Literature DB >> 24036868

Application of failure mode and effects analysis (FMEA) to pretreatment phases in tomotherapy.

Sara Broggi1, Marie Claire Cantone, Anna Chiara, Nadia Di Muzio, Barbara Longobardi, Paola Mangili, Ivan Veronese.   

Abstract

The aim of this paper was the application of the failure mode and effects analysis (FMEA) approach to assess the risks for patients undergoing radiotherapy treatments performed by means of a helical tomotherapy unit. FMEA was applied to the preplanning imaging, volume determination, and treatment planning stages of the tomotherapy process and consisted of three steps: 1) identification of the involved subprocesses; 2) identification and ranking of the potential failure modes, together with their causes and effects, using the risk probability number (RPN) scoring system; and 3) identification of additional safety measures to be proposed for process quality and safety improvement. RPN upper threshold for little concern of risk was set at 125. A total of 74 failure modes were identified: 38 in the stage of preplanning imaging and volume determination, and 36 in the stage of planning. The threshold of 125 for RPN was exceeded in four cases: one case only in the phase of preplanning imaging and volume determination, and three cases in the stage of planning. The most critical failures appeared related to (i) the wrong or missing definition and contouring of the overlapping regions, (ii) the wrong assignment of the overlap priority to each anatomical structure, (iii) the wrong choice of the computed tomography calibration curve for dose calculation, and (iv) the wrong (or not performed) choice of the number of fractions in the planning station. On the basis of these findings, in addition to the safety strategies already adopted in the clinical practice, novel solutions have been proposed for mitigating the risk of these failures and to increase patient safety.

Entities:  

Mesh:

Year:  2013        PMID: 24036868      PMCID: PMC5714567          DOI: 10.1120/jacmp.v14i5.4329

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


I. INTRODUCTION

The benefits of ionizing radiation in medicine are well accepted even though the risks, coupled with their use, cannot be entirely eliminated. While the diagnostic use of radiation requires suitable methodologies to minimize the dose without impairing the diagnostic quality, the optimization in radiotherapy must be achieved by maintaining sufficiently high doses to irradiated tumors and protecting, at the same time, the healthy tissues to the largest extent possible. In modern radiotherapy (RT), much effort is being invested to improve the conformity of dose distribution, as well as to integrate imaging techniques for tumor tracking and correction of inter‐ and intrafraction variations.( , ) Intensity‐modulated radiation therapy (IMRT) is becoming the standard technique for achieving highly conformal irradiation volumes in many RT treatments through the intersection of numerous beamlets. Most commercial IMRT systems evolved from conventional linear accelerators (linacs) equipped with multileaf collimators (MLCs). A specifically designed IMRT machine, which also integrates a highly image‐guided system, is the helical tomotherapy. Indeed, a tomotherapy accelerator basically combines the main features of a linear accelerator and a megavoltage computed tomography (MVCT) scanner. A common aspect of all the new technologies and methodologies introduced in the modern RT is the level of complexity, evidently much higher than the recent past (i.e., before IMRT became a major radiotherapy treatment modality). The increased complexity related to the technological and process changes in RT places new demands on quality assurance (QA) programs, as well as innovative instruments and detectors for beam characterization and checks.( , , ) Moreover, new approaches to safety are required, since complexity may also increase the sensitivity to uncertainties and risk for accidental exposures. Examples of radiotherapy‐related errors are unfortunately not uncommon, even in the countries with the highest level of health‐care resources. In order to fully assess and manage the risks of accidental exposures deriving from the use of innovative radiotherapy methodologies, prospective approaches, widely applied in high‐risk industry, should be implemented to determine all the elements that could go wrong and identify, a priori, all the potential hazards that might occur during a radiotherapy treatment. Prospective methods for risk analysis and patient safety improvement have been recently applied in various modern radiotherapy methodologies. , , , , , , However, as far as authors know, no studies specifically dedicated to tomotherapy are available in the literature. The aim of this paper was the application of the failure mode and effects analysis (FMEA) approach to assess the risks for patients during the pretreatment phases in tomotherapy. The applied procedure included the definition of the processes and fault trees, the identification and scoring of each potential failure mode, and finally the suggestion of additional safety measures for process improvement and risk mitigation.

