Literature DB >> 24008137

Monotonic and fatigue behavior of five clinically relevant conventional and highly crosslinked UHMWPEs in the presence of stress concentrations.

Michael C Sobieraj1, James E Murphy, Jennifer G Brinkman, Steve M Kurtz, Clare M Rimnac.   

Abstract

Five formulations of clinically relevant UHMWPE (conventional, moderately crosslinked annealed and remelted, and highly crosslinked annealed and remelted) were investigated in a physiologically relevant environment. Their monotonic stress-strain behavior in the presence of notches of two different severities and at two different displacement rates was examined using a custom developed video based system. It was found that both an elevation of yield stress and a truncation of orientation hardening took place under monotonic loading and that these changes were found to be material and elastic stress concentration factor dependent. The fatigue behavior of these materials was examined using the same geometries via a stress-life approach with failure defined as fracture of the specimen in the 1000 to 100,000 cycle lifetime range. The results were modeled using the Basquin relationship (σ=AN(b), where σ=stress and N=lifetime, and A and b are experimentally derived constants) via maximum likelihood estimation methods to account for specimen runout (no failure at 250,000cycles). The conventional material was found to have a greater slope, b, and intercept, A, than the crosslinked materials as well as appearing to have less variance in its failure distributions.
© 2013 Published by Elsevier Ltd.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24008137      PMCID: PMC3834138          DOI: 10.1016/j.jmbbm.2013.07.024

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  16 in total

1.  Effect of sterilization method and other modifications on the wear resistance of acetabular cups made of ultra-high molecular weight polyethylene. A hip-simulator study.

Authors:  H McKellop; F W Shen; B Lu; P Campbell; R Salovey
Journal:  J Bone Joint Surg Am       Date:  2000-12       Impact factor: 5.284

2.  Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements.

Authors:  H McKellop; F W Shen; B Lu; P Campbell; R Salovey
Journal:  J Orthop Res       Date:  1999-03       Impact factor: 3.494

3.  Notch strengthening and hardening behavior of conventional and highly crosslinked UHMWPE under applied tensile loading.

Authors:  M C Sobieraj; S M Kurtz; C M Rimnac
Journal:  Biomaterials       Date:  2005-06       Impact factor: 12.479

4.  Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030.

Authors:  Steven Kurtz; Kevin Ong; Edmund Lau; Fionna Mowat; Michael Halpern
Journal:  J Bone Joint Surg Am       Date:  2007-04       Impact factor: 5.284

5.  Comparison between highly cross-linked and conventional polyethylene in total knee arthroplasty.

Authors:  Yukihide Minoda; Masaharu Aihara; Akira Sakawa; Shinichi Fukuoka; Keiko Hayakawa; Masuhiro Tomita; Naoya Umeda; Kenji Ohzono
Journal:  Knee       Date:  2009-03-05       Impact factor: 2.199

6.  Highly crosslinked polyethylene is safe for use in total knee arthroplasty.

Authors:  Jeffrey T Hodrick; Erik P Severson; Deborah S McAlister; Brian Dahl; Aaron A Hofmann
Journal:  Clin Orthop Relat Res       Date:  2008-09-10       Impact factor: 4.176

7.  Notched stress-strain behavior of a conventional and a sequentially annealed highly crosslinked UHMWPE.

Authors:  Michael C Sobieraj; Steven M Kurtz; A Wang; Michael M Manley; Clare M Rimnac
Journal:  Biomaterials       Date:  2008-09-17       Impact factor: 12.479

8.  Tradeoffs amongst fatigue, wear, and oxidation resistance of cross-linked ultra-high molecular weight polyethylene.

Authors:  Sara A Atwood; Douglas W Van Citters; Eli W Patten; Jevan Furmanski; Michael D Ries; Lisa A Pruitt
Journal:  J Mech Behav Biomed Mater       Date:  2011-03-11

Review 9.  Ultra high molecular weight polyethylene: mechanics, morphology, and clinical behavior.

Authors:  M C Sobieraj; C M Rimnac
Journal:  J Mech Behav Biomed Mater       Date:  2008-12-25

10.  Comparative fatigue behavior and toughness of remelted and annealed highly crosslinked polyethylenes.

Authors:  Francisco J Medel; P Peña; José Cegoñino; E Gómez-Barrena; J A Puértolas
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2007-11       Impact factor: 3.368

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.