Removal of ring artifacts in CT imaging through detection and correction of stripes in the sinogram.

Emran Mohammad Abu Anas1, Soo Yeol Lee, Md Kamrul Hasan.   

Abstract

Due to malfunctioning and mis-calibration of cells in digital x-ray detectors as well as impurities on the scintillator screens, stripe artifacts arise in the sinogram which in turn generate ring artifacts in the reconstructed x-ray computed tomography images. In this paper, a novel technique is proposed for the detection and removal of stripe artifacts in a sinogram with a view to suppress the ring artifacts from the tomographic images. To accurately detect the stripe creating pixels using a derivative-based algorithm, at first the sinogram is windowed to create a sub-sinogram by keeping the pixel of examination at the center position in the sub-sinogram. The other pixels in the sub-sinogram are selected from a polyphase component of the sinogram. A new mathematical index is proposed here to isolate the strong and weak ring-generating stripes from the good ones. For the correction of strong ring artifacts resulting from the defective detector elements and dusty scintillator crystals, 2D variable window moving average and weighted moving average filters are proposed in this work. On the other hand, a conventionally trusted constant bias correction scheme is adopted to correct the responses of the mis-calibrated detector elements. To evaluate and compare the performance of the proposed algorithm, real micro-CT images acquired from two flat panel detectors under different operating conditions are used. Experimental results show that the proposed method can remove ring artifacts more effectively without imparting noticeable distortion in the image as compared to a recently reported technique in the literature.

Mesh:

Year:  2010        PMID: 21048294     DOI: 10.1088/0031-9155/55/22/020

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  9 in total

1.  A new iterative algorithm for ring artifact reduction in CT using ring total variation.

Authors:  Morteza Salehjahromi; Qian Wang; Yanbo Zhang; Lars A Gjesteby; Dan Harrison; Ge Wang; Peter M Edic; Hengyong Yu
Journal:  Med Phys       Date:  2019-09-20       Impact factor: 4.071

2.  Sparsity-based method for ring artifact elimination in computed tomography.

Authors:  Mona Selim; Essam A Rashed; Mohammed A Atiea; Hiroyuki Kudo
Journal:  PLoS One       Date:  2022-06-28       Impact factor: 3.752

3.  Comparison of ring artifact removal methods using flat panel detector based CT images.

Authors:  Emran M Abu Anas; Jae G Kim; Soo Y Lee; K Hasan
Journal:  Biomed Eng Online       Date:  2011-08-17       Impact factor: 2.819

4.  John's Equation-based Consistency Condition and Corrupted Projection Restoration in Circular Trajectory Cone Beam CT.

Authors:  Jianhui Ma; Shuyu Wu; Hongliang Qi; Bin Li; Hao Yan; Linghong Zhou; Yuan Xu
Journal:  Sci Rep       Date:  2017-07-07       Impact factor: 4.379

5.  A Ring Artifact Correction Method: Validation by Micro-CT Imaging with Flat-Panel Detectors and a 2D Photon-Counting Detector.

Authors:  Mohamed Elsayed Eldib; Mohamed Hegazy; Yang Ji Mun; Myung Hye Cho; Min Hyoung Cho; Soo Yeol Lee
Journal:  Sensors (Basel)       Date:  2017-01-30       Impact factor: 3.576

6.  GPU-accelerated ray-casting for 3D fiber orientation analysis.

Authors:  Roman Shkarin; Svetlana Shkarina; Venera Weinhardt; Roman A Surmenev; Maria A Surmeneva; Andrei Shkarin; Tilo Baumbach; Ralf Mikut
Journal:  PLoS One       Date:  2020-07-29       Impact factor: 3.240

7.  Complete Ring Artifacts Reduction Procedure for Lab-Based X-ray Nano CT Systems.

Authors:  Jakub Šalplachta; Tomáš Zikmund; Marek Zemek; Adam Břínek; Yoshihiro Takeda; Kazuhiko Omote; Jozef Kaiser
Journal:  Sensors (Basel)       Date:  2021-01-01       Impact factor: 3.576

8.  Ring artifact and Poisson noise attenuation via volumetric multiscale nonlocal collaborative filtering of spatially correlated noise.

Authors:  Ymir Mäkinen; Stefano Marchesini; Alessandro Foi
Journal:  J Synchrotron Radiat       Date:  2022-04-21       Impact factor: 2.557

9.  Identification of a unique cause of ring artifact seen in computed tomography trans-axial images.

Authors:  Ashish Kumar Jha; Nilendu C Purandare; Sneha Shah; Archi Agrawal; Ameya D Puranik; Venkatesh Rangarajan
Journal:  Indian J Nucl Med       Date:  2013-10
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.