II. MATERIALS AND METHODS

FMEA is a prospective risk analysis approach routinely employed in several manufacturing sectors, as well as in aviation. Recently FMEA was identified as a powerful tool in modern radiation oncology by the Task Group 100 of the American Association of Physicists in Medicine (AAPM). The use of the FMEA approach was also recommended by the International Commission on Radiological Protection (ICRP) as a resource for improving the safety of patients undergoing modern radiation therapy treatments.1(10) In this study, FMEA was applied to identify all the subprocesses involved in the stages of (i) preplanning imaging and volume determination, and of (ii) treatment planning, characterizing a RT process performed by using a helical tomotherapy unit (HTU). Afterwards, the potential failure modes (i.e., what could go wrong), together with their causes and effects, were identified and ranked in order of importance. Three indexes were assigned for each failure mode: the occurrence rating (O), the severity rating (S), and the detectability rating (D). A ten‐point scale was used to score each category, ten being the number indicating the most severe, most frequent, and least detectable failure mode, respectively. The ranking scales proposed by Ford et al. were adopted as guidelines. Finally, the risk probability number (RPN) was calculated as the product of the three attributes: . As already applied in previous FMEA studies in the value was considered as a threshold below which the risk can be considered acceptable. However, it must be pointed out that this value, derived from industry, still remains somehow arbitrary when applied to RT and deserves further investigation. The analysis was carried out by a working group (WG) composed by five people working at the San Raffaele Scientific Institute (three medical physicists and two radiation oncologists), and by two additional external physicists with experience and competences in radiation protection and in risk management strategies for radiotherapy. The delineation of the process trees and the identification of the potential failure modes, and causes and effects, were initially carried out through small group meetings. Afterwards, various collegial discussions were organized to revise the results the groups, and to identify and examine the additional safety measures for the risk mitigation. The risk attributes associated to each failure mode were initially conceived by each member of the WG in “blind mode”, then collectively revised during a dedicated plenary session to reach general consensus. The O, S, and D indexes for each failure mode were assigned by taking into account the current QA program and protocols developed at San Raffaele Scientific Institute and the safety measures already implemented. Indeed, San Raffaele Scientific Institute was assumed as a reference for the detailed definition of the process tree and the estimation of RPN numbers. The Institute is equipped with three HTUs: the first one (TT1, Hi‐Art II) installed in August 2004, a second (TT2, Hi‐Art II) installed in May 2007, and a third one (TT3, Tomotherapy HD) installed in March 2012 (TomoTherapy Inc., Madison, WI). Currently, only the two newer HTUs are operating; the first unit is at the moment off and will be replaced with a new one. The two operating HTUs (TT2 and TT3) are twinned each other; the same beam model is applied and each unit is characterized by a proper MLC with specific mechanical characteristics and specific output factors. Each HTU is connected with one specific operator station (OS) and with its specific tomotherapy planning station (ver. 4.2.1.2), a dedicated treatment planning system (TPS) equipped with an inverse planning algorithm able to optimize and calculate IMRT treatment plans. A planning transfer station (PTS) is also connected to each unit in order to allow the patient to transfer from different treatment machines. Each HTU, operator station, planning station, and PTS are connected into its own database. In addition to the HTUs, the radiotherapy department is equipped with two conventional linacs (one Clinac 2100CD and one Clinac iX; Varian Medical Systems, Palo Alto, CA) able to deliver IMRT and volumetric‐modulated arc treatments (RapidArc). A simulator Acuity (Varian) and a dedicated CT GE HiSpeed (GE Healthcare, Waukesha, WI) are also installed in the radiotherapy department. The staff of the Radiotherapy Department includes ten radiation oncologists and 18 radiographers. The Medical Physics Department includes nine medical physicists, one radiographer, and four technicians. Around 200 treatments were performed every year on each HTU.

III. RESULTS & DISCUSSION

The process tree of the preplanning imaging and volume determination stage is shown in Fig. 1; the process tree of the planning stage is reported in Fig. 2. Globally, 58 subprocesses were identified, starting from the patient identification up to the QA approval of the treatment plan. All the subprocesses were judged to be potentially prone to one or more failure modes.
Figure 1

Subprocesses of the preplanning imaging and volume determination stage in tomotherapy.

Figure 2

Subprocesses of the planning stage in tomotherapy.

A total of 74 failure modes were identified: 38 in the stage of preplanning, imaging, and volume determination, and 36 in the stage of planning. Fifty‐three failures (i.e., 72% of the cases) were considered of little concern in view of the RPN value lower or equal to 80. These failures are summarized in 1, 2 in a condensed form. A recurring minor failure in Table 1 was the improper imaging. It may occur in many subprocesses and, in most of cases, can be easily detected during the target/OAR contouring phase or simply during the upload and the visualization of the CT images. Typical examples are short CT slices acquisition, CT acquisition without an adequate patient preparation for special protocols (e.g., full/empty rectum, full/empty bladder), CT acquisition without an optimal immobilization system). In these cases the effect on patient safety is of little concern, consisting of additional CT exposure and consequent delay in the successive procedures. These same effects may occur when imaging data are lost, incomplete or corrupted during data transfer and saving procedures.
Table 1

Application of failure mode and effects analysis for the preplanning imaging and volume determination stage in tomotherapy. Failure modes having an assigned RPN ≤ 80 are listed

