| Literature DB >> 24176965 |
Zi Yi1, Hai-Bo Wang, Ke Chen, Qing Gao, Hao Tang, Ru-Qin Yu, Xia Chu.
Abstract
Telomerase has been considered to be an important tumor biomarker for early cancer diagnostics and a valuable target for therapy treatment. A novel electrochemical biosensor based on structure-switching DNA probe with ferrocene (Fc) as the electroactive reporter to detect telomerase activity was developed. The developed approach displayed desirable dynamic range from 10(2) to 6 × 10(4) Hela cells mL(-1) with a detection limit of 100 Hela cells mL(-1). This biosensor afforded good reproducibility, stability and simple operations. It provided a useful platform for practical use in quantitative telomerase activity assay for clinical applications. Telomerase inhibitor performance was also investigated and the results indicated the approach was suitable for telomerase inhibitor screening research.Entities:
Keywords: Electrochemical biosensor; Structure-switching DNA probe; Telomerase
Mesh:
Substances:
Year: 2013 PMID: 24176965 DOI: 10.1016/j.bios.2013.09.072
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618