Literature DB >> 24363750

Histopathological Comparison among Biolimus, Zotarolimus and Everolimus-Eluting Stents in Porcine Coronary Restenosis Model.

Kyung Seob Lim1, Myung Ho Jeong2, In Ho Bae3, Dae Sung Park4, Jong Min Kim4, Jung Ha Kim4, Dong Lyun Cho4, Doo Sun Sim4, Keun-Ho Park4, Young Joon Hong4, Youngkeun Ahn4.   

Abstract

BACKGROUND AND OBJECTIVES: The aim of this study was to examine the histolopathogical effects among the biolimus, zotarolimus, and everolimus eluting stent (EES) in the porcine coronary restenosis model. SUBJECTS AND METHODS: Pigs were randomized into three groups in which the coronary arteries (15 pigs, 10 coronaries in each group) had either a biolimus A9 eluting stent (BES, n=10), zotarolimus eluting stent (ZES, n=10) or an EES (n=10). Histopathologic analysis was performed at 28 days after stenting.
RESULTS: There were no significant differences in the injury score among the three groups. There was a significant difference in the internal elastic lamina, lumen area, neointima area, percent area stenosis, and the fibrin and inflammation score among the three groups (4.3±0.53 mm(2), 2.5±0.93 mm(2), 1.8±1.03 mm(2), 40.7±20.80%, 1.7±0.41, 1.4±0.72 in the BES group vs. 5.1±0.55 mm(2), 2.3±1.14 mm(2), 2.8±1.00 mm(2), 55.4±21.23%, 2.0±0.39, 1.6±0.76 in the ZES group vs. 4.4±0.53 mm(2), 1.7±1.22 mm(2), 2.8±1.23 mm(2), 64.0±26.00%, 1.8±0.76, 2.1±0.90 in the EES group, respectively). BES is more effective in inhibiting neointimal hyperplasia compared to ZES and EES (p<0.0001). According to the fibrin and inflammation score, BES and EES are more effective in decreasing the fibrin deposition compared to ZES (p<0.001). Moreover, BES and ZES are more effective in reducing the inflammatory reaction compared to EES (p<0.001).
CONCLUSION: The result demonstrates that BES shows better histopathological characteristics than ZES and EES at one month after stenting in the porcine coronary restenosis model.

Entities:  

Keywords:  Coronary restenosis; Drug-eluting stents; Inflammation; Percutaneous coronary intervention

Year:  2013        PMID: 24363750      PMCID: PMC3866314          DOI: 10.4070/kcj.2013.43.11.744

Source DB:  PubMed          Journal:  Korean Circ J        ISSN: 1738-5520            Impact factor:   3.243


Introduction

Currently, percutaneous coronary intervention (PCI) with the use of antiproliferative agents coated stents, such as the -limus and -taxol families, is the treatment of choice for patients with acute myocardial infarction.1) The introduction of drug eluting stents (DES) has provided improved clinical outcomes compared to the use of bare metal stents (BMS), by reducing neointimal hyperplasia and target vessel revascularizations.2)3) The clinical results of the sirolimus eluting stent (SES) as compared to BMS have provided clues for improved limus drug development.4-6) Biolimus A9, zotarolimus, and everolimus, a potent antiproliferative drug of the -limus families, has been shown to reduce clinical restenosis in clinical trials.7-9) The objective of the present study was to compare the histological effects among the biolimus, zotarolimus, and everolimus eluting stents in the porcine coronary restenosis model.

