Literature DB >> 27763628

Reliability of transcutaneous bilirubinometry from shielded skin in neonates receiving phototherapy: a prospective cohort study.

L Murli1, A Thukral1, M J Sankar1, S Vishnubhatla2, A K Deorari1, V K Paul1, A Sakariah1, R Agarwal1.   

Abstract

OBJECTIVE: To determine the agreement between transcutaneous bilirubin (TcB) measured from shielded skin and serum total bilirubin (STB) in infants (34 to 41 weeks of gestation) with hyperbilirubinemia receiving phototherapy (PT). STUDY
DESIGN: In this prospective cohort study, we shielded a small area of skin on sternum using a commercial photo-opaque patch (BilEclipseTM, Philips Respironics, Murrysville, PA, USA). The TcB from the shielded skin (TcBs) and STB were measured at four time points-before initiation, 12 and 24 h during and once after (12 h) cessation of PT. TcB was measured using multiwavelength transcutaneous bilirubinometer (BiliChek, Philips Children's Medical Ventures, Monroeville, PA, USA). The STB was measured in triplicate by spectrophotometry (Apel BR 5100, APEL, Japan). Bland and Altman plots were drawn to determine agreement between the TcBs and STB.
RESULTS: The gestation and birth weight of enrolled neonates were 37.0 (1.0) weeks and 2750 (458) g, respectively. The age at initiation and duration of PT were 75 (27 to 312) and 25.3 (4.4) h, respectively. Bland and Altman plot showed poor agreement between TcBs and STB at all time points. The gradient (median, range) between TcBs and STB at 0, 12, 24 h and 12 h after cessation of PT were -0.2 (-4.9 to 3.5), 1.4 (-4.7 to 4.0), 1.5 (-3.8 to 9.4) and 2 (-2.9 to 5.8) mg dl-1. The proportions of TcBs values outside ±1.5 mg dl-1 of STB ranged from 47 to 64% at four time points.
CONCLUSION: TcBs does not appear to be reliable for estimating serum bilirubin in late preterm and term neonates receiving PT.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27763628     DOI: 10.1038/jp.2016.189

Source DB:  PubMed          Journal:  J Perinatol        ISSN: 0743-8346            Impact factor:   2.521


  28 in total

1.  Non-invasive estimation of total serum bilirubin by multi-wavelength spectral reflectance in neonates.

Authors:  R Lodha; A K Deorari; V Jatana; V K Paul
Journal:  Indian Pediatr       Date:  2000-07       Impact factor: 1.411

2.  Incidence, course, and prediction of hyperbilirubinemia in near-term and term newborns.

Authors:  S Umit Sarici; Muhittin A Serdar; Ayse Korkmaz; Gülsen Erdem; Olcay Oran; Gülsevin Tekinalp; Murat Yurdakök; Sule Yigit
Journal:  Pediatrics       Date:  2004-04       Impact factor: 7.124

3.  Identifying newborns at risk of significant hyperbilirubinaemia: a comparison of two recommended approaches.

Authors:  R Keren; V K Bhutani; X Luan; S Nihtianova; A Cnaan; J S Schwartz
Journal:  Arch Dis Child       Date:  2005-04       Impact factor: 3.791

4.  Limitations and opportunities of transcutaneous bilirubin measurements.

Authors:  Nienke Bosschaart; Joke H Kok; Astrid M Newsum; Dagmar M Ouweneel; Rosaline Mentink; Ton G van Leeuwen; Maurice C G Aalders
Journal:  Pediatrics       Date:  2012-03-19       Impact factor: 7.124

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Transcutaneous bilirubinometry in preterm infants receiving phototherapy.

Authors:  Shakuntala Nanjundaswamy; Anna Petrova; Rajeev Mehta; Thomas Hegyi
Journal:  Am J Perinatol       Date:  2005-04       Impact factor: 1.862

7.  Noninvasive measurement of total serum bilirubin in a multiracial predischarge newborn population to assess the risk of severe hyperbilirubinemia.

Authors:  V K Bhutani; G R Gourley; S Adler; B Kreamer; C Dalin; L H Johnson
Journal:  Pediatrics       Date:  2000-08       Impact factor: 7.124

8.  Bilirubin bound to cells does not form photoisomers.

Authors:  T Christensen; G Kinn
Journal:  Acta Paediatr       Date:  1993-01       Impact factor: 2.299

9.  Transcutaneous bilirubinometry in the management of bilirubinemia in term neonates.

Authors:  A Kumar; M M Faridi; N Singh; S H Ahmad
Journal:  Indian J Med Res       Date:  1994-05       Impact factor: 2.375

10.  Transcutaneous bilirubinometry reduces the need for blood sampling in neonates with visible jaundice.

Authors:  S Mishra; D Chawla; R Agarwal; A K Deorari; V K Paul; V K Bhutani
Journal:  Acta Paediatr       Date:  2009-10-07       Impact factor: 2.299

View more
  5 in total

1.  Comparison between Bilistick System and transcutaneous bilirubin in assessing total bilirubin serum concentration in jaundiced newborns.

Authors:  C Greco; I F Iskander; D M Akmal; S Z El Houchi; D A Khairy; G Bedogni; R P Wennberg; C Tiribelli; C D Coda Zabetta
Journal:  J Perinatol       Date:  2017-06-15       Impact factor: 2.521

2.  Efficacy of transcutaneous bilirubinometry as compared to serum bilirubin in preterm newborn during phototherapy.

Authors:  Sunil Gothwal; Neelam Singh; Sadasivam Sitaraman; Ramesh Choudhary; Kailash Kumar Meena; Ghan Shyam Bairwa; Mohan Bairwa; Amrit Jeevan
Journal:  Eur J Pediatr       Date:  2021-06-12       Impact factor: 3.183

Review 3.  Point-of-care diagnostics to improve maternal and neonatal health in low-resource settings.

Authors:  Catherine E Majors; Chelsey A Smith; Mary E Natoli; Kathryn A Kundrod; Rebecca Richards-Kortum
Journal:  Lab Chip       Date:  2017-10-11       Impact factor: 6.799

4.  Reproducibility of BiliCare™ Transcutaneus Bilirrubin Meter in Mexican Newborns.

Authors:  David Oldak; Gabriela García; Estela E Gonzalez; Enver Aillon; Juan C Falcon; Ernesto Ayala; Bernardo Oldak
Journal:  Int J Pediatr       Date:  2019-01-01

5.  Transcutaneous Bilirubin Measurements Can Be Used to Measure Bilirubin Levels during Phototherapy.

Authors:  Saad Abdullah Alsaedi
Journal:  Int J Pediatr       Date:  2018-03-20
  5 in total