Adam M Strom1, Tanya C Garcia, Karl Jandrey, Michael L Huber, Susan M Stover.
Abstract
OBJECTIVE: To compare the bending structural stiffness and bending strength of thick and thin 2.4 mm limited contact dynamic compression plates (2.4 LC-DCP), 2.0 mm LC-DCP (2.0 LC-DCP), and 2.0 dynamic compression plates (2.0 DCP). STUDY
DESIGN: In vitro mechanical study.
METHODS: Two thicknesses of 2.4 LC-DCP, 2.0 LC-DCP, and 2.0 DCP stainless-steel plates were tested in 4-point bending. Data were collected during bending until implants plastically deformed. Bending structural stiffness and bending strength were determined from load displacement curves. Mechanical properties were compared between plates and the effects of plate type, size, and thickness on stiffness and strength were assessed using ANOVA.
RESULTS: The thick 2.4 LC-DCP implant was the stiffest and strongest; the thin 2.0 DCP implant was most compliant and weakest. Larger sized plates, thicker plates, and limited contact design of plates enhanced stiffness and strength. For the plates studied, plate size had a larger effect than plate type or thickness on stiffness and strength.
CONCLUSION: Increasing the size (width) and thickness of plates increases both the bending structural stiffness and strength. For the plates studied, LC-DCP implants were stiffer and stronger than DCP implants. CLINICAL RELEVANCE: Plate bending structural stiffness and strength can be most effectively enhanced by using a larger sized plate, but gains can also be achieved by using a thicker plate and/or an LC-DCP instead of a DCP implant when possible. © Copyright 2010 by The American College of Veterinary Surgeons.
OBJECTIVE: To compare the bending structural stiffness and bending strength of thick and thin 2.4 mm limited contact dynamic compression plates (2.4 LC-DCP), 2.0 mm LC-DCP (2.0 LC-DCP), and 2.0 dynamic compression plates (2.0 DCP). STUDY
DESIGN: In vitro mechanical study.
METHODS: Two thicknesses of 2.4 LC-DCP, 2.0 LC-DCP, and 2.0 DCP stainless-steel plates were tested in 4-point bending. Data were collected during bending until implants plastically deformed. Bending structural stiffness and bending strength were determined from load displacement curves. Mechanical properties were compared between plates and the effects of plate type, size, and thickness on stiffness and strength were assessed using ANOVA.
RESULTS: The thick 2.4 LC-DCP implant was the stiffest and strongest; the thin 2.0 DCP implant was most compliant and weakest. Larger sized plates, thicker plates, and limited contact design of plates enhanced stiffness and strength. For the plates studied, plate size had a larger effect than plate type or thickness on stiffness and strength.
CONCLUSION: Increasing the size (width) and thickness of plates increases both the bending structural stiffness and strength. For the plates studied, LC-DCP implants were stiffer and stronger than DCP implants. CLINICAL RELEVANCE: Plate bending structural stiffness and strength can be most effectively enhanced by using a larger sized plate, but gains can also be achieved by using a thicker plate and/or an LC-DCP instead of a DCP implant when possible. © Copyright 2010 by The American College of Veterinary Surgeons.
Mesh:
Substances:
Year: 2010
PMID: 20825594 DOI: 10.1111/j.1532-950X.2010.00736.x
Source DB: PubMed Journal: Vet Surg ISSN: 0161-3499 Impact factor: 1.495