Subjects and Methods

Animal preparation and stent implantation

The animal study was approved by the Ethics Committee of Chonnam National University Medical School and Chonnam National University Hospital (CNU IACUC-H-2010-18), and conformed to the Guidelines for the Care and Use of Laboratory Animals published by the US National Institutes of Health (NIH Publication No. 85-23, revised 1996). Study animals were castrated male pigs weighing 20-25 kg. To prevent acute thrombosis after stenting, premedication with aspirin 100 mg and clopidogrel 75 mg per day was applied for 5 days before the procedure. On the day of the procedure, pigs were anesthetized with zolazepam and tiletamine (2.5 mg/kg, Zoletil50®, Virbac, Caros, France), xylazine (3 mg/kg, Rompun®, Bayer AG, Leverkusen, Germany) and azaperone (6 mg/kg, Stresnil®, Janssen-Cilag, Neuss, Germany). They received supplemental oxygen continuously through oxygen masks. Subcutaneous 2% lidocaine at the cut-down site was administered, the left carotid artery was surgically exposed, and a 7 Fr sheath was inserted. Continuous hemodynamic and surface electrocardiographic monitoring was maintained throughout the procedure. Then, 5000 units of heparin was administered intravenously as a bolus prior to the procedure, the target coronary artery was engaged using standard 7 Fr guide catheters, and control angiograms of both coronary arteries were performed using a nonionic contrast agent in two orthogonal views. The stent was deployed by inflating the balloon and the resulting stent-to-artery ratio was 1.3 : 1. Coronary angiograms were obtained immediately after stent implantation. Thereafter, all equipment was removed and the carotid artery was ligated. Four weeks after stenting, the animals underwent follow-up angiography in the same orthogonal views as before death with 20 mL of a potassium chloride intracoronary injection. The hearts were removed, and the coronary arteries were pressure-perfusion fixed at 110 mm Hg in 10% neutral buffered formalin overnight. Arteries were step-sectioned, processed routinely for light microscopy, and stained for histological analysis.

Study groups

The pigs were randomly divided into 3 groups (Table 1): group 1 {biolimus A9-eluting stents (BES), BioMatrix®, Biosensors Interventional Technologies Pte Ltd., Singapore, 3.0×18 mm, n=10}, group 2 {zotarolimus-eluting stents (ZES), Endeavor Resolute®, Medtronic CardioVascular, Minneapolis, MN, USA, 3.0×18 mm, n=10}, and group 3 {everolimus-eluting stents (EES), Promus®, Boston Scientific, Natick, MA, USA, 3.0×18 mm, n=10}.
Table 1

Coronary artery morphometric measurements in stented porcine coronary arteries

BES: biolimus A9-eluting stents, ZES: zotarolimus-eluting stents, EES: everolimus-eluting stent, IEL: internal elastic lamina, NS: not significant

A total of 15 pigs were used in this study (15 pigs, 30 coronary arteries, 10 coronary arteries in each group). A BES, a ZES, and an EES were implanted in the left anterior descending artery and left circumflex artery in a randomized manner in each pig.

Histopathology and immunohistochemistry analysis

Histopathologic evaluations of each artery were performed by an experienced cardiovascular pathologist. The specimens were embedded, and sections of 50 to 100 µm in thickness were obtained at about 1 mm distances apart, and stained with Hematoxylin-Eosin (Fig. 1) and Carstairs' (Fig. 2) for histological analysis. Measurements of the histopathologic sections were performed using a calibrated microscope, digital video imaging system, and microcomputer program (Visus 2000 Visual Image Analysis System, IMT Tech, CA, USA). Borders were manually traced for the lumen area, the area circumscribed by the internal elastic lamina (IEL), and the innermost border of the external elastic lamina (external elastic lamina area). Morphometric analysis of the neointimal area for a given vessel was calculated as the measured IEL area minus the lumen area. The measurements were made on five cross-sections from the proximal and distal ends, and the three midpoints of each stented segment. Histopathologic stenosis was calculated as 100×{1-(lesion lumen area/lesion IEL area)}.10) Immunohistochemistry (IHC) was conducted through standard procedures, as previously described.11) Anti-smooth muscle actin monoclonal antibody (Sigma Aldrich, St. Louis, MO, USA) was used. IHC specimens were analyzed by fluorescence microscopy and digital photography. All histology and IHC results were interpreted by two independent pathologists in a blind manner.
Fig. 1

Representative images of H&E staining after 4 weeks of stenting. Specimen BES implanted (A: ×20, A-1: ×200), ZES implanted (B: ×20, B-1: ×200), and EES implanted (C: ×20, C-1: ×200). Inflammatory reaction was more severe in the EES stented artery compared to BES and ZES. BES: biolimus A9-eluting stents, ZES: zotarolimus-eluting stents, EES: everolimus-eluting stents.

