Stability of BUN and creatinine determinations on the Siemens Advia 1800 analyzer.

Jia Qin1, Huiying Wang, Anton Rets, Saul Harari, Herol Alexis, Ikram Eid, Matthew R Pincus.   

Abstract

BACKGROUND: Serum creatinine values of patients tend to change as a result of the use of different blanks used for creatinine determinations on the Advia 1650. After upgrading the analyzer to the Advia 1800, creatinine values tended to be more reproducible. As part of a quality assurance investigation to test the reproducibilities of creatinine values, we determined serial creatinine values in the sera of 13 patients whose initial values were either in the reference range or elevated (range 0.58-7.8 mg/dl). These values were determined concurrently with serum blood urea nitrogen (BUN) determinations (range 6.0-84.4 mg/dl) as these two analytes are used together in evaluation of renal function.
METHODS: We determined BUN and creatinine values, using the glutamate dehydrogenase lined enzyme assay system and the Jaffe method, respectively.
RESULTS: We find that all values for creatinine on samples stored at 4 °C were reproducible as were the corresponding BUN values, which is revealed by low values for the coefficients of variation (CVs), that is, mean CV of 4.55% for creatinine and 2.52% for BUN. One sample with relatively high CV (10.6%) for creatinine was found to have an initial value of 1.1 mg/dl, in the reference range; but, on repeat determinations, the obtained levels were as high as 1.5 mg/dl, above the reference range. BUN values for this sample remained in the reference range, suggesting that no renal disease was present.
CONCLUSION: We conclude that creatinine and BUN determinations are stable, but occasional spurious creatinine values can occur on the Advia 1800 analyzer.
© 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  Advia 1800 analyzer; BUN; coefficient of variation; creatinine; stability

Mesh:

Substances:

Year:  2013        PMID: 24218124      PMCID: PMC6807597          DOI: 10.1002/jcla.21624

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  2 in total

1.  Comparison of glucose determinations on blood samples collected in three types of tubes.

Authors:  Geling Li; Michael Cabanero; Zhenglong Wang; Huiying Wang; Tao Huang; Herol Alexis; Ikram Eid; Gilad Muth; Matthew R Pincus
Journal:  Ann Clin Lab Sci       Date:  2013       Impact factor: 1.256

2.  Stability of electrolyte determinations on the Siemens Advia 1800 analyzer.

Authors:  Rafael Garcia; Geling Li; Zhenglong Wang; Michael Cabanero; Matthew R Pincus
Journal:  J Clin Lab Anal       Date:  2012-09       Impact factor: 2.352

  2 in total
  3 in total

1.  Reproducibility of Serum Potassium Values in Serum From Blood Samples Stored for Increasing Times Prior to Centrifugation and Analysis.

Authors:  Aaron Harper; Chuanyong Lu; Yi Sun; Rafael Garcia; Anton Rets; Herol Alexis; Heba Saad; Ikram Eid; Loretta Harris; Barbara Marshall; Edlira Tafani; Matthew R Pincus
Journal:  J Clin Lab Anal       Date:  2015-04-13       Impact factor: 2.352

2.  Do Serum Creatinine Levels Show Clinically Significant Fluctuations on Serial Determinations on the Siemens Advia 1800 Analyzer?

Authors:  Daniel Levitan; Aaron E Harper; Yi Sun; Jose V Scarpa Carniello; Amir Momeni; Joshua Kagan; Herol Alexis; Ikram Eid; Loretta Harris; Barbara Marshal; Edlira Tafani; Matthew Pincus
Journal:  J Clin Lab Anal       Date:  2016-07-01       Impact factor: 2.352

3.  Can mobilization of bone marrow stem cells be an alternative regenerative therapy to stem cell injection in a rat model of chronic kidney disease?

Authors:  Shereen Morsy; Mona F Mansour; Mohamed Abdo; Yasser El-Wazir
Journal:  Physiol Rep       Date:  2022-09
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.