A comparison of skin and chest wall dose delivered with multicatheter, Contura multilumen balloon, and MammoSite breast brachytherapy.

Laurie W Cuttino1, Dorin Todor, Mihaela Rosu, Douglas W Arthur.   

Abstract

PURPOSE: Skin and chest wall doses have been correlated with toxicity in patients treated with breast brachytherapy . This investigation compared the ability to control skin and chest wall doses between patients treated with multicatheter (MC), Contura multilumen balloon (CMLB), and MammoSite (MS) brachytherapy. METHODS AND MATERIALS: 43 patients treated with the MC technique, 45 patients treated with the CMLB, and 83 patients treated with the MS were reviewed. The maximum doses delivered to the skin and chest wall were calculated for all patients.
RESULTS: The mean maximum skin doses for the MC, CMLB, and MS were 2.3 Gy (67% of prescription dose), 2.8 Gy (82% of prescription dose), and 3.2 Gy per fraction (94% of prescription dose), respectively. Although the skin distances were similar (p = 0.23) for the two balloon techniques, the mean skin dose with the CMLB was significantly lower than with the MS (p = 0.05). The mean maximum rib doses for the MC, CMLB, and MS were 2.3 Gy (67% of prescription dose), 2.8 Gy (82% of prescription dose), and 3.6 Gy per fraction (105% of prescription dose), respectively. Again, the mean rib dose with the CMLB was significantly lower than with the MS (p = 0.002).
CONCLUSION: The MC and CMLB techniques are associated with significantly lower mean skin and rib doses than is the MS. Treatment with the MS was associated with significantly more patients receiving doses to the skin or rib in excess of 125% of the prescription. Treatment with the CMLB may prove to yield less normal tissue toxicity than treatment with the MS.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20350796     DOI: 10.1016/j.ijrobp.2009.10.004

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  10 in total

1.  A Comparison of Skin Dose Delivered with MammoSite and Multicatheter Breast Brachytherapy.

Authors:  M Oshaghi; M Sadeghi; Sr Mahdavi; Ar Shirazi
Journal:  J Biomed Phys Eng       Date:  2013-12-02

Review 2.  Accelerated partial breast irradiation: advances and controversies.

Authors:  Mani Akhtari; Bin S Teh
Journal:  Chin J Cancer       Date:  2016-03-24

3.  Does ultrasound measurement improve the accuracy of electronic brachytherapy in the treatment of superficial non-melanomatous skin cancer?

Authors:  Uma Goyal; Junhan Pan; Haiyan Cui; Baldassarre Stea
Journal:  J Contemp Brachytherapy       Date:  2017-01-26

Review 4.  Dosimetric evaluation and systematic review of radiation therapy techniques for early stage node-negative breast cancer treatment.

Authors:  Tabitha Y Chan; Johann I Tang; Poh Wee Tan; Neill Roberts
Journal:  Cancer Manag Res       Date:  2018-10-23       Impact factor: 3.989

5.  Dose correction in lung for HDR breast brachytherapy.

Authors:  Eric Slessinger; Eric Pepin; Qingya Zhao; Li Zhao; Indra Das
Journal:  J Contemp Brachytherapy       Date:  2012-06-30

6.  Acute toxicity and early cosmetic outcome in patients treated with multilumen balloon brachytherapy with skin spacing ≤ 7.0 millimeters.

Authors:  Kiwhoon Lee; Amy Quillo; Dustin Dillon; Keith Sowards; Allison Hatmaker; Anthony Dragun
Journal:  J Contemp Brachytherapy       Date:  2012-03-30

7.  Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?

Authors:  Janusz Skowronek; Grzegorz Bielęda; Piotr Laski; Witold Kycler
Journal:  J Contemp Brachytherapy       Date:  2013-09-20

8.  The importance of the implant quality in APBI - Gliwice experience. Dosimetric evaluation.

Authors:  Agnieszka Cholewka; Marta Szlag; Brygida Białas; Sylwia Kellas-Ślęczka; Krzysztof Slosarek
Journal:  J Contemp Brachytherapy       Date:  2013-12-05

9.  Dose estimation for different skin models in interstitial breast brachytherapy.

Authors:  Judyta Lasota; Renata Kabacińska; Roman Makarewicz
Journal:  J Contemp Brachytherapy       Date:  2014-06-03

10.  Dosimetric evaluation of multilumen intracavitary balloon applicator rotation in high-dose-rate brachytherapy for breast cancer.

Authors:  Yongbok Kim; Mark G Trombetta
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.