Fig. 2

The Carstair fibrin stain of the low-power fields (magnitude, ×20, ×200) of fibrin infiltration in BES implanted (A and A-1), ZES implanted (B and B-1), and EES implanted (C and C-1). Fibrin deposition surrounding the stent struts was higher in ZES than in BES and EES cases. BES: biolimus A9-eluting stents, ZES: zotarolimus-eluting stents, EES: everolimus-eluting stents.

Evaluation of the arterial injury

The arterial injury at each strut site was determined by the anatomic structures penetrated by each strut. A numeric value was assigned, as previously described by Schwartz et al.:10) 0=no injury; 1=break in the internal elastic membrane; 2=perforation of the media; 3=perforation of the external elastic membrane to the adventitia. The average injury score for each segment was calculated by dividing the sum of the injury scores by the total number of struts at the examined section.

Evaluation of inflammation scores, neointimal reaction, and fibrin score

With regard to the inflammation score for each individual strut, the grading was as follows: 0=no inflammatory cells surrounding the strut; 1=light, noncircumferential lymphohistiocytic infiltrate surrounding the strut; 2=localized, moderate to dense cellular aggregate surrounding the strut noncircumferentially; 3=circumferential dense lymphohistiocytic cell infiltration of the strut. The inflammation score for each cross section was calculated by dividing the sum of the individual inflammation scores by the total number of struts at the examined section.12) Ordinal data for fibrin were collected on each stent section using a scale of 0 to 3, as previously reported.13)

Statistical analysis

Statistical analysis was performed with the aid of commercially available software {Statistical Package for the Social Sciences (SPSS) Version 15, SPSS Inc., Chicago, IL, USA}. The data were presented as mean value±SD. An unpaired Student's t-test was used for the comparison of each of the stent groups. Analysis of variance was used to make comparisons of the three stents groups. A value of p<0.05 was considered statistically significant.

Results

Analysis after stenting

Two stents were placed for two coronary arteries per swine. A total of thirty stents, including ten BES, ten ZES, and ten EES, were placed in the proximal left anterior descending and proximal circumflex artery for fifteen swine. The mortality rate in this study was zero. There was no significant difference in the stent-to-artery ratio among the three stent groups.

Histopathological and immunohistochemistry analysis

To exam the characteristics of smooth muscle cells (SMC) in neointima tissue, the stented coronary artery was stained by anti-SMC antibody. Vascular SMC were major components of neointima formation after stenting in all groups (Fig. 3).
Fig. 3

Representative images of immunohistochemistry using anti-smooth muscle actin monoclonal antibody in the neointima tissue. Immunofluorescence staining showing an expression of α-smooth muscle actin (bright red positive cells, ×200, A: BES, B: ZES, C: EES). BES: biolimus A9-eluting stents, ZES: zotarolimus-eluting stents, EES: everolimus-eluting stents.

