Literature DB >> 24070691

Oxinium femoral head damage generated by a metallic foreign body within the polyethylene cup following recurrent dislocation episodes.

E Gibon1, C Scemama, B David, M Hamadouche.   

Abstract

Oxinium femoral heads are supposed to be more scratch-resistant thanks to their oxidized layer. However, damages to this thin layer can jeopardize implant's properties. Following revision total hip arthroplasty performed for recurrent posterior dislocations, the Oxinium femoral head initially implanted was observed to be dramatically damaged. A metallic foreign body from a trochanteric fixation wire was found within the polyethylene cup. Only few cases of damaged Oxinium femoral heads have been reported and all were related to either dislocation or reduction of THA. The aim of this report is to describe a non-reported mechanism of damaged Oxinium femoral head due to a broken trochanteric fixation wire device. Any broken metallic wire from a transtrochanteric approach should be carefully followed to detect migration within the polyethylene cup. If such a migration occurs, revision surgery should be rapidly scheduled.
Copyright © 2013 Elsevier Masson SAS. All rights reserved.

Entities:  

Keywords:  Dislocation; Oxinium femoral head; Total hip arthroplasty; Transtrochanteric approach

Mesh:

Substances:

Year:  2013        PMID: 24070691     DOI: 10.1016/j.otsr.2013.04.011

Source DB:  PubMed          Journal:  Orthop Traumatol Surg Res        ISSN: 1877-0568            Impact factor:   2.256


  4 in total

1.  Do oxidized zirconium femoral heads reduce polyethylene wear in cemented THAs? A blinded randomized clinical trial.

Authors:  Amine Zaoui; Samer El Hage; Jean Langlois; Caroline Scemama; Jean Pierre Courpied; Moussa Hamadouche
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

2.  Distinctive damage patterns on THA metal bearing surfaces: case studies.

Authors:  Anneliese D Heiner; Nishant M Tikekar; Karen M Kruger; John J Lannutti; Thomas D Brown
Journal:  Iowa Orthop J       Date:  2014

3.  Low wear rates seen in THAs with highly crosslinked polyethylene at 9 to 14 years in patients younger than age 50 years.

Authors:  Kevin L Garvin; Tyler C White; Anand Dusad; Curtis W Hartman; John Martell
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

4.  Hip dislocation increases roughness of oxidized zirconium femoral heads in total hip arthroplasty: an analysis of 59 retrievals.

Authors:  Mohamed E Moussa; Christina I Esposito; Marcella E Elpers; Timothy M Wright; Douglas E Padgett
Journal:  J Arthroplasty       Date:  2014-11-11       Impact factor: 4.757

  4 in total