Literature DB >> 24197827

Impact of changes in systemic vascular resistance on a novel non-invasive continuous cardiac output measurement system based on pulse wave transit time: a report of two cases.

Hironori Ishihara1, Masato Tsutsui.   

Abstract

The inaccuracy of arterial waveform analysis for measuring continuos cardiac output (CCO) associated with changes in systemic vascular resistance (SVR) has been well documented. A new non-invasive continuous cardiac output monitoring system (esCCO) mainly utilizing pulse wave transit time (PWTT) in place of arterial waveform analysis has been developed. However, the trending ability of esCCO to measure cardiac output during changes in SVR remains unclear. After a previous multicenter study on esCCO measurement, we retrospectively identified two cases in which apparent changes in SVR developed in a short period during data collection. In each case, the trending ability of esCCO to measure cardiac output and time component of PWTT were analyzed. Recorded data suggest that the time component of PWTT may have a significant impact on the accuracy of estimating stroke volume during changes in SVR. However, further prospective clinical studies are required to test this hypothesis.

Entities:  

Mesh:

Year:  2013        PMID: 24197827      PMCID: PMC4119244          DOI: 10.1007/s10877-013-9529-3

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  11 in total

1.  A novel continuous cardiac output monitor based on pulse wave transit time.

Authors:  Yoshihiro Sugo; Teiji Ukawa; Sunao Takeda; Hironori Ishihara; Tomiei Kazama; Junzo Takeda
Journal:  Annu Int Conf IEEE Eng Med Biol Soc       Date:  2010

2.  Pulse wave velocity as a measure of blood pressure change.

Authors:  B Gribbin; A Steptoe; P Sleight
Journal:  Psychophysiology       Date:  1976-01       Impact factor: 4.016

3.  [Comparison of continuous cardiac output measurement methods: non-invasive estimated CCO using pulse wave transit time and CCO using thermodilution].

Authors:  Masato Tsutsui; Takashige Yamada; Yoshihiro Sugo; Tetsufumi Sato; Toshimasa Akazawa; Nobukazu Sato; Koichi Yamashita; Hironori Ishihara; Tomiei Kazama; Junzo Takeda
Journal:  Masui       Date:  2012-09

4.  The impact of phenylephrine, ephedrine, and increased preload on third-generation Vigileo-FloTrac and esophageal doppler cardiac output measurements.

Authors:  Lingzhong Meng; Nam Phuong Tran; Brenton S Alexander; Kathleen Laning; Guo Chen; Zeev N Kain; Maxime Cannesson
Journal:  Anesth Analg       Date:  2011-08-04       Impact factor: 5.108

5.  The peripheral pulse wave: information overlooked.

Authors:  W B Murray; P A Foster
Journal:  J Clin Monit       Date:  1996-09

6.  Multicenter study verifying a method of noninvasive continuous cardiac output measurement using pulse wave transit time: a comparison with intermittent bolus thermodilution cardiac output.

Authors:  Takashige Yamada; Masato Tsutsui; Yoshihiro Sugo; Tetsufumi Sato; Toshimasa Akazawa; Nobukazu Sato; Koichi Yamashita; Hironori Ishihara; Junzo Takeda
Journal:  Anesth Analg       Date:  2012-03-30       Impact factor: 5.108

7.  Effect of afterload on left ventricular performance in experimental animals. Comparison of the pre-ejection period and other indices of left ventricular contractility.

Authors:  H Boudoulas; P E Karayannacos; R P Lewis; C V Leier; J S Vasko
Journal:  J Med       Date:  1982

8.  The relationship between effective arterial capacitance and pulse wave velocity is dependent on left ventricular stroke volume.

Authors:  Oladipupo Olafiranye; Ghazanfar Qureshi; Louis Salciccioli; Kinda Vernon-Jones; Charles Philip; John Kassotis; Jason M Lazar
Journal:  Angiology       Date:  2008-05-25       Impact factor: 3.619

9.  Acute haemodynamic effects of felodipine and verapamil in man, singly and with metoprolol.

Authors:  O Rönn; B Bengtsson; B Edgar; S Raner
Journal:  Drugs       Date:  1985       Impact factor: 9.546

10.  Uncalibrated arterial pulse contour analysis versus continuous thermodilution technique: effects of alterations in arterial waveform.

Authors:  Suraphong Lorsomradee; Sratwadee Lorsomradee; Stefanie Cromheecke; Stefan G De Hert
Journal:  J Cardiothorac Vasc Anesth       Date:  2007-05-07       Impact factor: 2.628

View more
  4 in total

1.  Comparison of the ability of two continuous cardiac output monitors to measure trends in cardiac output: estimated continuous cardiac output measured by modified pulse wave transit time and an arterial pulse contour-based cardiac output device.

Authors:  Takashi Terada; Ayano Oiwa; Yumi Maemura; Samuna Robert; Sayaka Kessoku; Ryoichi Ochiai
Journal:  J Clin Monit Comput       Date:  2015-09-14       Impact factor: 2.502

Review 2.  Year in review in journal of clinical monitoring and computing 2014: cardiovascular and hemodynamic monitoring.

Authors:  Karim Bendjelid; Steffen Rex; Thomas Scheeren; Lester Critchley
Journal:  J Clin Monit Comput       Date:  2015-03-05       Impact factor: 2.502

3.  Estimated continuous cardiac output based on pulse wave transit time in off-pump coronary artery bypass grafting: a comparison with transpulmonary thermodilution.

Authors:  Alexey A Smetkin; Ayyaz Hussain; Evgenia V Fot; Viktor I Zakharov; Natalia N Izotova; Angelika S Yudina; Zinaida A Dityateva; Yanina V Gromova; Vsevolod V Kuzkov; Lars J Bjertnæs; Mikhail Y Kirov
Journal:  J Clin Monit Comput       Date:  2016-03-07       Impact factor: 2.502

4.  Factors Associated with Variability in Pulse Wave Transit Time Using Pulse Oximetry: A Retrospective Study.

Authors:  Hilmanda Budiman; Ryo Wakita; Takaya Ito; Shigeru Maeda
Journal:  J Clin Med       Date:  2022-07-07       Impact factor: 4.964

  4 in total