Improved quantification of myocardial blood flow using highly constrained back projection reconstruction.

David Chen1, Behzad Sharif, Rohan Dharmakumar, Louise E J Thomson, C Noel Bairey Merz, Daniel S Berman, Debiao Li.   

Abstract

PURPOSE: To improve quantification of myocardial blood flow using a fast T1 mapping technique using highly constrained back projection reconstruction (HYPR)-accelerated acquisition.
METHODS: A major source of error in the measurement of myocardial blood flow (MBF) using MRI is the nonlinear relationship between image signal intensity and contrast agent (CA) concentration. HYPR-accelerated radial acquisition was used to generate pixel-wise T1 maps with a temporal resolution of one heartbeat. HYPR produces images with a temporal footprint of 40 ms and four images within 188 ms. T1 values were converted into CA concentrations by the known linear relationship between CA concentration and T1 . The T1 mapping technique was used to quantify MBF in 10 healthy subjects and compared with MBF found using image signal intensity as well as MBF reported in the literature.
RESULTS: The MBF measured using the proposed method was more consistent with that previously reported in the literature and was significantly lower (P = 0.002) than that calculated using image signal intensity (1.11 ± 0.27 mL/min/g versus 1.88 ± 0.45 mL/min/g, respectively).
CONCLUSION: We developed a fast T1 mapping method for MBF quantification using radial sampling and HYPR. Further validation is required to determine its clinical value in assessing myocardial perfusion deficit in coronary artery disease.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  MR studies; coronary artery disease; myocardial blood flow quantification; myocardial perfusion; radial acquisition

Mesh:

Substances:

Year:  2013        PMID: 24122950      PMCID: PMC3972375          DOI: 10.1002/mrm.24958

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  33 in total

1.  Improved contrast agent bolus tracking using T1 FARM.

Authors:  C A McKenzie; R S Pereira; F S Prato; Z Chen; D J Drost
Journal:  Magn Reson Med       Date:  1999-03       Impact factor: 4.668

Review 2.  Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association.

Authors:  Manuel D Cerqueira; Neil J Weissman; Vasken Dilsizian; Alice K Jacobs; Sanjiv Kaul; Warren K Laskey; Dudley J Pennell; John A Rumberger; Thomas Ryan; Mario S Verani
Journal:  J Nucl Cardiol       Date:  2002 Mar-Apr       Impact factor: 5.952

3.  A simplified explanation of the theory of indicator-dilution for measurement of fluid flow and volume and other distributive phenomena.

Authors:  K L ZIERLER
Journal:  Bull Johns Hopkins Hosp       Date:  1958-10

4.  Absolute myocardial perfusion in canines measured by using dual-bolus first-pass MR imaging.

Authors:  Timothy F Christian; Dan W Rettmann; Anthony H Aletras; Steve L Liao; Joni L Taylor; Robert S Balaban; Andrew E Arai
Journal:  Radiology       Date:  2004-07-29       Impact factor: 11.105

5.  Accurate assessment of the arterial input function during high-dose myocardial perfusion cardiovascular magnetic resonance.

Authors:  Peter D Gatehouse; Andrew G Elkington; Nicholas A Ablitt; Guang-Zhong Yang; Dudley J Pennell; David N Firmin
Journal:  J Magn Reson Imaging       Date:  2004-07       Impact factor: 4.813

6.  Highly constrained backprojection for time-resolved MRI.

Authors:  C A Mistretta; O Wieben; J Velikina; W Block; J Perry; Y Wu; K Johnson; Y Wu
Journal:  Magn Reson Med       Date:  2006-01       Impact factor: 4.668

7.  Signal strength in subsecond FLASH magnetic resonance imaging: the dynamic approach to steady state.

Authors:  W Hänicke; K D Merboldt; D Chien; M L Gyngell; H Bruhn; J Frahm
Journal:  Med Phys       Date:  1990 Nov-Dec       Impact factor: 4.071

8.  Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart.

Authors:  Daniel R Messroghli; Aleksandra Radjenovic; Sebastian Kozerke; David M Higgins; Mohan U Sivananthan; John P Ridgway
Journal:  Magn Reson Med       Date:  2004-07       Impact factor: 4.668

9.  Reduced myocardial perfusion reserve and transmural perfusion gradient in heart transplant arteriopathy assessed by magnetic resonance imaging.

Authors:  Olaf M Muehling; Norbert M Wilke; Prasad Panse; Michael Jerosch-Herold; Betsy V Wilson; Robert F Wilson; Leslie W Miller
Journal:  J Am Coll Cardiol       Date:  2003-09-17       Impact factor: 24.094

10.  Towards elimination of the dark-rim artifact in first-pass myocardial perfusion MRI: removing Gibbs ringing effects using optimized radial imaging.

Authors:  Behzad Sharif; Rohan Dharmakumar; Troy LaBounty; Reza Arsanjani; Chrisandra Shufelt; Louise Thomson; C Noel Bairey Merz; Daniel S Berman; Debiao Li
Journal:  Magn Reson Med       Date:  2013-09-12       Impact factor: 4.668

View more
  3 in total

1.  Quantification of myocardial blood flow using non-electrocardiogram-triggered MRI with three-slice coverage.

Authors:  David Chen; Behzad Sharif; Xiaoming Bi; Janet Wei; Louise E J Thomson; C Noel Bairey Merz; Daniel S Berman; Debiao Li
Journal:  Magn Reson Med       Date:  2015-06-08       Impact factor: 4.668

2.  All-systolic non-ECG-gated myocardial perfusion MRI: Feasibility of multi-slice continuous first-pass imaging.

Authors:  Behzad Sharif; Reza Arsanjani; Rohan Dharmakumar; C Noel Bairey Merz; Daniel S Berman; Debiao Li
Journal:  Magn Reson Med       Date:  2015-06-06       Impact factor: 4.668

3.  Quantification of myocardial blood flow using non-ECG-triggered MR imaging.

Authors:  David Chen; Behzad Sharif; Rohan Dharmakumar; Louise E J Thomson; C Noel Bairey Merz; Daniel S Berman; Debiao Li
Journal:  Magn Reson Med       Date:  2014-09-16       Impact factor: 4.668

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.