Literature DB >> 21335901

X-ray analog pixel array detector for single synchrotron bunch time-resolved imaging.

Lucas J Koerner1, Sol M Gruner.   

Abstract

Dynamic X-ray studies can reach temporal resolutions limited by only the X-ray pulse duration if the detector is fast enough to segregate synchrotron pulses. An analog integrating pixel array detector with in-pixel storage and temporal resolution of around 150 ns, sufficient to isolate pulses, is presented. Analog integration minimizes count-rate limitations and in-pixel storage captures successive pulses. Fundamental tests of noise and linearity as well as high-speed laser measurements are shown. The detector resolved individual bunch trains at the Cornell High Energy Synchrotron Source at levels of up to 3.7 × 10(3) X-rays per pixel per train. When applied to turn-by-turn X-ray beam characterization, single-shot intensity measurements were made with a repeatability of 0.4% and horizontal oscillations of the positron cloud were detected.

Entities:  

Mesh:

Year:  2010        PMID: 21335901      PMCID: PMC3042326          DOI: 10.1107/S090904951004104X

Source DB:  PubMed          Journal:  J Synchrotron Radiat        ISSN: 0909-0495            Impact factor:   2.616


  5 in total

1.  Time-resolved x-ray scattering from laser-molten indium antimonide.

Authors:  R Nüske; C v Korff Schmising; A Jurgilaitis; H Enquist; H Navirian; P Sondhauss; J Larsson
Journal:  Rev Sci Instrum       Date:  2010-01       Impact factor: 1.523

2.  X-ray synchrotron studies of ultrafast crystalline dynamics.

Authors:  Matthew F DeCamp; David A Reis; David M Fritz; Philip H Bucksbaum; Eric M Dufresne; Roy Clarke
Journal:  J Synchrotron Radiat       Date:  2005-02-22       Impact factor: 2.616

3.  Analog pixel array detectors.

Authors:  A Ercan; M W Tate; S M Gruner
Journal:  J Synchrotron Radiat       Date:  2006-02-17       Impact factor: 2.616

4.  Chopper system for time resolved experiments with synchrotron radiation.

Authors:  Marco Cammarata; Laurent Eybert; Friederike Ewald; Wolfgang Reichenbach; Michael Wulff; Philip Anfinrud; Friedrich Schotte; Anton Plech; Qingyu Kong; Maciej Lorenc; Bernd Lindenau; Jürgen Räbiger; Stephan Polachowski
Journal:  Rev Sci Instrum       Date:  2009-01       Impact factor: 1.523

5.  Picosecond time-resolved laser pump/X-ray probe experiments using a gated single-photon-counting area detector.

Authors:  T Ejdrup; H T Lemke; K Haldrup; T N Nielsen; D A Arms; D A Walko; A Miceli; E C Landahl; E M Dufresne; M M Nielsen
Journal:  J Synchrotron Radiat       Date:  2009-03-20       Impact factor: 2.616

  5 in total
  7 in total

1.  Global radiation damage at 300 and 260 K with dose rates approaching 1 MGy s⁻¹.

Authors:  Matthew Warkentin; Ryan Badeau; Jesse B Hopkins; Anne M Mulichak; Lisa J Keefe; Robert E Thorne
Journal:  Acta Crystallogr D Biol Crystallogr       Date:  2012-01-17

2.  A prototype direct-detection CCD for protein crystallography.

Authors:  Katherine S Green; Doletha M E Szebenyi; Kasey Boggs; Richard Bredthauer; Mark W Tate; Sol M Gruner
Journal:  J Appl Crystallogr       Date:  2013-07-18       Impact factor: 3.304

Review 3.  X-ray imaging detectors for synchrotron and XFEL sources.

Authors:  Takaki Hatsui; Heinz Graafsma
Journal:  IUCrJ       Date:  2015-04-10       Impact factor: 4.769

4.  Determination of crystallographic intensities from sparse data.

Authors:  Kartik Ayyer; Hugh T Philipp; Mark W Tate; Jennifer L Wierman; Veit Elser; Sol M Gruner
Journal:  IUCrJ       Date:  2015-01-01       Impact factor: 4.769

5.  High-speed X-ray imaging pixel array detector for synchrotron bunch isolation.

Authors:  Hugh T Philipp; Mark W Tate; Prafull Purohit; Katherine S Shanks; Joel T Weiss; Sol M Gruner
Journal:  J Synchrotron Radiat       Date:  2016-01-28       Impact factor: 2.616

6.  Pulse-resolved multi-photon X-ray detection at 31 MHz based on a quadrant avalanche photodiode.

Authors:  Tobias Reusch; Markus Osterhoff; Johannes Agricola; Tim Salditt
Journal:  J Synchrotron Radiat       Date:  2014-06-03       Impact factor: 2.616

7.  Evaluating scintillator performance in time-resolved hard X-ray studies at synchrotron light sources.

Authors:  Michael E Rutherford; David J Chapman; Thomas G White; Michael Drakopoulos; Alexander Rack; Daniel E Eakins
Journal:  J Synchrotron Radiat       Date:  2016-03-24       Impact factor: 2.616