Subprocess Potential Failure Mode
IPatient identificationImaging of the wrong patient at a different anatomical site
Imaging of the wrong patient at the correct anatomical site
IIReading information on the RT record: recommended immobilization systemImproper imaging
IIIReading information on the RT record: patient position (prone‐supine)Improper imaging
IVReading information on the RT record: anatomical sitesImproper imaging
Improper planning optimization
VReading information on the RT record: imaging choiceImproper imaging
VIDefinition of the positioning system (not customized system) of the patient on the CT simulator couchImproper immobilization
VIIIdentification of the customized positioning system of the patient on the CT simulator couchImproper immobilization
VIIIPositioning/alignment of the patient on the CT couch on the basis of the laserImproper imaging
IXDefinition of the temporary isocenter on the basis of the anatomical site to be treated and placement of 3 reference markersImproper imaging due to impossibility to define red laser position
XIRecall the patient identity record from the RIS systemIncorrect association CT imaging ‐ patient
XIISelection of the CT scan protocolImproper imaging
XIIIAcquisition scout and scanning and CT imaging reconstructionImproper imaging ‐ wrong anatomical region
XIVCheck of patient preparationImproper imaging
XVPatient tatoos in the position of the 3 markers at the temporary isocenterWrong isocenter
XVISave imaging data of the patientData loss
XVIISending CT data and imaging to PACSIncomplete or corrupt data
Inability to find the patient data, research on other nodes, and/or resending
XVIIIPossible other imaging procedures (MRI, CT‐PET with definition of BTV, CT with contrast medium)Incomplete or corrupt data ‐ inability to recover the images
Difficulty in the image fusion
XIXPlacement of the RT record in the folder “Planning to be done” of the referring physicianRT record loss
XXRecall of the list of the CT imagesCT images not found
XXIRecovery of the CT images from PACS to TPSImages associated to the wrong patient Impossibility to reconstruct previous treatments
XXIIPossible recovery of other images (from PACS via DICOM or tomotherapy station)Incomplete or corrupt data
XXVIAutomatic contour of the OARWrong OAR definition
XXVIIManual contour of the OARWrong OAR definition
Table 2

Application of failure mode and effects analysis for the planning stage in tomotherapy. Failure modes having an assigned are listed

Subprocess Potential Failure Mode
XXXIIn case of retreatment, reconstruction of the previous treatment planningWrong reconstruction Reconstruction not possible Reconstruction suboptimal
XXXIIPrescription of PTV dose (from protocol or personalized)Missing prescription in the record
XXXIIIPrescription of dose to overlapping regions (from protocol or personalized)Missing prescription in the record
XXXIVDefinition of the fractionation of PTVs: number of fractions and daily dosesMissing prescription in the record
XXXVDefinition of specific dose limits for OAR not to be exceededMissing prescription in the record
XXXVIPlacement of the RT record in the folder “Planning to be done” according to the various therapy units.Missing record Planning for the wrong unit
XXXVIIDefinition of help structure regions to make easier the optimizationIncorrect help structure region definition
XXXIXSend the contoured CT images from the Somavision/Eclipse station to the TPS of the chosen tomotherapy unitWrong data sent
XLIAssignment of the contoured structures as target and OARWrong assignment
XLIIIPlacement of the mobile red lasers at the tattoos, if the distance between red and green lasers is higher that a max value, then placement of red lasers in other site and notification the distance markers/red lasers to the RT technician.Wrong positioning
XLIVChoice of the filed dimensionSuboptimal choice
XLVChoice of the pitchSuboptimal choice
XLVIIChoice of the target structure for the dose prescription, and choice of the dose prescription modalityWrong choice
XLIXChoice of the modulation factorSuboptimal choice
LDefinition of dose‐volume constraints to be associated to the targets and to the OAR for the optimization phaseSuboptimal choice
LIChoice of the weights and penalties to be associated to the structures and to each dose‐volume constraintSuboptimal choice
LIIBeamlet calculationBeamlet recalculation
LIIIOptimizationIncomplete optimization
LVVideo approval of planning (without this approval it is not possible to treat the patient)Approval missing
LVITreatment plan printingPrinted copy of the treatment plan missing
LVIIVideo signature of the planningSignature missing
LVIIIQA approvalQA approval missing
Subprocesses of the preplanning imaging and volume determination stage in tomotherapy. In other cases, improper imaging could be more difficult to detect during the planning stage, as attested by severe incidents reported in the recent past. In particular, a possible failure is the reversal of images as a consequence of incorrect positioning of the patient and/or incorrect selection of the CT scan protocol (i.e., “head first” vs. “feet first”). However, in tomotherapy, thanks to the check of the patient position performed just before the treatment by means of the MVCT scanner, the detectability of this failure can be considered sufficiently high to make the overall risk acceptable. Furthermore, the MVCT pre‐treatment imaging permits identification of possible discrepancies between the position taken by the patient during the preplanning imaging and that required for the treatment (e.g., arms up/down). Subprocesses of the planning stage in tomotherapy. Possible minor failures in the planning stage (Table 2) caused by lack of attention are the missing of prescription data such as dose to planned target volume (PTV) or to overlapping regions, number of fractions, and dose limits to organs at risk (OAR). Since these data are strictly required by the medical physicist to start the treatment planning, their absence will only reflect in a delay of the process flow. Application of failure mode and effects analysis for the preplanning imaging and volume determination stage in tomotherapy. Failure modes having an assigned RPN ≤ 80 are listed Inadequate skill or lack of attention is also cause of other minor failures in the planning stage, such as the suboptimal choice of one or more parameters of the TPS (e.g., calculation matrix, field dimension, pitch, modulation factor, constrains). The possible effect of these failures was the delivery of a suboptimal treatment. Application of failure mode and effects analysis for the planning stage in tomotherapy. Failure modes having an assigned are listed Twenty‐one failure modes (28% of cases) were characterized by a RPN score higher than 80. Ten of them, shown in Table 3, were identified in the stage of preplanning imaging and volume determination. The remaining 11 failures, which may occur during the stage of plan‐ning, are shown in Table 4. The mean values of the risk indexes of Table 3 were (range 3–4), (range 5–8), and (rage 3–6). Similarly, the mean values of the indexes of Table 4 were: (range 2–4), (range 5–8), and (range 4–9). These data suggest that such events, on average, are infrequent (once or a few times a year) and not too difficult to detect (especially for the stage of preplanning imaging and volume determination), but potentially severe in terms of patient safety. The threshold of 125 for RPN was exceeded in four cases: one case only in the phase of preplanning imaging and volume determination, and three cases in the stage of planning. Actually, treatment planning is known to be one of the most critical phases within the whole RT process.
Table 3

