Anisotropy measurements of a high dose rate Ir-192 source in air and in polystyrene.

R Muller-Runkel1, S H Cho.   

Abstract

Anisotropy measurements in air and in polystyrene were performed for a Nucletron microSelectron 10 Ci Ir-192 source. TLD rods of similar sensitivity (+/- 3%) were placed on a semicircle around the source center which had been precisely localized by either radiographic means or direct measurement. The anisotropy, expressed as a ratio of TLD reading at a given angle relative to the longitudinal source axis, over the reading at 90 degrees, was measured for distances between 1 and 10 cm from the source center. For both media, a marked anisotropy of the high dose rate source was found. In air, there is no detectable variation in anisotropy with distance up to 10 cm from the source. In polystyrene, however, the anisotropy becomes less pronounced with increasing distance due to scatter. For both media, no difference, within the estimated experimental accuracy of +/- 5%, was found between three consecutive sources installed by the vendor.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7968845     DOI: 10.1118/1.597339

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Evaluation of Gafchromic EBT2 film for the measurement of anisotropy function for high-dose-rate (192)Ir brachytherapy source with respect to thermoluminescent dosimetry.

Authors:  Satish C Uniyal; Umesh C Naithani; Sunil D Sharma
Journal:  Rep Pract Oncol Radiother       Date:  2010-12-07

2.  Determination of the tissue inhomogeneity correction in high dose rate Brachytherapy for Iridium-192 source.

Authors:  Barlanka Ravikumar; S Lakshminarayana
Journal:  J Med Phys       Date:  2012-01

3.  Physics-aspects of dose accuracy in high dose rate (HDR) brachytherapy: source dosimetry, treatment planning, equipment performance and in vivo verification techniques.

Authors:  Antony Palmer; David Bradley; Andrew Nisbet
Journal:  J Contemp Brachytherapy       Date:  2012-06-30

4.  A study on the dose distributions in various materials from an Ir-192 HDR brachytherapy source.

Authors:  Shih-Ming Hsu; Chin-Hui Wu; Jeng-Hung Lee; Ya-Ju Hsieh; Chun-Yen Yu; Yi-Jen Liao; Li-Cheng Kuo; Ji-An Liang; David Y C Huang
Journal:  PLoS One       Date:  2012-09-05       Impact factor: 3.240