Literature DB >> 24061715

Navigated total knee arthroplasty: is it error-free?

Kerk Hsiang Zackary Chua1, Yongsheng Chen, Krishna Lingaraj.   

Abstract

PURPOSE: The aim of this study was to determine whether errors do occur in navigated total knee arthroplasty (TKAs) and to study whether errors in bone resection or implantation contribute to these errors.
METHODS: A series of 20 TKAs was studied using computer navigation. The coronal and sagittal alignments of the femoral and tibial cutting guides, the coronal and sagittal alignments of the final tibial implant and the coronal alignment of the final femoral implant were compared with that of the respective bone resections. To determine the post-implantation mechanical alignment of the limb, the coronal alignment of the femoral and tibial implants was combined.
RESULTS: The median deviation between the femoral cutting guide and bone resection was 0° (range -0.5° to +0.5°) in the coronal plane and 1.0° (range -2.0° to +1.0°) in the sagittal plane. The median deviation between the tibial cutting guide and bone resection was 0.5° (range -1.0° to +1.5°) in the coronal plane and 1.0° (range -1.0° to +3.5°) in the sagittal plane. The median deviation between the femoral bone resection and the final implant was 0.25° (range -2.0° to 3.0°) in the coronal plane. The median deviation between the tibial bone resection and the final implant was 0.75° (range -3.0° to +1.5°) in the coronal plane and 1.75° (range -4.0° to +2.0°) in the sagittal plane. The median post-implantation mechanical alignment of the limb was 0.25° (range -3.0° to +2.0°).
CONCLUSIONS: When navigation is used only to guide the positioning of the cutting jig, errors may arise in the manual, non-navigated steps of the procedure. Our study showed increased cutting errors in the sagittal plane for both the femur and the tibia, and following implantation, the greatest error was seen in the sagittal alignment of the tibial component. Computer navigation should be used not only to guide the positioning of the cutting jig, but also to check the bone resection and implant position during TKA. LEVEL OF EVIDENCE: IV.

Mesh:

Year:  2013        PMID: 24061715     DOI: 10.1007/s00167-013-2641-6

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


  32 in total

1.  Bone cutting errors in total knee arthroplasty.

Authors:  Christopher Plaskos; Antony J Hodgson; Kevin Inkpen; Robert W McGraw
Journal:  J Arthroplasty       Date:  2002-09       Impact factor: 4.757

2.  Alignment in total knee arthroplasty. A comparison of computer-assisted surgery with the conventional technique.

Authors:  H Bäthis; L Perlick; M Tingart; C Lüring; D Zurakowski; J Grifka
Journal:  J Bone Joint Surg Br       Date:  2004-07

3.  Intraoperative cutting errors in total knee arthroplasty.

Authors:  H Bäthis; L Perlick; M Tingart; C Perlick; C Lüring; J Grifka
Journal:  Arch Orthop Trauma Surg       Date:  2004-11-09       Impact factor: 3.067

4.  Computer assisted navigation in total knee arthroplasty: comparison with conventional methods.

Authors:  Kevin C Anderson; Knute C Buehler; David C Markel
Journal:  J Arthroplasty       Date:  2005-10       Impact factor: 4.757

5.  Alignments and clinical results in conventional and navigated total knee arthroplasty.

Authors:  A Ensini; F Catani; A Leardini; M Romagnoli; S Giannini
Journal:  Clin Orthop Relat Res       Date:  2007-04       Impact factor: 4.176

6.  Sources of error in total knee arthroplasty.

Authors:  Loukas Koyonos; S David Stulberg; Todd C Moen; Gina Bart; Michael Granieri
Journal:  Orthopedics       Date:  2009-05       Impact factor: 1.390

7.  Alignment deviation between bone resection and final implant positioning in computer-navigated total knee arthroplasty.

Authors:  Fabio Catani; Nicola Biasca; Andrea Ensini; Alberto Leardini; Luca Bianchi; Vitantonio Digennaro; Sandro Giannini
Journal:  J Bone Joint Surg Am       Date:  2008-04       Impact factor: 5.284

8.  Cutting errors in total knee replacement: assessment by computer assisted surgery.

Authors:  W P Yau; K Y Chiu
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2008-05-14       Impact factor: 4.342

9.  Relationship between cutting errors and learning curve in computer-assisted total knee replacement.

Authors:  Alfonso Manzotti; Pietro Cerveri; Elena De Momi; Chris Pullen; Norberto Confalonieri
Journal:  Int Orthop       Date:  2009-06-10       Impact factor: 3.075

10.  Cutting and implanting errors in minimally invasive total knee arthroplasty using a navigation system.

Authors:  Masahiro Hasegawa; Kakunoshin Yoshida; Hiroki Wakabayashi; Akihiro Sudo
Journal:  Int Orthop       Date:  2012-10-28       Impact factor: 3.075

View more
  9 in total

1.  Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions.

Authors:  Carmen Zietz; Joern Reinders; Jens Schwiesau; Alexander Paulus; Jan Philippe Kretzer; Thomas Grupp; Sandra Utzschneider; Rainer Bader
Journal:  J Mater Sci Mater Med       Date:  2015-02-26       Impact factor: 3.896

2.  Case-related factors affecting cutting errors of the proximal tibia in total knee arthroplasty assessed by computer navigation.

Authors:  Tadashi Tsukeoka; Yoshikazu Tsuneizumi; Kensuke Yoshino; Mashiko Suzuki
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-12-22       Impact factor: 4.342

3.  A modified technique to reduce tibial keel cutting errors during an Oxford unicompartmental knee arthroplasty.

Authors:  Hiroshi Inui; Shuji Taketomi; Keitarou Tahara; Ryota Yamagami; Takaki Sanada; Sakae Tanaka
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-05-02       Impact factor: 4.342

4.  Does final component alignment correlate with alignment of the bone resection surfaces in cemented total knee arthroplasty?

Authors:  Seung-Suk Seo; Chang-Wan Kim; Chang-Rack Lee; Jin-Hyuk Seo; Do-Hun Kim; Ok-Gul Kim; Young-Kyoung Min
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-11-08       Impact factor: 4.342

5.  Preserving the PCL during the tibial cut in total knee arthroplasty.

Authors:  G Cinotti; P Sessa; M Amato; F R Ripani; G Giannicola
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-11-04       Impact factor: 4.342

6.  Valgus position of the femoral component causes abnormal kinematics in the presence of medial looseness in total knee arthroplasty: a computer simulation model of TKA for valgus knee osteoarthritis.

Authors:  Kohei Nishitani; Shinichi Kuriyama; Shinichiro Nakamura; Yugo Morita; Hiromu Ito; Shuichi Matsuda
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2018-10-29       Impact factor: 4.342

7.  Articular surface mounted navigated total knee arthroplasty improves the reliability of component alignment.

Authors:  N D Clement; D MacDonald; A G Burgess; C R Howie
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-02-16       Impact factor: 4.342

8.  The use of imageless navigation to quantify cutting error in total knee arthroplasty.

Authors:  Ran Schwarzkopf; Morteza Meftah; Scott E Marwin; Michelle A Zabat; Jeffrey M Muir; Iain R Lamb
Journal:  Knee Surg Relat Res       Date:  2021-12-04

9.  Notching is less, if femoral component sagittal positioning is planned perpendicular to distal femur anterior cortex axis, in navigated TKA.

Authors:  Raj Kanna; Chandramohan Ravichandran; Gautam M Shetty
Journal:  Knee Surg Relat Res       Date:  2021-12-24
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.