Application of failure mode and effects analysis for the preplanning imaging and volume determination stage in tomotherapy. Failure modes having an assigned are reported

Subprocess No Potential Failure Mode Potential Causes of Failure Potential Effects of Failure O S D RPN
VIII Positioning/alignment of the patient on the CT couch on the basis of the laser1Wrong definition of the isocenterInadequate CT laser alignmentSystematic shift of the patient position48396
X Reference markers positioning to indicate previous treatment fields2Missing marker positionsLack of attention or incomplete compilation of the RT recordPrevious treatment not taken into consideration/suboptimal planning46396
XXIII Registration of possible other images from different techniques and check of consistency3Wrong registrationConsistency not verifiedWrong PTV and OAR definition375105
XXIV Definition of the GTVs and/or CTVs contour on the basis of the anatomical and/or functional information (BTV)4Wrong CTV definitionLack of attention/inadequate skillWrong dose distribution375105
XXV Definition of ITV in the case of 4D CT acquisition5Incorrect ITV constructionLack of attention/inadequate skillWrong PTV36590
XXVI Automatic contour of the OAR6Missing OAR definitionLack of attention/inadequate skillUnintended normal tissue irradiation385120
XXVII Manual contour of the OAR7Missing OAR definitionLack of attention/inadequate skillUnintended normal tissue irradiation385120
XXVIII Defnition of the planning structure at risk through automatic expansion of OAR8Wrong expansionLack of attention/inadequate skillUnintended normal tissue irradiation35696
XXIX Automatic expansion with definition of the CTVs and PTVs margins9Wrong expansionLack of attention/inadequate skillUnintended normal tissue irradiation35696
XXX Definition and contouring of overlapping regions10Wrong/missed definitionLack of attention/inadequate skill or not detailed information on previous treatmentUnintended normal tissue irradiation or wrong dose distribution475140
Table 4

Application of failure mode and effects analysis for the planning stage in tomotherapy. Failure modes having an assigned are reported

