Sahar Elahi1, Marc D Feldman, Jouke Dijkstra, Thomas E Milner.
Abstract
Effect of beam size and catheter position on the apparent size and apposition of metallic stent struts in IVOCT images were examined. Micro-CT data was employed to determine light - stent strut interactions. Simulated results suggest that location of the reflecting regions depend on relative orientation and position of stent struts to the IVOCT beam. Erroneous stent apposition measurements can occur when the IVOCT catheter is at an eccentric position. A method that mitigates stent strut apposition measurement errors is proposed.Keywords: (110.0110) Imaging systems; (110.4500) Optical coherence tomography
Year: 2013 PMID: 24156050 PMCID: PMC3799652 DOI: 10.1364/BOE.4.001876
Source DB: PubMed Journal: Biomed Opt Express ISSN: 2156-7085 Impact factor: 3.732