Literature DB >> 24305785

Mechanical behaviour of low-cost dynamic compression plates correlates with manufacturing quality standards.

Edward Aluede1, Erik McDonald, Harry Jergesen, Thomas Penoyar, Kayla Calvert.   

Abstract

PURPOSE: This study compares the mechanical properties of low-cost stainless steel dynamic compression plates (DCPs) from developing-world manufacturers, adhering to varying manufacturing quality standards, with those of high-cost DCPs manufactured for use in the developed world.
METHODS: Standard-design ten-hole DCPs from six developing-world manufacturers and high-cost DCPs from two manufacturers in the developed world were studied. Nine plates from each manufacturer underwent mechanical testing: six in four-point monotonic bending to assess strength and stiffness and three in four-point bending fatigue. Statistical comparisons of the group means of monotonic bending test data were made, and a qualitative comparison was performed to assess failures in fatigue.
RESULTS: Low-cost DCPs from manufacturers with at least one manufacturing quality standard had significantly higher bending strength and fewer failures in fatigue than did those from low-cost manufacturers with no recognised quality standards. High-cost DCPs demonstrated greater bending strength than did those in both low-cost groups. There were no differences in stiffness and fatigue failure between high-cost DCPs and those low-cost DCPs with quality standards. However, high-cost DCPs were significantly less stiff and had fewer fatigue failures than low-cost DCPs manufactured without such standards.
CONCLUSION: Significant differences were found in the mechanical properties of ten-hole DCP plates from selected manufacturers in the developing and developed worlds. These differences correlated with reported quality certification in the manufacturing process. Mechanical analysis of low-cost implants may provide information useful in determining which manufacturers produce implants with the best potential for benefit relative to cost.

Entities:  

Mesh:

Year:  2013        PMID: 24305785      PMCID: PMC3890111          DOI: 10.1007/s00264-013-2148-2

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  8 in total

1.  In vitro mechanical comparison of 2.0 and 2.4 limited-contact dynamic compression plates and 2.0 dynamic compression plates of different thicknesses.

Authors:  Adam M Strom; Tanya C Garcia; Karl Jandrey; Michael L Huber; Susan M Stover
Journal:  Vet Surg       Date:  2010-09-02       Impact factor: 1.495

2.  A preliminary study of bending stiffness alteration in shape changing nitinol plates for fracture fixation.

Authors:  Gavin Olender; Ronny Pfeifer; Christian W Müller; Thomas Gösling; Stephan Barcikowski; Christof Hurschler
Journal:  Ann Biomed Eng       Date:  2011-02-01       Impact factor: 3.934

3.  Representation of developing countries in orthopaedic journals: a survey of four influential orthopaedic journals.

Authors:  Edward E Aluede; Jonathan Phillips; Jamie Bleyer; Harry E Jergesen; Richard Coughlin
Journal:  Clin Orthop Relat Res       Date:  2012-05-16       Impact factor: 4.176

4.  Fatigue analysis of plates used for fracture stabilization in small dogs and cats.

Authors:  Scott P Hammel; G Elizabeth Pluhar; Roberto E Novo; Craig A Bourgeault; Larry J Wallace
Journal:  Vet Surg       Date:  2006-08       Impact factor: 1.495

5.  Mechanical failure in intramedullary interlocking nails.

Authors:  A K Bhat; S K Rao; K Bhaskaranand
Journal:  J Orthop Surg (Hong Kong)       Date:  2006-08       Impact factor: 1.118

6.  Fatigue fracture of bone plates.

Authors:  H Brunner; J P Simpson
Journal:  Injury       Date:  1980-02       Impact factor: 2.586

7.  Mechanical comparison of 3.5 mm broad dynamic compression plate, broad limited-contact dynamic compression plate, and narrow locking compression plate systems using interfragmentary gap models.

Authors:  Justin M Uhl; Bernard Seguin; Amy S Kapatkin; Kurt S Schulz; Tanya C Garcia; Susan M Stover
Journal:  Vet Surg       Date:  2008-10       Impact factor: 1.495

Review 8.  Failure of fracture plate fixation.

Authors:  Michael J Gardner; Jason M Evans; Robert P Dunbar
Journal:  J Am Acad Orthop Surg       Date:  2009-10       Impact factor: 3.020

  8 in total
  1 in total

1.  Comparison of conventional reconstruction plate versus direct metal laser sintering plate: an in vitro mechanical characteristics study.

Authors:  Pusheng Xie; Hanbin Ouyang; Yuping Deng; Yang Yang; Jing Xu; Wenhua Huang
Journal:  J Orthop Surg Res       Date:  2017-09-02       Impact factor: 2.359

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.