Subprocess No Potential Failure Mode Potential Causes of Failure Potential Effects of Failure O S D RPN
XXXII Prescription of PTV dose (from protocol or personalized)11Wrong prescription in the recordLack of attention/inadequate skillWrong dose delivery38496
XXXIII Prescription of dose to overlapping regions (from protocol or personalized)12Wrong prescription in the recordLack of attention/inadequate skillPossible wrong dose distribution/wrong dose delivery38496
XXXIV Definition of the fractionation of PTVs: number of fractions and daily doses13Wrong prescription in the recordLack of attention/inadequate skillWrong dose delivery38496
XXXV Definition of specific dose limits for OAR, not to be exceeded14Wrong prescription in the recordLack of attention/inadequate skillUnintended normal tissue irradiated38496
XXXVIII Possible contouring of the couch15Incorrect dose calculationIncorrect positioning in the imagingWrong dose delivery26784
XL If not automatically done, replacement in the imaging of the CT couch (diagnostic) with tomoterapy couch16Incorrect dose calculationIncorrect introduction of the couch position in the CT imagesWrong dose delivery26896
XLII Assignment of the Overlap Priority to each structure17Wrong assignmentLack of attention/inadequate skillWrong dose distribution477196
XLVI Choice of the calibration curve nCT‐nHU18Wrong choice (kV‐MV)Lack of attention/inadequate skillWrong dose calculation/wrong dose delivery289144
19Wrong choice (kV‐MV)Lack of attention/inadequate skillWrong dose calculation/wrong dose delivery25990
XLVIII Choice of the calculation matrix20Suboptimal choiceLack of attention/inadequate skillSuboptimal treatment35690
LIV Introduction of the number of fractions and automatic generation of number of sessions21Wrong or not performed choice (erroneous use of the default value)Lack of attention/inadequate skillWrong dose delivery386144
Application of failure mode and effects analysis for the preplanning imaging and volume determination stage in tomotherapy. Failure modes having an assigned are reported Application of failure mode and effects analysis for the planning stage in tomotherapy. Failure modes having an assigned are reported The failure mode with the highest risk (i.e., ) within the stage of preplanning imaging and volume determination was the wrong or missing definition and contouring of the overlapping regions. Overlapping regions are, in general, defined to limit the prescription dose to some structures, both in case of overlapping between target volume and dose limit to organs at risk, and in case of retreatment. Typical examples of overlapping regions are the boundaries between PTV and chiasma and/or optical structures, PTV and anterior rectal wall, and PTV and duodenum and/or stomach. In tomotherapy, the overlapping regions of adjacent anatomical structures need to be defined and contoured, and a priority level has to be assigned during planning in order to achieve the correct dose distribution to the PTV and, at the same time, to spare the OAR. If one or more overlapping regions are not properly defined or contoured by the radiation oncologist, the potential effect for the patient could be an unintended normal tissue irradiation or, more likely, a wrong dose distribution. The main cause of this failure was identified as a lack of attention or inadequate skill of the radiation oncologist in charge of defining and contouring the volumes of interest. Moreover, this failure might occur as a consequence of a lack of exhaustive clinical documentation about possible previous RT treatments. Indeed, the doses received by the various anatomical structures during previous irradiations have to be taken into account for defining the overlapping regions. An additional safety measure was suggested by the working group for reducing the probability of occurrence of this failure, and therefore for mitigating the overall risk. It consisted of improving the RT record of the patient through the systematic introduction of a sheet to be flagged containing detailed information about the prescribed dose to target volumes, OARs, and overlapping regions contoured with relative dose constraints. The most critical failure in the stage of planning was the wrong assignment of the overlap priority to each anatomical structure by the medical physicist. The TPS prompts the user to divide structures or regions of interest (ROI) into two groups: tumors and organs at risk. A ROI from the tumor group can overlap with a ROI in the OAR group, but the ROIs within the same group are not allowed to own common voxels. When two contours from the same group overlap with each other, the overlap priority setting governs to which structure the voxel belongs. The structure with the higher overlap priority will own the voxels in the overlap region, and the structure with lower overlap priority will lose these voxels for dose volume computations, optimization, and dose‐volume histograms (DVHs) in the overlapping mode visualization. Therefore, overlap priorities must be carefully selected since, in case of failure, wrong dose optimization can be generated with possible serious consequences for the patient. As for many other failures, this error was recognized as the result of a lack of attention or inadequate skill of the medical physicist in charge of planning. The probability of this failure going undetected was estimated to be relatively high if the final plan evaluation is based on only the DVH in the overlapping mode (i.e., the usual way to visualize the DVH during plan optimization). For this reason, the additional safety measure recommended by the WG consisted of a systematic final plan evaluation considering both the DVH in the overlapping mode and in the standard mode (i.e., the mode that provides the statistics of the entire volumes), regardless the defined overlapping priorities. It is interesting to note that the possibility to visualize and to print out the DVHs in the two different modes is an option of one of the latest TPS release. A further critical failure identified in the planning stage was the wrong choice of the CT calibration curve. This failure might accordingly occur when more than one CT scanning system is available, each of them with the proper calibration curve. Since the dose calculation process by the TPS is based on the data of the CT calibration, the wrong choice of calibration curve could reflect in a wrong dose calculation and, possibly, in a wrong dose delivery. This failure could be particularly critical in tomotherapy since, in the case of using MVCT instead of kVCT images, or vice‐versa, a significant variation in the calculated dose (i.e., of the order of 10%) is expected. Indeed, the relationship of MVCT number to electron density is different from that observed in common kVCT scanners as a consequence of the difference in physical interaction probabilities of photons within the two energy ranges. This failure was considered extremely hard to detect (i.e., ). Indeed, the information about the CT