytical detection limit. For this reason, optimum voltammetric parameters were examined via cyclic and stripping voltammetry. Moreover, better and inexpensive tap water was used instead of the expensive electrolyte using 10 ml non-treated water with −2 V initial +2 V switching potentials and a 0.5 mV/s scan potential. The cyclic peak potentials are shown in the Fig. 1A inset curves, and the main curves show the results of the cathodic stripping voltammetry, wherein the water blank is simple and no peak appeared. Thus, the 1 mg/L Cu(II) standard 1,000 ml concentration was spiked with a 0-sec accumulation time, while there was no peak. Thus, a 30-sec accumulation time was used. The second curve is that of 0.319 × 10−6 A, where a −0.5 V cathodic potential appeared, after which spiking was continued at a 2~7 mg/L standard. The statistical linear equation was y = 0.302x + 0.247, and the relative standard was R2 = 0.955. Under these conditions, anodic stripping voltammetry was performed, and the voltammetric results are shown in Fig. 1B. The spiking was varied at 1, 2, 3, 4, 5, 6, and 7 mg/L, and the peak current increased continually from 0.549 to 1.751 × 10−6 A. The working curve was y = 0.2x + 0.2999, and the statistic was R2 = 0.9844, which is usable for high ranges, though reduction stripping is more sensitive and a big current appeared. Cyclic oxidation is also better than reduction. Under these conditions, the stripping voltammetric optimization was examined, and the final results were a 0.09 V amplitude, a 40 Hz frequency, a 0.01 V incremental potential, and a 60-sec accumulation time. These conditions were used to determine the analytically possible interference effects, and were calibrated using standard addition methods. Under these conditions, the analytical working ranges and diagnostic applications were used.
Fig. 1.

Sensor properties. (A) 1~8mg/L Cu(II) variation with a cathodic SW; the inset curve is the CV for a 1~7mg/L continuous spike. (B) 1~7mg/L variations via SW anodic stripping, under a 10 ml tap water electrolyte. For the CV, the following conditions were used: a −2.0 V initial potential, a 2.0 V switching potential, and a 0.5 V/s scan rate. For the SW, the optimum parameters were used.

Under the optimized SW conditions, the analytical working ranges were eamined using a 120 s accumulation time. Fig. 2A shows the two ranges of the working curves. The first curve was obtained at 10, 20, 30, 40, 50, 60, 70, and 80 μg/L. The inset voltammograms show the anodic stripping, wherein the statistic slop was Δx/Δy = 13.75, and the relative standard deviation was R2 = 0.97, which was applied to drinking water. In this figure, a higher range was obtained at 50, 100, 150, 200, 250, 300, and 350 μg/L, and the slop sensitive was Δx/Δy = 5.48 and the relative standard deviation, R2 = 0.99. That can be used for food composition and concentrated soil analysis. Fig. 2B shows the applications for the contaminated water sample, which was fixed by a solution that contained the standard 1 mg/L Cu(II), in which 5 ml of the known sample and 5 ml of the tap water solution were stripped under the optimum parameters. Anodic stripping is also shown in this figure. The first curve shows a −0.4 V peak potential with 2,412 × 10−6 A obtained, after which the 1,000 mg/L standard 0.01 ml spiked sequential anodic voltammograms increased at the same −0.4 V potential from 7.31, 12.49, 17.18, and 21.29 × 10−6 A. The statistical working curve was y = 5.16x + 2.80, R2 = 0.99 and the error rate was 9%. 1.09 mg/L Cu(II) obtained, the results were sensitive, and the peak voltammograms were linearly sharp. They can be used for food or any other system. The probe was statistically examined, and the 7th repeated standard addition methods were examined using a 1 mg/L tap water constant. Fig. 2C shows the results on a histogram. Each of the results was calculated at 12, 8, 7, 9, 6, 8, and 5% error rates, and the relative standard deviation was 0.023. Under these conditions, a real sample was examined to develop a commercial-market food drinking water.
Fig. 2.

(A): Square wave micro-working range of 50~350 ppb μg/L and 10~80 μg/L Cu(II) variations. (B): Standard addition method for 1.0mg/L Cu(II) in tap water with a pH of 7.0, using an SW amplitude of 0.08 V, a frequency of 25 Hz, an incremental potential of 10 mV, an accumulation potential of −1.1 V, and an accumulation time of 120 s. (C): statistic 7th repeated standard addition methods of a 1mg/L tap water constant.

