Jianzhi Cai, Haitao Huang, Yongxin Zhou, Yunqing Mei, Jie Shao, Yongwu Wang1.
Abstract
BACKGROUND: To evaluated the feasibility and safety of new aortic valved stents in transcatheter aortic valve implantation (TAVI) using retrograde approach by in vitro testing and animal implantation.Entities:
Mesh:
Year: 2013 PMID: 24219844 PMCID: PMC3842839 DOI: 10.1186/1749-8090-8-210
Source DB: PubMed Journal: J Cardiothorac Surg ISSN: 1749-8090 Impact factor: 1.637
Figure 1The animal preparation for the implantation. The healthy adult pig (A); the aortic inner diameter of experimental pig (B).
Figure 2Manufacture of the new type of aortic valved stent. The module of aortic valved stent (A); side view of the self- expandable stent (B); top view of the self- expandable stent (C); side view of the new aortic valved stent (D); top view of the new aortic valved stent (E). The old aortic valved stents in a side view (F) and a top view (G).
Figure 3The static and releasing tests of aortic valved stent in PVC tube.
Figure 4Radiography of valved stent. The radiographs of aortic valved stent in the ascending aorta (A) and during implantation (B).
The coronary flow rates pre-implantation and post-implantation
| A | Left (n = 12) | 25.38 ± 5.50 | 19.71 ± 4.80 | 5.67 ± 1.52 | 0.0000* |
| Right (n = 12) | 21.30 ± 5.41 | 17.74 ± 4.10 | 3.56 ± 1.88 | 0.0001* | |
| B | Left (n = 12) | 27.71 ± 6.28 | 27.58 ± 6.16 | 0.14 ± 1.35 | 0.7426 |
| Right (n = 12) | 22.37 ± 4.85 | 21.73 ± 4.77 | 0.64 ± 1.36 | 0.1487 |
A: the new type of aortic valved stent released within the native valve leaflets; B: the new type of aortic valved stent released in the ascending aorta. The coronary flow rates pre-implantation VS post-implantation, *P < 0.05.
Comparison of transvalvular pressure gradient and effective orifice area for the new type of valved stents implanted within the native valve leaflets or in the ascending aorta
| 1 | 0.64 ± 0.09 | 0.63 ± 0.08 | 0.9087 | 0.77 ± 0.17 | 0.77 ± 0.15 | 0.8885 |
| 2 | 0.69 ± 0.10 | 0.70 ± 0.06 | 0.7785 | 0.82 ± 0.19 | 0.81 ± 0.16 | 0.8302 |
| 3 | 0.75 ± 0.09 | 0.76 ± 0.07 | 0.8057 | 0.86 ± 0.16 | 0.87 ± 0.19 | 0.4519 |
| 4 | 0.80 ± 0.08 | 0.78 ± 0.07 | 0.0643 | 0.92 ± 0.17 | 0.91 ± 0.16 | 0.8736 |
| 5 | 0.86 ± 0.08 | 0.83 ± 0.07 | 0.2431 | 1.00 ± 0.17 | 0.99 ± 0.17 | 0.6475 |
| 6 | 0.90 ± 0.09 | 0.91 ± 0.10 | 0.5549 | 1.03 ± 0.15 | 1.02 ± 0.15 | 0.7291 |
| 7 | 0.93 ± 0.10 | 0.96 ± 0.10 | 0.5165 | 1.09 ± 0.16 | 1.08 ± 0.16 | 0.9161 |
| 8 | 1.04 ± 0.09 | 1.06 ± 0.10 | 0.6707 | 1.13 ± 0.16 | 1.12 ± 0.17 | 0.5842 |
| 9 | 1.10 ± 0.09 | 1.12 ± 0.10 | 0.6892 | 1.13 ± 0.15 | 1.12 ± 0.16 | 0.5234 |
Comparison of aortic regurgitant volume under different pressures between pre-implantation and post-implantation of new valved stents
| 40 | 576.17 ± 51.71 | 514.83 ± 52.26 | 0.0193 |
| 60 | 703.50 ± 138.68 | 653.50 ± 117.14 | 0.0296 |
| 80 | 887.17 ± 150.84 | 791.50 ± 151.66 | 0.0422 |
| 100 | 985.34 ± 145.22 | 845.20 ± 132.31 | 0.0403 |
| 120 | 1243.07 ± 142.17 | 1095.36 ± 159.37 | 0.0395 |
Comparison of the static leakage and perivalvular leakage for the old and new valved stents under different pressures
| 40 | Old | 322.00 ± 57.30 | 151.50 ± 24.52 | 170.50 ± 41.28 |
| New | 319.67 ± 59.61 | 135.83 ± 29.96 | 183.83 ± 47.62 | |
| P value | 0.9558 | 0.2943 | 0.6898 | |
| 60 | Old | 511.50 ± 48.71 | 218.83 ± 32.82 | 292.67 ± 66.27 |
| New | 476.17 ± 51.71 | 228.83 ± 38.97 | 247.33 ± 56.15 | |
| P value | 0.1190 | 0.6007 | 0.0220* | |
| 80 | Old | 700.17 ± 134.90 | 308.17 ± 84.37 | 392.00 ± 87.79 |
| New | 654.50 ± 120.09 | 306.33 ± 93.18 | 331.50 ± 77.77 | |
| P value | 0.0195* | 0.9205 | 0.0260* | |
| 100 | Old | 875.83 ± 170.19 | 358.17 ± 88.48 | 517.67 ± 117.83 |
| New | 760.17 ± 129.37 | 346.50 ± 87.20 | 413.67 ± 100.39 | |
| P value | 0.0301* | 0.2045 | 0.0373* | |
| 120 | Old | 1065.67 ± 149.71 | 399.17 ± 82.87 | 666.50 ± 114.87 |
| New | 1035.00 ± 140.59 | 404.17 ± 83.56 | 630.83 ± 101.45 | |
| P value | 0.3376 | 0.7375 | 0.3969 |
*P < 0.05, new aortic valved stents compared with the old stents.
Figure 5The post implantation morphological characteristics of aortic valved stent and the thin layer of endothelial cells on aortic valved stent was pointed out using white arrow.
Figure 6The post implantation anatomical features of three valve leaflets of aortic valved stent.