Prospective real-time head motion correction using inductively coupled wireless NMR probes.

Saikat Sengupta1, Sasidhar Tadanki, John C Gore, E Brian Welch.   

Abstract

PURPOSE: Head motion continues to be a major source of artifacts and data quality degradation in MRI. The goal of this work was to develop and demonstrate a novel technique for prospective, 6 degrees of freedom (6DOF) rigid body motion estimation and real-time motion correction using inductively coupled wireless nuclear magnetic resonance (NMR) probe markers.
METHODS: Three wireless probes that are inductively coupled with the scanner's RF setup serve as fiducials on the subject's head. A 12-ms linear navigator module is interleaved with the imaging sequence for head position estimation, and scan geometry is updated in real time for motion compensation. Flip angle amplification in the markers allows the use of extremely small navigator flip angles (∼1°). A novel algorithm is presented to identify marker positions in the absence of marker specific receive channels. Motion correction is demonstrated in high resolution 2D and 3D gradient recalled echo experiments in a phantom and humans.
RESULTS: Significant improvement of image quality is demonstrated in phantoms and human volunteers under different motion conditions.
CONCLUSION: A novel real-time 6DOF head motion correction technique based on wireless NMR probes is demonstrated in high resolution imaging at 7 Tesla.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  inductively coupled probes; prospective rigid body motion correction; wireless NMR markers

Mesh:

Year:  2013        PMID: 24243810      PMCID: PMC4304771          DOI: 10.1002/mrm.25001

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  37 in total

1.  Prospective multiaxial motion correction for fMRI.

Authors:  H A Ward; S J Riederer; R C Grimm; R L Ehman; J P Felmlee; C R Jack
Journal:  Magn Reson Med       Date:  2000-03       Impact factor: 4.668

2.  Motion correction with PROPELLER MRI: application to head motion and free-breathing cardiac imaging.

Authors:  J G Pipe
Journal:  Magn Reson Med       Date:  1999-11       Impact factor: 4.668

3.  A method for fast 3D tracking using tuned fiducial markers and a limited projection reconstruction FISP (LPR-FISP) sequence.

Authors:  C Flask; D Elgort; E Wong; A Shankaranarayanan; J Lewin; M Wendt; J L Duerk
Journal:  J Magn Reson Imaging       Date:  2001-11       Impact factor: 4.813

4.  Spherical navigator echoes for full 3D rigid body motion measurement in MRI.

Authors:  Edward Brian Welch; Armando Manduca; Roger C Grimm; Heidi A Ward; Clifford R Jack
Journal:  Magn Reson Med       Date:  2002-01       Impact factor: 4.668

5.  A quantitative comparison of motion detection algorithms in fMRI.

Authors:  B A Ardekani; A H Bachman; J A Helpern
Journal:  Magn Reson Imaging       Date:  2001-09       Impact factor: 2.546

6.  Prospective acquisition correction for head motion with image-based tracking for real-time fMRI.

Authors:  S Thesen; O Heid; E Mueller; L R Schad
Journal:  Magn Reson Med       Date:  2000-09       Impact factor: 4.668

7.  Image-based tracking of optically detunable parallel resonant circuits.

Authors:  Holger Eggers; Steffen Weiss; Peter Boernert; Peter Boesiger
Journal:  Magn Reson Med       Date:  2003-06       Impact factor: 4.668

8.  An embedded optical tracking system for motion-corrected magnetic resonance imaging at 7T.

Authors:  Jessica Schulz; Thomas Siegert; Enrico Reimer; Christian Labadie; Julian Maclaren; Michael Herbst; Maxim Zaitsev; Robert Turner
Journal:  MAGMA       Date:  2012-06-13       Impact factor: 2.310

9.  Adaptive technique for high-definition MR imaging of moving structures.

Authors:  R L Ehman; J P Felmlee
Journal:  Radiology       Date:  1989-10       Impact factor: 11.105

10.  Catheter tracking and visualization using 19F nuclear magnetic resonance.

Authors:  Sebastian Kozerke; Sanjeet Hegde; Tobias Schaeffter; Rolf Lamerichs; Reza Razavi; Derek L Hill
Journal:  Magn Reson Med       Date:  2004-09       Impact factor: 4.668

View more
  14 in total

Review 1.  Motion artifacts in MRI: A complex problem with many partial solutions.

Authors:  Maxim Zaitsev; Julian Maclaren; Michael Herbst
Journal:  J Magn Reson Imaging       Date:  2015-01-28       Impact factor: 4.813

2.  MR-based motion correction for PET imaging using wired active MR microcoils in simultaneous PET-MR: phantom study.

Authors:  Chuan Huang; Jerome L Ackerman; Yoann Petibon; Thomas J Brady; Georges El Fakhri; Jinsong Ouyang
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

3.  Motion compensation for brain PET imaging using wireless MR active markers in simultaneous PET-MR: phantom and non-human primate studies.

Authors:  Chuan Huang; Jerome L Ackerman; Yoann Petibon; Marc D Normandin; Thomas J Brady; Georges El Fakhri; Jinsong Ouyang
Journal:  Neuroimage       Date:  2014-01-10       Impact factor: 6.556

Review 4.  Prospective motion correction in functional MRI.

Authors:  Maxim Zaitsev; Burak Akin; Pierre LeVan; Benjamin R Knowles
Journal:  Neuroimage       Date:  2016-11-11       Impact factor: 6.556

5.  Optical tracking with two markers for robust prospective motion correction for brain imaging.

Authors:  Aditya Singh; Benjamin Zahneisen; Brian Keating; Michael Herbst; Linda Chang; Maxim Zaitsev; Thomas Ernst
Journal:  MAGMA       Date:  2015-06-30       Impact factor: 2.310

Review 6.  Motion correction in MRI of the brain.

Authors:  F Godenschweger; U Kägebein; D Stucht; U Yarach; A Sciarra; R Yakupov; F Lüsebrink; P Schulze; O Speck
Journal:  Phys Med Biol       Date:  2016-02-11       Impact factor: 3.609

7.  MR-conditional steerable needle robot for intracerebral hemorrhage removal.

Authors:  Yue Chen; Isuru S Godage; Saikat Sengupta; Cindy Lin Liu; Kyle D Weaver; Eric J Barth
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-09-01       Impact factor: 2.924

8.  Motion-Robust Diffusion-Weighted Brain MRI Reconstruction Through Slice-Level Registration-Based Motion Tracking.

Authors:  Bahram Marami; Benoit Scherrer; Onur Afacan; Burak Erem; Simon K Warfield; Ali Gholipour
Journal:  IEEE Trans Med Imaging       Date:  2016-10       Impact factor: 10.048

Review 9.  Functional Magnetic Resonance Imaging Methods.

Authors:  Jingyuan E Chen; Gary H Glover
Journal:  Neuropsychol Rev       Date:  2015-08-07       Impact factor: 7.444

10.  Design and Fabrication of MR-Tracked Metallic Stylet for Gynecologic Brachytherapy.

Authors:  Yue Chen; Wei Wang; Ehud J Schmidt; Ka-Wai Kwok; Akila N Viswanathan; Robert Cormack; Zion Tsz Ho Tse
Journal:  IEEE ASME Trans Mechatron       Date:  2015-11-24       Impact factor: 5.303

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.