calibration curve chosen for dose calculation cannot be included in the printed copy of the treatment plan. Therefore, the correctness of this parameter cannot be checked during the plan evaluation before the final approval. On the basis of this analysis, the WG decided, as additional strategy, to suggest the vendor implement a new function in the TPS software, consisting of the possibility for the user to include the information on the CT calibration curve in the printed copy of the treatment plan. The last major failure (with a ) which might occur during the planning stage, was the wrong (or not performed) choice of the number of fractions composing the treatments. A default number of fractions equal to 30 is set in the TPS. This value must be changed by the planner according to the particular prescription. If the planner fails to perform this task, the consequences for the patient could be particularly severe, since a wrong daily dose could be delivered during the treatment. The additional safety measure recommended by the WG is to increase the probability of failure detection by the RT technician before the treatment start through the systematic check of the agreement between the number of fractions actually prescribed in the radiotherapy sheet and the number of procedures visualized on the screen of the tomotherapy control unit. In addition to the failure modes with the highest overall risks, specific attention should be paid to the failures which could lead to severe injuries to the patients, independently of the RPN value. In this analysis, seven failure modes characterized by a severity index and were identified (i.e., failures No. 1, 6 and 7 in Table 3 and Nos. 11–14 in Table 4). The wrong definition of the isocenter during the positioning of the patient on the CT couch, even though potentially dangerous, appeared easily detectable, since the laser alignment is daily checked through the use of the MVCT scanner. No further safety measures are, therefore, required. The risk of missing OAR definitions can be simply mitigated through the implementation of the safety measures formerly proposed. In particular, the suggested improvement of the RT record could significantly reduce the number of occurrences of these failures. Finally, additional safety measures are required to deal with the remaining failures — the wrong prescription of dose and/or dose limits. Systematic double‐check of these data by the radiation therapy staff, and/or the definition of reference protocols, could be an adequate solution increasing the detection rating of these failures. The failure modes shown in 1, 2, 3, 4 were identified by analyzing the typical sequence of events and procedures that characterize the standard RT process. However, in clinical practice, some deviations from the standard process flow might occur as a consequence of specific medical or organizational needs. Unfortunately, such process changes cannot be considered free from additional potential failures. In particular, three events which could affect the patient safety were identified by the working group and are briefly described below. 1. When the prescribed dose to the PTV is high, it may occur that the HTU is not able to deliver the entire dose in a single fraction; the physicist has to duplicate the number of fractions and write in the radiotherapy sheet to deliver two fractions every daily treatment session. If the indication is not clearly reported, this can result in an uncorrected daily dose delivery to the patient. 2. Sometimes, it could happen that the radiation oncologist decides to deliver more fractions than planned (e.g., if the treatment needs to be interrupted in a case where the patient experienced severe side effects). The TPS does not allow the physicist to add treatment fractions without replanning and recalculating the plan, but only those QA fractions on the operator station. So the patient can be treated, but the final treatment report does not record the added fractions and only the radiotherapy sheet can attest the real delivered dose. 3. As previously reported, at San Raffaele Scientific Institute two HTUs are currently operating. The two machines are twinned in order to have the possibility to transfer the patient from one unit to the other one (in a case where one machine needs maintenance). In such a case, the patient should be immediately cancelled from the original unit, transferred to the second one, treated, deleted from the TPS of the second unit, and finally transferred again to the original unit. In fact, this procedure would permit the tracking of the treatments actually done by the patient on the two units. However, it must be considered that every transfer entails a new final dose calculation (taking into account the specific output factors and MLC mechanical characteristics of each unit); moreover, each of these operations takes a lot of time to be completed. So, usually the patient record remains simultaneously on the two TPS for few days and, in this way, the only possibility to check the number of fractions actually delivered is through the radiotherapy sheet of the patient. In all these circumstances, more flexibility of the system should let the operator to modify the schedule of the treatment so as to avoid communication errors that can induce an incorrect dose to patients. One solution could be the implementation of a record and verify system; a session, daily, and total dose accumulation point could also be a safety issue. In addition to the safety measures specifically related to tomotherapy proposed by the WG, the results of this analysis confirmed the soundness of the general lessons and recommendations provided by ICRP for preventing accidental exposures from new external beam radiation therapy technologies. Indeed, during the identification of the causes of failures and the estimation of their occurrence probability, it came out that the competence and skill of the various professional figures involved in the RT process, as well as the work environment, play an important role both for the quality of the treatment and for patient safety. Regular updating of the knowledge and maintenance of the skills of the personnel can, therefore, represent an efficient instrument for patient safety improvement. In particular, the development of specific professional training schemes on the functionalities and limits of the various systems and software used in the RT process, as well as on the procedures and protocols, can be considered a general action for reducing the frequency of failures and, consequently, the overall risk of accidents in tomotherapy. Furthermore, excessive workload should be avoided, and a suitable work environment that encourages working with awareness, facilitates concentration, and avoids distraction should be provided. Finally, as in other modern RT techniques, in tomotherapy in vivo dosimetry could provide a reliable way of detecting serious anomalies potentially leading to substantial overdoses or underdoses to the patient.

