Literature DB >> 24091416

Extracortical plate fixation with new plate inserts and cerclage wires for the treatment of periprosthetic hip fractures.

Johannes D Bastian1, Andre Butscher, Gianni Bigolin, Matthias A Zumstein, Hubert P Nötzli.   

Abstract

PURPOSE: Fixation of periprosthetic hip fractures with intracortical anchorage might not be feasible in cases with bulky implants and/or poor bone stock.
METHODS: Rotational stability of new plate inserts with extracortical anchorage for cerclage fixation was measured and compared to the stability found using a standard technique in a biomechanical setup using a torsion testing machine. In a synthetic PUR bone model, transverse fractures were fixed distally using screws and proximally by wire cerclages attached to the plates using "new" (extracortical anchorage) or "standard" (intracortical anchorage) plate inserts. Time to fracture consolidation and complications were assessed in a consecutive series of 18 patients (18 female; mean age 81 years, range 55-92) with periprosthetic hip fractures (ten type B1, eight type C-Vancouver) treated with the new device between July 2003 and July 2010.
RESULTS: The "new" device showed a higher rotational stability than the "standard" technique (p < 0.001). Fractures showed radiographic consolidation after 14 ± 5 weeks (mean ± SD) postoperatively in patients. Revision surgery was necessary in four patients, unrelated to the new technique.
CONCLUSION: In periprosthetic hip fractures in which fixation with intracortical anchorage using conventional means might be difficult due to bulky revision stems and/or poor bone stock, the new device may be an addition to the range of existing implants.

Entities:  

Mesh:

Year:  2013        PMID: 24091416      PMCID: PMC3936090          DOI: 10.1007/s00264-013-2114-z

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  19 in total

1.  Mennen plate fixation for the treatment of periprosthetic femoral fractures: a multicenter study of thirty-six fractures.

Authors:  Robbert J P Noorda; Paul I J M Wuisman
Journal:  J Bone Joint Surg Am       Date:  2002-12       Impact factor: 5.284

2.  Surgical management of intra- and postoperative fractures of the femur about the tip of the stem in total hip arthroplasty.

Authors:  H Schmotzer; G H Tchejeyan; D M Dall
Journal:  J Arthroplasty       Date:  1996-09       Impact factor: 4.757

Review 3.  Periprosthetic fracture of the femur after total hip arthroplasty: treatment and results to date.

Authors:  D G Lewallen; D J Berry
Journal:  Instr Course Lect       Date:  1998

4.  A biomechanical study of three wiring techniques for cerclage-plating.

Authors:  S S Stevens; A J Irish; J G Vachtsevanos; J Csongradi; G S Beaupré
Journal:  J Orthop Trauma       Date:  1995       Impact factor: 2.512

Review 5.  Epidemiology: hip and knee.

Authors:  D J Berry
Journal:  Orthop Clin North Am       Date:  1999-04       Impact factor: 2.472

Review 6.  Classification of the hip.

Authors:  O H Brady; D S Garbuz; B A Masri; C P Duncan
Journal:  Orthop Clin North Am       Date:  1999-04       Impact factor: 2.472

7.  Periprosthetic fractures of the femur. An analysis of 93 fractures.

Authors:  R K Beals; S S Tower
Journal:  Clin Orthop Relat Res       Date:  1996-06       Impact factor: 4.176

Review 8.  Periprosthetic fractures evaluation and treatment.

Authors:  Bassam A Masri; R M Dominic Meek; Clive P Duncan
Journal:  Clin Orthop Relat Res       Date:  2004-03       Impact factor: 4.176

9.  Fixation of periprosthetic femoral shaft fractures associated with cemented femoral stems: a biomechanical comparison of locked plating and conventional cable plates.

Authors:  Eric Fulkerson; Kenneth Koval; Charles F Preston; Kazuho Iesaka; Frederick J Kummer; Kenneth A Egol
Journal:  J Orthop Trauma       Date:  2006-02       Impact factor: 2.512

10.  Dall-Miles plates for periprosthetic femoral fractures. A critical review of 16 cases.

Authors:  Eleftherios Tsiridis; Fares S Haddad; Graham A Gie
Journal:  Injury       Date:  2003-02       Impact factor: 2.586

View more
  3 in total

1.  High incidence of stem loosening in association with periprosthetic femur fractures in previously well-fixed cementless grit-blasted tapered-wedge stems.

Authors:  Seung-Jae Lim; Kyung-Jae Lee; Byung-Woo Min; Joo-Hyoun Song; Sang-Yeon So; Youn-Soo Park
Journal:  Int Orthop       Date:  2014-11-11       Impact factor: 3.075

Review 2.  The management of type B1 periprosthetic femoral fractures: when to fix and when to revise.

Authors:  Adam T Yasen; Fares S Haddad
Journal:  Int Orthop       Date:  2014-12-16       Impact factor: 3.075

3.  Clinical and radiological results of patients treated with orthogonal double plating for periprosthetic femoral fractures.

Authors:  Franz Josef Müller; Michael Galler; Bernd Füchtmeier
Journal:  Int Orthop       Date:  2014-08-12       Impact factor: 3.075

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.