The developed results were used to detect the trace range in the copper-rich soil, at which site the same samples were examined five times. Fig. 3A shows a picture of the 5th statistical analysis results for the radish sample for the different purchased soil samples. The calculated values were 0.93, 0.98, 1.08, 1.09, and 1.12 mg/L, which indicate that the same slops and the same sensitivity can be applied to other types of soil. Then more advanced applications were examined. Fig. 3B shows the following standard addition methods for market cola. The first curve represents the 5 ml electrolyte with 9.0 μg/L Cu(II) contain and 5 ml drinking cola that had a voltammogram, for a 1.61 × 10−6 A peak current at a −0.18 V peak potential. Thus, the sequential 0.5, 1.0, 1.5, and 2.0 mg/L standards were spiked, and a continuous peak current was obtained from 2.67, 3.61, 4.66, and 5.52 × 10−6 A. The linear equation was y = 0.019x + 1.651, the statistic was R2 = 0.9987, and the containing was 8.68 μg/L. A more advanced experiment was performed directly on a real non-treated water sample. Fig. 3C shows raw voltammograms of commercial drink solutions for soda pop. The voltammetry was directly performed using 4.0 μg/L Cu(II) contained non-treated water. First, an undiluted 10 ml original fluid was oxidized via stripping voltammetry. The voltammogram is shown in this figure. The first curve shows the same potential of 0.2 V and the current of 0.311 × 10−5 A. Thus, standard 50, 100, 150, and 200 μg/L sequence spiking procedures were performed, and the oxidation peaks appeared at 1.93, 2.69, 3.35, and 4.01 × 10−6 A, the statistic linear equation was y = 0.0176x + 0.6976, the relative standard deviation was R2 = 0.9534, and the calculated value was 3.963 μg/L. Thus, the developed methods can be applied directly using carbonated soda drinks.
Fig. 3.

(A) SW application for standard addition by a copperrich soil radish using five different purchased samples. (B) Market drinking water sample of cola. (C) Soda pop using an SW amplitude of 0.09 V, a frequency of 40 Hz, an incremental potential of 10 mV, an accumulation potential of −1.1 V, and an accumulation time of 60 s.

The novel experiments described in this paper showed that trace metal ranges in MCW can be investigated for drinking water systems. Under optimum conditions, the developed parameters were a 40 Hz SW frequency, a 0.01 V incremental SW potential, an 0.08 V amplitude, and a 1.0 V anodic accumulation potential. These conditions, the simplified probe sensors were shown to be more sensitive than common spectro photometric and voltammetric methods. The handmade combination sensors were successfully used to directly detect trace metals in non-treated raw water without using any electrolyte.
  8 in total

1.  Synchrotron-based infrared and X-ray imaging shows focalized accumulation of Cu and Zn co-localized with beta-amyloid deposits in Alzheimer's disease.

Authors:  Lisa M Miller; Qi Wang; Tejas P Telivala; Randy J Smith; Antonio Lanzirotti; Judit Miklossy
Journal:  J Struct Biol       Date:  2005-11-14       Impact factor: 2.867

2.  Detection of dopamine in the pharmacy with a carbon nanotube paste electrode using voltammetry.

Authors:  Suw Young Ly
Journal:  Bioelectrochemistry       Date:  2005-11-23       Impact factor: 5.373

3.  Toxicity of copper on rice growth and accumulation of copper in rice grain in copper contaminated soil.

Authors:  Jiakuan Xu; Lianxin Yang; Ziqiang Wang; Guichun Dong; Jianye Huang; Yulong Wang
Journal:  Chemosphere       Date:  2005-08-02       Impact factor: 7.086

4.  Aluminum and copper in drinking water enhance inflammatory or oxidative events specifically in the brain.

Authors:  Angelica Becaria; Debomoy K Lahiri; Stephen C Bondy; DeMao Chen; Ali Hamadeh; Huihui Li; Russell Taylor; Arezoo Campbell
Journal:  J Neuroimmunol       Date:  2006-05-11       Impact factor: 3.478

5.  Trace element determination of Argentine wines using ETAAS and USN-ICP-OES.

Authors:  R Lara; S Cerutti; J A Salonia; R A Olsina; L D Martinez
Journal:  Food Chem Toxicol       Date:  2005-02       Impact factor: 6.023

6.  Serum copper as a marker of inflammation in prediction of short term outcome in high risk patients with chronic heart failure.

Authors:  F Malek; E Jiresova; A Dohnalova; H Koprivova; R Spacek
Journal:  Int J Cardiol       Date:  2006-07-14       Impact factor: 4.164

7.  Copper determination in ethanol fuel by differential pulse anodic stripping voltammetry at a solid paraffin-based carbon paste electrode modified with 2-aminothiazole organofunctionalized silica.

Authors:  Regina M Takeuchi; André L Santos; Pedro M Padilha; Nelson R Stradiotto
Journal:  Talanta       Date:  2006-06-27       Impact factor: 6.057

8.  Simultaneous determination of copper and iron in automotive gasoline by X-ray fluorescence after pre-concentration on cellulose paper.

Authors:  Leonardo S G Teixeira; Rodrigo B S Rocha; Eledir V Sobrinho; Paulo R B Guimarães; Luiz A M Pontes; Josanaide S R Teixeira
Journal:  Talanta       Date:  2007-01-09       Impact factor: 6.057

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.