Literature DB >> 24231583

Dynamic fixation of distal femur fractures using far cortical locking screws: a prospective observational study.

Michael Bottlang1, Daniel C Fitzpatrick, Dan Sheerin, Erik Kubiak, Richard Gellman, Corey Vande Zandschulp, Josef Doornink, Kathleen Earley, Steven M Madey.   

Abstract

OBJECTIVES: Document fixation and healing of distal femur fractures stabilized by plate osteosynthesis using far cortical locking (FCL) screws.
DESIGN: Prospective and observational.
SETTING: Two level 1 and 1 level 2 trauma centers. PATIENTS: Thirty-two consecutive patients with 33 distal femur fractures (AO/OTA types 33A, 33C). INTERVENTION: Fractures were stabilized by plate osteosynthesis with MotionLoc FCL screws without supplemental bone graft or bone morphogenic proteins. Patients were followed up for a minimum of 1 year with functional and radiographic assessments obtained at postoperative weeks 6, 12, and 24 and computed tomography scans at week 12. If union was not confirmed within 1 year, follow-up was continued until union or revision surgery. MAIN OUTCOME MEASURES: The primary end point was fracture union in the absence of complications and secondary interventions. Fracture healing was defined by resolution of pain at the fracture site and cortical bridging on biplanar radiographs. Complications were defined by fixation failure, loss of reduction, implant breakage, infection, nonunion, and need for revision.
RESULTS: Thirty-one fractures were available for follow-up. None of the 125 FCL screws used for diaphyseal fixation broke or lost fixation. One of the 31 fractures displaced into varus (ΔVarus = 5.8 degrees). Thirty of the 31 fractures healed within 15.6 ± 6.2 weeks. At an average follow-up of 17 ± 4 months, there were 2 revisions: one to correct a malrotation at day 5 and one to treat a nonunion at 6 months.
CONCLUSIONS: Absence of implant and fixation failure suggests that dynamic plating of distal femur fractures with FCL screws provides safe and effective fixation.

Entities:  

Mesh:

Year:  2014        PMID: 24231583     DOI: 10.1097/01.bot.0000438368.44077.04

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  18 in total

1.  Dynamic locked plating for fixation of distal femur fractures using near- cortical over-drilling: Preliminary results of a prospective observational study.

Authors:  Sherif Galal
Journal:  J Clin Orthop Trauma       Date:  2017-07-15

2.  From Bench to Bedside: How Stiff is Too Stiff? Far-cortical Locking or Dynamic Locked Plating May Obviate the Question.

Authors:  Benjamin K Potter
Journal:  Clin Orthop Relat Res       Date:  2016-05-10       Impact factor: 4.176

3.  Incidence of nonunion after surgery of distal femoral fractures using contemporary fixation device: a meta-analysis.

Authors:  Byung-Ho Yoon; In Keun Park; Youngwoo Kim; Hyoung-Keun Oh; Suk Kyu Choo; Yerl-Bo Sung
Journal:  Arch Orthop Trauma Surg       Date:  2020-05-09       Impact factor: 3.067

4.  Dynamic Stabilization with Active Locking Plates Delivers Faster, Stronger, and More Symmetric Fracture-Healing.

Authors:  Michael Bottlang; Stanley Tsai; Emily K Bliven; Brigitte von Rechenberg; Karina Klein; Peter Augat; Julia Henschel; Daniel C Fitzpatrick; Steven M Madey
Journal:  J Bone Joint Surg Am       Date:  2016-03-16       Impact factor: 5.284

Review 5.  Internal fixation of osteoporotic fractures.

Authors:  David L Rothberg; Mark A Lee
Journal:  Curr Osteoporos Rep       Date:  2015-02       Impact factor: 5.096

6.  Lateral locked plating for distal femur fractures by low-energy trauma: what makes a difference in healing?

Authors:  Sang-Min Kim; Jae-Woo Yeom; Hyung Keun Song; Kyu-Tae Hwang; Ji-Hyo Hwang; Je-Hyun Yoo
Journal:  Int Orthop       Date:  2018-03-16       Impact factor: 3.075

7.  Dynamic Fixation of Humeral Shaft Fractures Using Active Locking Plates: A Prospective Observational Study.

Authors:  Steven M Madey; Stanley Tsai; Daniel C Fitzpatrick; Kathleen Earley; Michael Lutsch; Michael Bottlang
Journal:  Iowa Orthop J       Date:  2017

Review 8.  Locking plate fixation of distal femoral fractures is a challenging technique: a retrospective review.

Authors:  Giuseppe Toro; Giampiero Calabrò; Antonio Toro; Alessandro de Sire; Giovanni Iolascon
Journal:  Clin Cases Miner Bone Metab       Date:  2016-04-07

9.  Load-sharing through elastic micro-motion accelerates bone formation and interbody fusion.

Authors:  Eric H Ledet; Glenn P Sanders; Darryl J DiRisio; Joseph C Glennon
Journal:  Spine J       Date:  2018-02-13       Impact factor: 4.166

Review 10.  Clinical Management of Osteoporotic Fractures.

Authors:  Adam Z Khan; Richard D Rames; Anna N Miller
Journal:  Curr Osteoporos Rep       Date:  2018-06       Impact factor: 5.096