Giovanna Zacchetti1, Anselm Wiskott2, Joël Cugnoni3, John Botsis3, Patrick Ammann4.
Abstract
PURPOSE: To assess the effect of external mechanical microstimuli of controlled magnitude on the microarchitecture of the peri-implant bone beds in rat tibiae.Entities:
Mesh:
Substances:
Year: 2013 PMID: 24369009 PMCID: PMC3866820 DOI: 10.1155/2013/234093
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Radiograph of the implants inserted into the proximal tibial metaphysis. Diagrammatic view showing the position of the test (left) and the anchorage (right) implants. The 3 regions of interest (ROI) are shown. (a) Peri-implant band of bone located around the implant's most apical segment that was still exclusively surrounded by trabecular bone. Height: 56 slices (ca. 1.12 mm), band width: 0.5 mm. (b) ROI number two. Volume of 50 slices (ca. 0.98 mm) including both the cortical and trabecular compartment of bone. Slice number 1 was located immediately outside the distal surface of the test implant. (c) ROI number three. Volume of 100 slices (ca. 1.98 mm) contoured midway between the two implants and also including both the cortical and the trabecular compartments of bone.
Figure 2Effect of 4 weeks of in vivo mechanical loading on pull-out strength. The pull-out strength measured in the 2 N group was significantly higher compared to NS. *P < 0.05. Means ± SEM. NS: nonstimulated.
Effect of 4 weeks of mechanical loading on bone microarchitecture measured in a 0.5 mm circular band selected around the implant test in rat tibiae.
| NS | 1 N | 2 N | 3 N | |
|---|---|---|---|---|
| BV/TV% | 44.2 ± 3.1 | 50.5 ± 2.3 | 48.5 ± 1.9 | 48.0 ± 3.9 |
| Tb.N (1/mm) | 6.23 ± 0.177 | 6.53 ± 0.200 | 6.02 ± 0.234 | 6.27 ± 0.294 |
| Tb.Th (mm) | 0.142 ± 0.003 | 0.153 ± 0.006 | 0.145 ± 0.003 | 0.149 ± 0.005 |
| Tb.Sp (mm3) | 0.169 ± 0.008 | 0.160 ± 0.012 | 0.174 ± 0.010 | 0.168 ± 0.015 |
| SMI | −0.155 ± 0.316 | −0.595 ± 0.344 | −0.487 ± 0.222 | −0.388 ± 0.370 |
| Contact (%) | 76.02 ± 2.66 | 78.73 ± 2.25 | 78.35 ± 2.19 | 76.85 ± 3.70 |
| POS (N) | 39.57 ± 2.23 | 40.82 ± 3.12 | 46.63 ± 2.21* | 43.81 ± 3.41 |
Mean ± SEM.
*P < 0.05 compared to NS.
POS: pull-out strength.
Effect of 4 weeks mechanical loading on parameters of bone microarchitecture analyzed in a region of 50 lines selected near the implant test in rat tibiae.
| NS | 1 N | 2 N | 3 N | |
|---|---|---|---|---|
| BV/TV% | 14.6 ± 2.2 | 20.1 ± 2.2# | 16.9 ± 2.1 | 13.8 ± 2.2 |
| Tb.N (1/mm) | 3.74 ± 0.27 | 4.26 ± 0.29 | 4.03 ± 0.23 | 3.67 ± 0.14 |
| Tb.Th (mm) | 0.074 ± 0.002 | 0.080 ± 0.003 | 0.074 ± 0.003 | 0.071 ± 0.004 |
| Tb.Sp (mm3) | 0.28 ± 0.03 | 0.24 ± 0.02 | 0.25 ± 0.02 | 0.28 ± 0.01 |
| SMI | 2.7 ± 0.15 | 2.3 ± 0.15 | 2.5 ± 0.14 | 2.7 ± 0.18 |
| Ct.Th (mm) | 0.44 ± 0.01 | 0.43 ± 0.01 | 0.43 ± 0.01 | 0.46 ± 0.01* |
Mean ± SEM.
# P < 0.05 versus 3 N; *P < 0.05 versus 1 N.
Effect of 4 weeks of mechanical loading on parameters of bone microarchitecture analyzed in a region of 100 lines selected between the test and the anchorage implants in rat tibiae.
| NS | 1 N | 2 N | 3 N | |
|---|---|---|---|---|
| BV/TV% | 8.3 ± 1.4 | 11.5 ± 1.2# | 8.2 ± 1.3 | 6.1 ± 1.1 |
| Tb.N (1/mm) | 2.77 ± 0.22 | 3.20 ± 0.22# | 2.58 ± 0.29 | 2.19 ± 0.17 |
| Tb.Th (mm) | 0.071 ± 0.002 | 0.071 ± 0.002 | 0.071 ± 0.002 | 0.071 ± 0.005 |
| Tb.Sp (mm3) | 0.38 ± 0.03 | 0.33 ± 0.03 | 0.47 ± 0.12 | 0.48 ± 0.03 |
| SMI | 2.99 ± 0.10 | 2.78 ± 0.10 | 3.00 ± 0.15 | 3.19 ± 0.14° |
| Ct.Th (mm) | 0.49 ± 0.02 | 0.48 ± 0.02 | 0.48 ± 0.03 | 0.51 ± 0.02 |
Mean ± SEM.
# P < 0.01 versus 3 N; °P < 0.05 versus 1 N.
Effect of 4 weeks of mechanical loading on intrinsic quality of bone analyzed between the test and the anchorage implants in rat tibiae.
| NS | 1 N | 2 N | 3 N | |
|---|---|---|---|---|
| E (GPa) | 14.2 ± 0.5 | 14.5 ± 0.4 | 16.3 ± 0.4∗∗a,b | 14.4 ± 0.4 |
| H (MPa) | 552 ± 20 | 578 ± 17 | 604 ± 17∗b | 528 ± 14 |
| W (pJ) | 3491 ± 115 | 4014 ± 97** | 3785 ± 135 | 4024 ± 92** |
Mean ± SEM.
*P < 0.05 NS; **P < 0.001 versus NS.
a P < 0.01 versus 1 N; b P < 0.01 versus 3 N.
E: modulus of elasticity; H: hardness; W: working energy.
The sum of the effects of 4 weeks of mechanical loading on the determinants of pull-out strength in rat tibiae.
| 1 N | 2 N | 3 N | |
|---|---|---|---|
| Trabecular bone | + | 0 | − |
| Cortical bone | 0 | 0 | + |
| Intrinsic bone quality | ± | + | ± |
| Pull-out strength | 0 | + | ± |