There was no statistically significant difference in the injury score among the three groups (1.4±0.37 in the BES group vs. 1.4±0.35 in the ZES group vs. 1.5±0.36 in the EES group, p=not significant). There was a statistically significant difference in the IEL among the three groups (4.3±0.53 mm2 in the BES group vs. 5.1±0.55 mm2 in the ZES group vs. 4.4±0.53 mm2 in the EES group, p<0.0001). There was a statistically significant difference in the lumen area among the three groups (2.5±0.93 mm2 in the BES group vs. 2.3±1.14 mm2 in the ZES group vs. 1.7±1.22 mm2 in the EES group, p<0.001). There was a statistically significant difference in the neointima area among the three groups (1.8±1.03 mm2 in the BES group vs. 2.8±1.00 mm2 in the ZES group vs. 2.8±1.23 mm2 in the EES group, p<0.0001). There was a statistically significant difference in percentage area stenosis among the three groups (40.7±20.80% in the BES group vs. 55.4±21.23% in the ZES group vs. 64.0±26.00% in the EES group, p<0.0001). There was a statistically significant difference in the fibrin score among the three groups (1.7±0.41 in the BES group vs. 2.0±0.39 in the ZES group vs. 1.8±0.76 in the EES, p<0.001). There was a statistically significant difference in the inflammation score among the three groups (1.4±0.72 in the BES group vs. 1.6±0.76 in the ZES group vs. 2.1±0.90 in the EES group, p<0.001) (Table 1, Fig. 4).
Fig. 4

Injury score (A), internal elastic lamina (B), lumen area (C), neointima area (D), % area stenosis (E) fibrin score (F) and inflammation score (G), in the BES, ZES, and EES groups. BES: biolimus A9-eluting stents, ZES: zotarolimus-eluting stents, EES: everolimus-eluting stents.

Discussion

This study was conducted in order to compare the histopathological differentiation of the BES, ZES, and EES in the porcine coronary restenosis model. Our study demonstrated that BES was more effective in reducing neointima proliferation compared to ZES and EES. In the fibrin score, which indicates delayed arterial healing, ZES was inferior to BES and EES. BES and ZES were more effective in inhibiting the inflammatory reaction compared to EES, according to the inflammation score. The results demonstrated that BES displayed improved histopathological characteristics in the three limus families. Biolimus A9-eluting stents, EES, and ZES were compared to BMS and/or other generation DES in several clinical trials. Two types of second generation stents, ZES and EES, have shown promising results in clinical trials and registries compared with BMS, SES, and paclitaxel eluting stent (PES).8)14)15) Major adverse cardiac events, death, and myocardial infarction were lower for ZES vs. SES and PES in patients with diabetes mellitus.16) In a comparison study between ZES and EES using optical coherence tomography, neointima proliferation was greater in the ZES group than in the EES group at both 3 and 12 months.17) Other clinical trials comparing ZES and EES, however, have found that both stents demonstrated comparable levels of safety and efficacy.18)19) One year clinical results after 3rd generation BES implantation were as safe and efficacious as those after 2nd generation EES implantation.20) Both stents displayed an excellent low rate of target lesion revascularization and an extremely low rate of stent thrombosis.21) In a 4-year long-term follow-up, BES has shown improved safety and efficacy compared with SES.22) In our previous study using BES, BES appeared to be reliable in terms of inflammation at overlapping segments, as well as at nonoverlapping segments.23) In clinical research, BES displayed a lower rate of the composite of major adverse cardiac events in patients with ST-elevation myocardial infarction undergoing primary PCI compared with BMS.24) The major difference between 2nd and 3rd generation stents is the biodegradable polymer used. The polymers are potentially linked with neointima hyperplasia, inflammation, and late stent thrombosis.25)26) The polymer-free BES demonstrated the equivalent early and superior late inhibition of neointima hyperplasia compared with SES in a porcine model.27) The biolimus-eluting stent with biodegradable polymer (BES) was developed as a third generation DES elutes biolimus A9 from a bioabsorbable polylactic acid (PLA) polymer.28)29) BES releases biolimus A9 into the artery wall while the PLA polymer is absorbed by the contacted coronary vessel tissues. Therefore, this study suggests that the biodegradable polymer of BES achieved superior histopathologic results compared to the permanent polymer of ZES and EES in the porcine coronary restenosis model.