IV. CONCLUSIONS

The application of FMEA to the preplanning imaging, volume determination, and treatment planning stages in tomotherapy led to the identification and deep investigation of various failure modes. The assignment of a score assessing the potential risk for each event permitted a ranking of these failure modes in order of importance and the ability to define priorities for risk mitigation, with the aim of optimizing quality management workflow. In addition to the safety strategies already adopted in the clinical practice, novel solutions have been proposed to increase patient safety. Although this study was carried out considering the specific processes implemented at San Raffaele Scientific Institute, the proposed methodology, as well as likely most of the findings, can be generalized and made useful to other RT centers equipped with HTU. On the basis of the results obtained in this study and of the experience accrued by the WG, further stages of the RT process, such as treatment delivery and treatment monitoring and verification, will be analyzed by means of the FMEA approach in the near future.
  14 in total

1.  Failure mode and effect analysis for delivery of lung stereotactic body radiation therapy.

Authors:  Julian R Perks; Sinisa Stanic; Robin L Stern; Barbara Henk; Marsha S Nelson; Rick D Harse; Mathew Mathai; James A Purdy; Richard K Valicenti; Allan D Siefkin; Allen M Chen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-12-22       Impact factor: 7.038

2.  QA for helical tomotherapy: report of the AAPM Task Group 148.

Authors:  Katja M Langen; Niko Papanikolaou; John Balog; Richard Crilly; David Followill; S Murty Goddu; Walter Grant; Gustavo Olivera; Chester R Ramsey; Chengyu Shi
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

3.  Probabilistic fault tree analysis of a radiation treatment system.

Authors:  Edidiong Ekaette; Robert C Lee; David L Cooke; Sandra Iftody; Peter Craighead
Journal:  Risk Anal       Date:  2007-12       Impact factor: 4.000

4.  ICRP publication 112. A report of preventing accidental exposures from new external beam radiation therapy technologies.

Authors:  P Ortiz López; J M Cosset; P Dunscombe; O Holmberg; J C Rosenwald; L Pinillos Ashton; J J Vilaragut Llanes; S Vatnitsky
Journal:  Ann ICRP       Date:  2009-08

Review 5.  The European Society of Therapeutic Radiology and Oncology-European Institute of Radiotherapy (ESTRO-EIR) report on 3D CT-based in-room image guidance systems: a practical and technical review and guide.

Authors:  Stine Korreman; Coen Rasch; Helen McNair; Dirk Verellen; Uwe Oelfke; Philippe Maingon; Ben Mijnheer; Vincent Khoo
Journal:  Radiother Oncol       Date:  2010-02-12       Impact factor: 6.280

6.  Application of failure mode and effects analysis to intraoperative radiation therapy using mobile electron linear accelerators.

Authors:  Mario Ciocca; Marie-Claire Cantone; Ivan Veronese; Federica Cattani; Guido Pedroli; Silvia Molinelli; Viviana Vitolo; Roberto Orecchia
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-06-25       Impact factor: 7.038

7.  Evaluation of safety in a radiation oncology setting using failure mode and effects analysis.

Authors:  Eric C Ford; Ray Gaudette; Lee Myers; Bruce Vanderver; Lilly Engineer; Richard Zellars; Danny Y Song; John Wong; Theodore L Deweese
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-05-04       Impact factor: 7.038

8.  A method for evaluating quality assurance needs in radiation therapy.

Authors:  M Saiful Huq; Benedick A Fraass; Peter B Dunscombe; John P Gibbons; Geoffrey S Ibbott; Paul M Medin; Arno Mundt; Sassa Mutic; Jatinder R Palta; Bruce R Thomadsen; Jeffrey F Williamson; Ellen D Yorke
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008       Impact factor: 7.038

9.  Cone beam CT pre- and post-daily treatment for assessing geometrical and dosimetric intrafraction variability during radiotherapy of prostate cancer.

