Literature DB >> 16409392

Uniformity of the fluid flow velocities within hollow fiber membranes of blood oxygenation devices.

Ali R Mazaheri1, Goodarz Ahmadi.   

Abstract

A finite volume-based computational model was developed to investigate the uniformity of the fluid flow across the hollow fiber membranes in blood oxygenation devices. A two-dimensional annular cross section of a blood oxygenation device including about 3,300 hollow fiber membranes was used in the computation model. The equations governing the steady incompressible laminar flow in the blood oxygenation device were solved numerically and the results were compared with those obtained from the equivalent porous medium approximation. For the porous medium approximation, the Ergun equation was used for evaluating the permeability. The simulation results showed that the fluid molecules spend about six times longer in the fiber bundle region than that in its equivalent porous medium approximation model. The computational model also provides a more detailed fluid flow pattern in the membrane compartment of the blood oxygenator.

Mesh:

Substances:

Year:  2006        PMID: 16409392     DOI: 10.1111/j.1525-1594.2006.00150.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  4 in total

1.  Computational study of the blood flow in three types of 3D hollow fiber membrane bundles.

Authors:  Jiafeng Zhang; Xiaobing Chen; Jun Ding; Katharine H Fraser; M Ertan Taskin; Bartley P Griffith; Zhongjun J Wu
Journal:  J Biomech Eng       Date:  2013-12       Impact factor: 2.097

2.  Micro-scale Modeling of Flow and Oxygen Transfer in Hollow Fiber Membrane Bundle.

Authors:  M Ertan Taskin; Katharine H Fraser; Tao Zhang; Bartley P Griffith; Zhongjun J Wu
Journal:  J Memb Sci       Date:  2010-10-15       Impact factor: 8.742

3.  THE ROLE OF POROUS MEDIA IN MODELING FLUID FLOW WITHIN HOLLOW FIBER MEMBRANES OF THE TOTAL ARTIFICIAL LUNG.

Authors:  Khalil Khanafer; Keith Cook; Alia Marafie
Journal:  J Porous Media       Date:  2010-08-23       Impact factor: 1.663

Review 4.  Advances in extracorporeal membrane oxygenator design for artificial placenta technology.

Authors:  David G Blauvelt; Emily N Abada; Peter Oishi; Shuvo Roy
Journal:  Artif Organs       Date:  2020-11-04       Impact factor: 3.094

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.