Defining the limitations of transcutaneous bilirubin measurement in late preterm newborns.

K L Fine1, W A Carey1, J A W Schuster1, S C Bryant2, W J Cook1, B S Karon3.   

Abstract

OBJECTIVE: The objective of this study is to determine the impact of postnatal age on the bias between transcutaneous (TcB) and total serum bilirubin (TSB), and evaluate a TcB screening protocol. STUDY
DESIGN: Preterm and term infants had paired TcB and TSB performed on days 1 to 3 of life; a subset of preterm infants had measurements on days 4 to 7. Sensitivity and specificity of TcB (plotted on an age-specific TSB nomogram) for prediction of high-intermediate (HIR) or high-risk TSB were calculated.
RESULTS: Median TcB bias was 2.6 and 2.5 mg dl-1 for term and preterm infants in the first 3 days of life, respectively. However, median bias was 2.2 mg dl-1 for preterm infants at 4 to 7 days of life. TcB in preterm infants predicted HIR or high-risk TSB with 94% sensitivity and 56% specificity.
CONCLUSION: TcB screening protocols developed for term infants can be used for late preterm infants in the first 3 days of life.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28206994     DOI: 10.1038/jp.2017.8

Source DB:  PubMed          Journal:  J Perinatol        ISSN: 0743-8346            Impact factor:   2.521


  15 in total

1.  Transcutaneous bilirubin levels in late preterm neonates.

Authors:  Sotirios Fouzas; Ageliki A Karatza; Eleni Skylogianni; Lito Mantagou; Anastasia Varvarigou
Journal:  J Pediatr       Date:  2010-06-17       Impact factor: 4.406

2.  Transcutaneous bilirubinometry: effect of postnatal age.

Authors:  Y Yamauchi; I Yamanouchi
Journal:  Acta Paediatr Jpn       Date:  1991-10

3.  Estimation of bilirubin using BiliChektrade mark, a transcutaneous bilirubin measurement device: Effects of gestational age and use of phototherapy.

Authors:  Ka Jangaard; H Curtis; Rb Goldbloom
Journal:  Paediatr Child Health       Date:  2006-02       Impact factor: 2.253

Review 4.  Hyperbilirubinemia in the newborn infant > or =35 weeks' gestation: an update with clarifications.

Authors:  M Jeffrey Maisels; Vinod K Bhutani; Debra Bogen; Thomas B Newman; Ann R Stark; Jon F Watchko
Journal:  Pediatrics       Date:  2009-09-28       Impact factor: 7.124

5.  Comparison between transcutaneous bilirubinometry and total serum bilirubin measurements in preterm infants <35 weeks gestation.

Authors:  M Ahmed; S Mostafa; G Fisher; T M Reynolds
Journal:  Ann Clin Biochem       Date:  2009-11-25       Impact factor: 2.057

6.  Improved confidence intervals for the difference between binomial proportions based on paired data.

Authors:  R G Newcombe
Journal:  Stat Med       Date:  1998-11-30       Impact factor: 2.373

7.  A new transcutaneous bilirubinometer, BiliCheck, used in the neonatal intensive care unit and the maternity ward.

Authors:  F Ebbesen; L M Rasmussen; P D Wimberley
Journal:  Acta Paediatr       Date:  2002       Impact factor: 2.299

8.  Discrepancies between transcutaneous and serum bilirubin measurements.

Authors:  James A Taylor; Anthony E Burgos; Valerie Flaherman; Esther K Chung; Elizabeth A Simpson; Neera K Goyal; Isabelle Von Kohorn; Nui Dhepyasuwan
Journal:  Pediatrics       Date:  2015-01-19       Impact factor: 7.124

9.  Evaluation of transcutaneous bilirubinometry in preterm neonates.

Authors:  E T Schmidt; C A Wheeler; G L Jackson; W D Engle
Journal:  J Perinatol       Date:  2009-03-26       Impact factor: 2.521

10.  Predictive ability of a predischarge hour-specific serum bilirubin for subsequent significant hyperbilirubinemia in healthy term and near-term newborns.

Authors:  V K Bhutani; L Johnson; E M Sivieri
Journal:  Pediatrics       Date:  1999-01       Impact factor: 7.124