| Literature DB >> 24143147 |
Janusz Skowronek1, Grzegorz Bielęda, Piotr Laski, Witold Kycler.
Abstract
PURPOSE: The aim of the study was to verify dose distribution parameters for multi-lumen, and artificially created single-lumen balloon applicator used for the same patient with two optimization algorithms: inverse planning simulated annealing (IPSA) and dose point optimization with distance option.Entities:
Keywords: APBI; Contura; IPSA; brachytherapy; breast cancer
Year: 2013 PMID: 24143147 PMCID: PMC3797409 DOI: 10.5114/jcb.2013.37776
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1Contura® multi-lumen balloon (SenoRx, Inc., Aliso Viejo, CA, USA)
Fig. 2Anatomy reconstruction in Contura® application
Fig. 3Dose distribution visualization
Dose distribution parameters (mean values and SD) for real and virtual plans
| Multi-lumen IPSA | Multi-lumen DP | Single-lumen IPSA | Single-lumen DP | |
|---|---|---|---|---|
|
| 99.32 ± 6.75 | 83.50 ± 8.16 | 97.82 ± 7.51 | 85.52 ± 10.21 |
|
| 99.95 ± 19.61 | 88.40 ± 7.96 | 102.56 ± 7.13 | 90.27 ± 9.82 |
|
| 85.83 ± 28.59 | 81.34 ± 28.43 | 85.31 ± 32.31 | 85.07 ± 43.20 |
|
| 92.52 ± 33.86 | 84.59 ± 24.13 | 105.56 ± 33.34 | 96.05 ± 34.59 |
|
| 133.41 ± 18.96 | 118.70 ± 21.06 | 132.28 ± 21.15 | 119.99 ± 23.86 |
|
| 190.62 ± 9.63 | 193.96 ± 18.46 | 190.07 ± 12.42 | 185.05 ± 14.21 |
|
| 34.72 ± 4.84 | 24.20 ± 4.01 | 33.60 ± 5.25 | 24.45 ± 4.76 |
|
| 7.37 ± 4.25 | 8.78 ± 3.04 | 7.20 ± 4.72 | 7.19 ± 3.44 |
IPSA – Inverse Planning Simulated Annealing, DP – dose points optimization, D – dose (dose that covers % of the target volume), Dmax – maximal dose received by skin or rib, V – volume (V150, that is the percentage of the target volume that has received 50% more than the prescribed dose, V200 – 100% more)
Fig. 4Mean values and SD for doses deposited to target (D95, D50, V150) and OAR (Dmaxskin) for plans prepared using different methods
Comparison of p values: real plan with virtual plans
| Multi-lumen DP | Single-lumen IPSA | Single-lumen DP | |
|---|---|---|---|
|
| < 0.0001 | 0.0367 | < 0.0001 |
|
| 0.0145 | 0.5440 | 0.0350 |
|
| 0.0816 | 0.7131 | 0.8763 |
|
| 0.1452 | 0.0067 | 0.04467 |
|
| < 0.0001 | 0.2245 | < 0.0001 |
|
| 0.3290 | 0.7787 | 0.0333 |
|
| < 0.0001 | 0.2369 | < 0.0001 |
|
| 0.0181 | 0.6926 | 0.8001 |
IPSA – Inverse Planning Simulated Annealing, DP – dose points optimization, D – dose (dose that covers % of the target volume), Dmax – maximal dose received by skin or rib, V – volume (V150, that is the percentage of the target volume that has received 50% more than the prescribed dose, V200 – 100% more)