| Literature DB >> 24332342 |
B E Zimmerman1, L Pibida2, L E King2, D E Bergeron2, J T Cessna2, M M Mille3.
Abstract
We have developed a methodology to calibrate the (68)Ge activity concentration in large (9L) cylindrical epoxy phantoms in a way that is traceable to national standards. The method was tested on two prototype cylindrical phantoms that are being used in a clinical trial and gave (68)Ge activity concentration values with combined standard uncertainties of about 1.1%. Imaging data from the phantoms using a calibrated PET-CT scanner gave values consistent with the calibrated activity concentrations within experimental uncertainties.Keywords: Germanium-68; PET; Positron emission tomography; Standards; Traceability
Year: 2013 PMID: 24332342 DOI: 10.1016/j.apradiso.2013.11.049
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513