Study limitations

Our study had some limitations. First, we used normal porcine coronary arteries without atherosclerotic lesions, unlike in human clinical situations with pre-existing atherosclerosis. Second, we examined the inflammatory reaction based on H&E stain. IHC techniques were the standard for such studies.30) Third, we did not perform long-term follow-up experiments, such as over 6 months using minipigs. In conclusion, this study shows that BES is more effective in inhibiting neointima hyperplasia compared to ZES and EES. According to fibrin and inflammation score, BES and EES are more effective in decreasing fibrin deposition compared to ZES. Moreover, BES and ZES are more effective in reducing the inflammatory reaction compared to EES. The result demonstrates that BES shows superior histopathological characteristics in BES, ZES, and EES at one month after stenting in the porcine coronary restenosis model.
  30 in total

1.  Final 5-year outcomes from the Endeavor zotarolimus-eluting stent clinical trial program: comparison of safety and efficacy with first-generation drug-eluting and bare-metal stents.

Authors:  David E Kandzari; Martin B Leon; Ian Meredith; Jean Fajadet; William Wijns; Laura Mauri
Journal:  JACC Cardiovasc Interv       Date:  2013-04-17       Impact factor: 11.195

2.  Safety and efficacy of second-generation everolimus-eluting Xience V stents versus zotarolimus-eluting resolute stents in real-world practice: patient-related and stent-related outcomes from the multicenter prospective EXCELLENT and RESOLUTE-Korea registries.

Authors:  Kyung Woo Park; Joo Myung Lee; Si-Hyuck Kang; Hyo-Suk Ahn; Han-Mo Yang; Hae-Young Lee; Hyun-Jae Kang; Bon-Kwon Koo; Janghyun Cho; Hyeon-Cheol Gwon; Sung Yoon Lee; In-Ho Chae; Tae-Jin Youn; Jei Keon Chae; Kyoo-Rok Han; Cheol Woong Yu; Hyo-Soo Kim
Journal:  J Am Coll Cardiol       Date:  2012-12-26       Impact factor: 24.094

3.  Polymer-free biolimus a9-coated stent demonstrates more sustained intimal inhibition, improved healing, and reduced inflammation compared with a polymer-coated sirolimus-eluting cypher stent in a porcine model.

Authors:  Norio Tada; Renu Virmani; Gordon Grant; Lauren Bartlett; Alexander Black; Claudia Clavijo; Uwe Christians; Ron Betts; Doug Savage; Shih-Horng Su; John Shulze; Saibal Kar
Journal:  Circ Cardiovasc Interv       Date:  2010-04       Impact factor: 6.546

Review 4.  Polymer coatings and delayed arterial healing following drug-eluting stent implantation.

Authors:  R A Byrne; M Joner; A Kastrati
Journal:  Minerva Cardioangiol       Date:  2009-10       Impact factor: 1.347

5.  Five-year long-term clinical follow-up of the XIENCE V everolimus-eluting coronary stent system in the treatment of patients with de novo coronary artery disease: the SPIRIT II trial.

Authors:  Yoshinobu Onuma; Karine Miquel-Hebert; Patrick W Serruys
Journal:  EuroIntervention       Date:  2013-01-22       Impact factor: 6.534

6.  Abluminal biodegradable polymer biolimus-eluting stent versus durable polymer everolimus-eluting stent (COMPARE II): a randomised, controlled, non-inferiority trial.

Authors:  Pieter Cornelis Smits; Sjoerd Hofma; Mario Togni; Nicolás Vázquez; Mariano Valdés; Vassilis Voudris; Ton Slagboom; Jean-Jaques Goy; Andre Vuillomenet; Antoni Serra; Ramiro Trillo Nouche; Peter den Heijer; Martin van der Ent
Journal:  Lancet       Date:  2013-01-30       Impact factor: 79.321

7.  Long-term efficacy and safety of Zotarolimus-eluting stent in patients with diabetes mellitus: pooled 5-year results from the ENDEAVOR III and IV trials.

Authors:  Moshe Vardi; David A Burke; Sripal Bangalore; Michael J Pencina; Laura Mauri; David E Kandzari; Martin B Leon; Donald E Cutlip
Journal:  Catheter Cardiovasc Interv       Date:  2013-07-04       Impact factor: 2.692

8.  The pharmacokinetics of Biolimus A9 after elution from the Nobori stent in patients with coronary artery disease: the NOBORI PK study.