Authors:  Giacomo Reggiori; Pietro Mancosu; Angelo Tozzi; Marie C Cantone; Simona Castiglioni; Paola Lattuada; Francesca Lobefalo; Luca Cozzi; Antonella Fogliata; Piera Navarria; Marta Scorsetti
Journal:  J Appl Clin Med Phys       Date:  2010-12-02       Impact factor: 2.102

10.  Application of failure mode and effects analysis to treatment planning in scanned proton beam radiotherapy.

Authors:  Marie Claire Cantone; Mario Ciocca; Francesco Dionisi; Piero Fossati; Stefano Lorentini; Marco Krengli; Silvia Molinelli; Roberto Orecchia; Marco Schwarz; Ivan Veronese; Viviana Vitolo
Journal:  Radiat Oncol       Date:  2013-05-24       Impact factor: 3.481

View more
  9 in total

1.  Failure modes and effects analysis of total skin electron irradiation technique.

Authors:  B Ibanez-Rosello; J A Bautista; J Bonaque; J Perez-Calatayud; A Gonzalez-Sanchis; J Lopez-Torrecilla; L Brualla-Gonzalez; T Garcia-Hernandez; A Vicedo-Gonzalez; D Granero; A Serrano; B Borderia; C Solera; J Rosello
Journal:  Clin Transl Oncol       Date:  2017-08-04       Impact factor: 3.405

2.  The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality management.

Authors:  M Saiful Huq; Benedick A Fraass; Peter B Dunscombe; John P Gibbons; Geoffrey S Ibbott; Arno J Mundt; Sasa Mutic; Jatinder R Palta; Frank Rath; Bruce R Thomadsen; Jeffrey F Williamson; Ellen D Yorke
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

3.  Identifying risk characteristics using failure mode and effect analysis for risk management in online magnetic resonance-guided adaptive radiation therapy.

Authors:  Shie Nishioka; Hiroyuki Okamoto; Takahito Chiba; Tatsuya Sakasai; Kae Okuma; Junichi Kuwahara; Daisuke Fujiyama; Satoshi Nakamura; Kotaro Iijima; Hiroki Nakayama; Mihiro Takemori; Yuuki Tsunoda; Keita Kaga; Hiroshi Igaki
Journal:  Phys Imaging Radiat Oncol       Date:  2022-06-06

4.  Multi-institutional application of Failure Mode and Effects Analysis (FMEA) to CyberKnife Stereotactic Body Radiation Therapy (SBRT).

Authors:  Ivan Veronese; Elena De Martin; Anna Stefania Martinotti; Maria Luisa Fumagalli; Cristina Vite; Irene Redaelli; Tiziana Malatesta; Pietro Mancosu; Giancarlo Beltramo; Laura Fariselli; Marie Claire Cantone
Journal:  Radiat Oncol       Date:  2015-06-13       Impact factor: 3.481

5.  Failure mode and effects analysis: A community practice perspective.

Authors:  Bradley W Schuller; Angi Burns; Elizabeth A Ceilley; Alan King; Joan LeTourneau; Alexander Markovic; Lynda Sterkel; Brigid Taplin; Jennifer Wanner; Jeffrey M Albert
Journal:  J Appl Clin Med Phys       Date:  2017-09-25       Impact factor: 2.102

6.  Failure mode and effects analysis of skin electronic brachytherapy using Esteya® unit.

Authors:  Blanca Ibanez-Rosello; Juan Antonio Bautista-Ballesteros; Jorge Bonaque; Francisco Celada; Françoise Lliso; Vicente Carmona; Jose Gimeno-Olmos; Zoubir Ouhib; Joan Rosello; Jose Perez-Calatayud
Journal:  J Contemp Brachytherapy       Date:  2016-12-20

7.  Failure modes and effects analysis (FMEA) for Gamma Knife radiosurgery.

Authors:  Andy Yuanguang Xu; Jagdish Bhatnagar; Greg Bednarz; John Flickinger; Yoshio Arai; Jonet Vacsulka; Wenzheng Feng; Edward Monaco; Ajay Niranjan; L Dade Lunsford; M Saiful Huq
Journal:  J Appl Clin Med Phys       Date:  2017-10-29       Impact factor: 2.102

8.  A Modular System for Treating Moving Anatomical Targets With Scanned Ion Beams at Multiple Facilities: Pre-Clinical Testing for Quality and Safety of Beam Delivery.

Authors:  Michelle Lis; Wayne Newhauser; Marco Donetti; Moritz Wolf; Timo Steinsberger; Athena Paz; Marco Durante; Christian Graeff
Journal:  Front Oncol       Date:  2021-03-19       Impact factor: 6.244

9.  Evaluation and improvement the safety of total marrow irradiation with helical tomotherapy using repeat failure mode and effects analysis.

Authors:  Jiuling Shen; Xiaoyong Wang; Di Deng; Jian Gong; Kang Tan; Hongli Zhao; Zhirong Bao; Jinping Xiao; An Liu; Yunfeng Zhou; Hui Liu; Conghua Xie
Journal:  Radiat Oncol       Date:  2019-12-27       Impact factor: 3.481

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.