Evaluation of Gafchromic EBT2 film for the measurement of anisotropy function for high-dose-rate (192)Ir brachytherapy source with respect to thermoluminescent dosimetry.

Satish C Uniyal1, Umesh C Naithani2, Sunil D Sharma3.   

Abstract

AIM: The aim of this work was to assess the suitability of the use of a Gafchromic EBT2 film for the measurement of anisotropy function for microSelectron HDR (192)Ir (classic) source with a comparative dosimetry method using a Gafchromic EBT2 film and thermoluminescence dosimeters (TLDs).
BACKGROUND: Sealed linear radiation sources are commonly used for high dose rate (HDR) brachytherapy treatments. Due to self-absorption and oblique filtration of radiation in the source capsule material, an inherent anisotropy is present in the dose distribution around the source which can be described by a measurable two-dimensional anisotropy function, F(r, θ).
MATERIALS AND METHODS: Measurements were carried out in a specially designed and locally fabricated PMMA phantom with provisions to accommodate miniature LiF TLD rods and EBT2 film dosimeters at identical radial distances with respect to the (192)Ir source.
RESULTS: The data of anisotropy function generated by the use of the Gafchromic EBT2 film method are in agreement with their TLD measured values within 4%. The produced data are also consistent with their experimental and Monte Carlo calculated results for this source available in the literature.
CONCLUSION: Gafchromic EBT2 film was found to be a feasible dosimeter in determining anisotropy in the dose distribution of (192)Ir source. It offers high resolution and is a viable alternative to TLD dosimetry at discrete points. The method described in this paper is useful for comparing the performances of detectors and can be applied for other brachytherapy sources as well.

Keywords:  Anisotropy function; Gafchromic EBT2 film; HDR 192Ir source; TL dosimetry

Year:  2010        PMID: 24376951      PMCID: PMC3863320          DOI: 10.1016/j.rpor.2010.11.003

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  14 in total

1.  In-phantom response of LiF TLD-100 for dosimetry of 192Ir HDR source.

Authors:  A S Pradhan; U Quast
Journal:  Med Phys       Date:  2000-05       Impact factor: 4.071

2.  Radiochromic film measurement of anisotropy function for high-dose-rate Ir-192 brachytherapy source.

Authors:  S D Sharma; C Bianchi; L Conte; R Novario; B C Bhatt
Journal:  Phys Med Biol       Date:  2004-09-07       Impact factor: 3.609

3.  Dosimetry parameters of BARC OcuProsta I-125 seed source.

Authors:  S D Sharma; M Basu; A Shanta; T Palani Selvam; U B Tripathi; B C Bhatt
Journal:  Australas Phys Eng Sci Med       Date:  2005-03       Impact factor: 1.430

4.  Monte Carlo and TLD dosimetry of an 192Ir high dose-rate brachytherapy source.

Authors:  P Karaiskos; A Angelopoulos; L Sakelliou; P Sandilos; C Antypas; L Vlachos; E Koutsouveli
Journal:  Med Phys       Date:  1998-10       Impact factor: 4.071

5.  Experimental determination of dosimetry functions of Ir-192 sources.

Authors:  J C Anctil; B G Clark; C J Arsenault
Journal:  Med Phys       Date:  1998-12       Impact factor: 4.071

6.  Dosimetric characteristics of a new high-intensity 192Ir source for remote afterloading.

Authors:  A S Meigooni; M T Kleiman; J L Johnson; D Mazloomdoost; G S Ibbott
Journal:  Med Phys       Date:  1997-12       Impact factor: 4.071

7.  Anisotropy of an 192iridium high dose rate source measured with a miniature ionization chamber.

Authors:  V Mishra; F M Waterman; N Suntharalingam
Journal:  Med Phys       Date:  1997-05       Impact factor: 4.071

8.  Monte Carlo aided dosimetry of the microselectron pulsed and high dose-rate 192Ir sources.

Authors:  J F Williamson; Z Li
Journal:  Med Phys       Date:  1995-06       Impact factor: 4.071

9.  Anisotropy measurements of a high dose rate Ir-192 source in air and in polystyrene.

Authors:  R Muller-Runkel; S H Cho
Journal:  Med Phys       Date:  1994-07       Impact factor: 4.071

10.  Dosimetric characteristics of 192Ir sources used in interstitial brachytherapy.

Authors:  M Ghiassi-Nejad; M Jafarizadeh; M R Ahmadian-Pour; A R Ghahramani
Journal:  Appl Radiat Isot       Date:  2001-08       Impact factor: 1.513

View more
  1 in total

1.  Gafchromic film dosimetry of a new HDR 192Ir brachytherapy source.

Authors:  Navid Ayoobian; Akbar Sarabi Asl; Hosein Poorbaygi; Mohammad Reza Javanshir
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.