Authors:  Miodrag Ostojic; Dragan Sagic; Robert Jung; Yan-Ling Zhang; Milan Nedeljkovic; Ljupco Mangovski; Sinisa Stojkovic; Dragan Debeljacki; Mirko Colic; Branko Beleslin; Bratislav Milosavljevic; Dejan Orlic; Dragan Topic; Nevena Karanovic; Dragica Paunovic; Uwe Christians
Journal:  Catheter Cardiovasc Interv       Date:  2008-12-01       Impact factor: 2.692

9.  Short- and long-term outcomes with drug-eluting and bare-metal coronary stents: a mixed-treatment comparison analysis of 117 762 patient-years of follow-up from randomized trials.

Authors:  Sripal Bangalore; Sunil Kumar; Mario Fusaro; Nicholas Amoroso; Michael J Attubato; Frederick Feit; Deepak L Bhatt; James Slater
Journal:  Circulation       Date:  2012-05-14       Impact factor: 29.690

10.  Safety and efficacy of biolimus-eluting stent with biodegradable polymer: insights from EINSTEIN (Evaluation of Next-generation drug-eluting STEnt IN patients with coronary artery disease) Registry.

Authors:  Cristiano Freitas de Souza; Anwar Mohamed El Mouallem; Fábio Sândoli de Brito Júnior; Alexandre Antônio Cunha Abizaid; Breno Oliveira Almeida; Amanda Gonçalves Almeida; Teresa Cristina Dias Cunha Nascimento; Marco Antonio Perin; Adriano Caixeta
Journal:  Einstein (Sao Paulo)       Date:  2013 Jul-Sep
View more
  3 in total

1.  Effect of pretreatment of ezetimibe/simvastatin on arterial healing and endothelialization after drug-eluting stent implantation in a porcine coronary restenosis model.

Authors:  Doo Sun Sim; Myung Ho Jeong; Dae Sung Park; Jung Ha Kim; Kyung Seob Lim; Hyun Kuk Kim; Sung Soo Kim; Jae Yeong Cho; Hae Chang Jeong; Keun Ho Park; Young Joon Hong; Ju Han Kim; Youngkeun Ahn; Jeong Gwan Cho; Jong Chun Park
Journal:  Korean Circ J       Date:  2015-03-24       Impact factor: 3.243

2.  Effectiveness and Safety of Biolimus A9™-Eluting stEnt in Patients with AcUTe Coronary sYndrome; A Multicenter, Observational Study (BEAUTY Study).

Authors:  Keun Ho Park; Myung Ho Jeong; Young Joon Hong; Youngkeun Ahn; Hyun Kuk Kim; Young Yub Koh; Doo Il Kim; Sang Wook Kim; Weon Kim; Seung Woon Rha; Jay Young Rhew; Jong Seon Park; Hun Sik Park; Jang Ho Bae; Jang Whan Bae; Seok Kyu Oh; Sung Yun Lee; Seung Wook Lee; Jae Hwan Lee; Sang Yeob Lim; Jang Hyun Cho; Kwang Soo Cha; Jai Keon Chae; Seung Ho Hur; Sun Ho Hwang; Jin Yong Hwang
Journal:  Yonsei Med J       Date:  2018-01       Impact factor: 2.759

3.  It Is Not Mandatory to Use Triple Rather Than Dual Anti-Platelet Therapy After a Percutaneous Coronary Intervention With a Second-Generation Drug-Eluting Stent.

Authors:  Ju-Youn Kim; Yun-Seok Choi; Ami Kwon; Woo-Baek Chung; Chul-Soo Park; Hee-Yeol Kim; Kiyuk Chang; Man-Young Lee; Wook-Sung Chung; Ki-Bae Seung
Journal:  Medicine (Baltimore)       Date:  2015-11       Impact factor: 1.817

  3 in total