Stability of electrolyte determinations on the Siemens Advia 1800 analyzer.

Rafael Garcia1, Geling Li, Zhenglong Wang, Michael Cabanero, Matthew R Pincus.   

Abstract

BACKGROUND: It is sometimes necessary for the laboratory to re-test samples for critical serum electrolyte levels. It is important to assure reproducibility of results when testing is performed on stored, refrigerated samples. We have tested the reproducibility of results for the critical electrolytes, Na, K, Cl and Ca, from ten randomly selected patients'sera over our maximum storage period of nine (9) days on the Siemens Advia 1800 analyzer. The ranges for each electrolyte were 131-150 meq/L (Na), 3.4-5.2 meq/L (K), 101-123 meq/L (Cl) and 7.3-9.9 mg/dL (Ca).
METHODS: We used ion-selective electrodes for Na, K and Cl and the ortho-cresolphthalein dye method for Ca.
RESULTS: We find that the reproducibility of determinations for all of these electrolytes was excellent, i.e. the coefficients of variation for each electrolyte determination for each patient were low.
CONCLUSION: The methods of measurement for these electrolytes on the Advia 1800 are reliable and reproducible.
© 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23001983      PMCID: PMC6807598          DOI: 10.1002/jcla.21533

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  2 in total

Review 1.  Mitigation and detection of spurious potassium and sodium results.

Authors:  R A Sulaiman; P J Twomey; R Gama
Journal:  Clin Chim Acta       Date:  2010-08-25       Impact factor: 3.786

2.  Reliability of IL Monarch ion-selective electrode module for sodium, potassium, and chloride measurements.

Authors:  E Guagnellini; G Spagliardi; G Bernardi; P Stella
Journal:  Clin Chem       Date:  1988-04       Impact factor: 8.327

  2 in total
  3 in total

1.  Stability of BUN and creatinine determinations on the Siemens Advia 1800 analyzer.

Authors:  Jia Qin; Huiying Wang; Anton Rets; Saul Harari; Herol Alexis; Ikram Eid; Matthew R Pincus
Journal:  J Clin Lab Anal       Date:  2013-11       Impact factor: 2.352

2.  Reproducibility of Serum Potassium Values in Serum From Blood Samples Stored for Increasing Times Prior to Centrifugation and Analysis.

Authors:  Aaron Harper; Chuanyong Lu; Yi Sun; Rafael Garcia; Anton Rets; Herol Alexis; Heba Saad; Ikram Eid; Loretta Harris; Barbara Marshall; Edlira Tafani; Matthew R Pincus
Journal:  J Clin Lab Anal       Date:  2015-04-13       Impact factor: 2.352

3.  Do Serum Creatinine Levels Show Clinically Significant Fluctuations on Serial Determinations on the Siemens Advia 1800 Analyzer?

Authors:  Daniel Levitan; Aaron E Harper; Yi Sun; Jose V Scarpa Carniello; Amir Momeni; Joshua Kagan; Herol Alexis; Ikram Eid; Loretta Harris; Barbara Marshal; Edlira Tafani; Matthew Pincus
Journal:  J Clin Lab Anal       Date:  2016-07-01       Impact factor: 2.352

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.