Ashish Kumar Jha1, Nilendu C Purandare1, Sneha Shah1, Archi Agrawal1, Ameya D Puranik1, Venkatesh Rangarajan1.
Abstract
Artifacts present in computed tomography (CT) image often degrade the image quality and ultimately, the diagnostic outcome. Ring artifact in trans-axial image is caused by either miscalibrated or defective detector element of detector row, which is often categorized as scanner based artifact. A ring artifact detected on trans-axial CT image of positron emission tomography/computed tomography (PET/CT), was caused by contamination of CT tube aperture by droplet of injectable contrast medium. This artifact was corrected by removal of contrast droplet from CT tube aperture. The ring artifact is a very common artifact, commonly cited in the literature. Our case puts forward an uncommon cause of this artifact and its method of correction, which also, has no mention in the existing literature.Entities:
Keywords: Artifact; computed tomography tube; computed tomography tube aperture; mylar window; trans-axial
Year: 2013 PMID: 24379535 PMCID: PMC3866670 DOI: 10.4103/0972-3919.121970
Source DB: PubMed Journal: Indian J Nucl Med ISSN: 0974-0244
Figure 1Clinical image (a) Trans-axial image showing a ring artifact (arrows), (b) trans-axial image showing no artifact