Effectiveness of VF induction with DC fibber versus conventional induction methods in patients on chronic amiodarone therapy.

Jason C Rubenstein1, Manish S Gupta, Michael H Kim.   

Abstract

BACKGROUND: Amiodarone therapy, especially chronic, can result in difficult ventricular fibrillation (VF) inductions during implantable cardioverter defibrillator (ICD) testing. The efficacy of various VF induction methods on patients treated with amiodarone has not been well described. This prospective analysis evaluated the impact of direct current (DC) fibber, burst fibber, and synchronized T-wave shock VF induction methods.
METHODS: Data were collected from one study and two enrolling centers totaling 14 ICD patients (92.9 % male, age 64.3 years (range 23.6-81.2)). A minimum of two successful VF inductions from each patient were required to be included in the study analysis. Each VF induction method was attempted in each patient. Non-sustained VF resulting in an aborted shock was considered an induction failure. All 14 patients were on chronic amiodarone therapy.
RESULTS: From a total of 42 attempted inductions, 25 (59.5 %) were successful at inducing VF. The success rates of effective induction of VF during the first attempt using DC, burst, and Shock-on-T are 100, 50, and 28.6 %, respectively. DC fibber had significantly higher VF induction rates than Shock-on-T (p = 0.004) or burst fibber (p = 0.02). There was no difference between Shock-on-T and burst fibber (p = 0.45).
CONCLUSION: The success rate of VF induction using DC fibber method is significantly higher than either Shock-on-T or burst fibber induction methods for patients on chronic amiodarone therapy. This may facilitate defibrillation threshold testing in such patients.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24013703     DOI: 10.1007/s10840-013-9819-1

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  13 in total

1.  Myocardial stunning following defibrillation threshold testing.

Authors:  Michael Mollerus; Leslee Naslund
Journal:  J Interv Card Electrophysiol       Date:  2007-09-11       Impact factor: 1.900

2.  Is defibrillation testing safe?

Authors:  I Singer; J van der Laken; H L Edmonds; A D Slater; E Austin; C B Shields; J Kupersmith
Journal:  Pacing Clin Electrophysiol       Date:  1991-11       Impact factor: 1.976

3.  Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction.

Authors:  Arthur J Moss; Wojciech Zareba; W Jackson Hall; Helmut Klein; David J Wilber; David S Cannom; James P Daubert; Steven L Higgins; Mary W Brown; Mark L Andrews
Journal:  N Engl J Med       Date:  2002-03-19       Impact factor: 91.245

4.  Low voltage direct current delivered through unipolar transvenous leads: an alternate method for the induction of ventricular fibrillation.

Authors:  D E Euler; T A Whitman; P R Roberts; M J Kallok
Journal:  Pacing Clin Electrophysiol       Date:  1999-06       Impact factor: 1.976

5.  The defibrillation threshold: how many measurements are enough?

Authors:  O Fujimura; D L Jones; G J Klein
Journal:  Am Heart J       Date:  1989-04       Impact factor: 4.749

6.  Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators.

Authors:  A J Moss; W J Hall; D S Cannom; J P Daubert; S L Higgins; H Klein; J H Levine; S Saksena; A L Waldo; D Wilber; M W Brown; M Heo
Journal:  N Engl J Med       Date:  1996-12-26       Impact factor: 91.245

7.  A comparison of standard and high-dose regimens for the initiation of amiodarone therapy.

Authors:  J Summitt; F Morady; A Kadish
Journal:  Am Heart J       Date:  1992-08       Impact factor: 4.749

8.  Efficacy of low-energy T wave shocks for induction of ventricular fibrillation in patients with implantable cardioverter defibrillators.

Authors:  A K Bhandari; N Isber; M Estioko; T Ziccardi; D S Cannom; Y Park; R D Lerman; C Prejean; G W Sun
Journal:  J Electrocardiol       Date:  1998-01       Impact factor: 1.438

9.  Shock on T versus direct current voltage for induction of ventricular fibrillation: a randomized prospective comparison.

Authors:  Arjun D Sharma; Eric Fain; P Gearoid O'Neill; Anne Skadsen; Roger Damle; Jim Baker; Vinod Chauhan; Meir Mazuz; Terrance Ross; Ziaozheng Zhang
Journal:  Pacing Clin Electrophysiol       Date:  2004-01       Impact factor: 1.976

10.  Complications associated with defibrillation threshold testing: the Canadian experience.

Authors:  David Birnie; Stanley Tung; Christopher Simpson; Eugene Crystal; Derek Exner; Felix-Alejandro Ayala Paredes; Andrew Krahn; Ratika Parkash; Yaariv Khaykin; Francois Philippon; Peter Guerra; Shane Kimber; Douglas Cameron; Jeffrey S Healey
Journal:  Heart Rhythm       Date:  2007-11-28       Impact factor: 6.343

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.