Katherine S Green1, Doletha M E Szebenyi, Kasey Boggs, Richard Bredthauer, Mark W Tate, Sol M Gruner.
Abstract
The fabrication and testing of a prototype deep-depletion direct-conversion X-ray CCD detector are described. The device is fabricated on 600 µm-thick high-resistivity silicon, with 24 × 24 µm pixels in a 4k × 4k pixel format. Calibration measurements and the results of initial protein crystallography experiments at the Cornell High Energy Synchrotron Source (CHESS) F1 beamline are described, as well as suggested improvements for future versions of the detector.Entities:
Keywords: CCD area detectors; instrumentation; protein crystallography
Year: 2013 PMID: 24046505 PMCID: PMC3769057 DOI: 10.1107/S0021889813016592
Source DB: PubMed Journal: J Appl Crystallogr ISSN: 0021-8898 Impact factor: 3.304