Literature DB >> 23981817

Effect of vertebroplasty on the compressive strength of vertebral bodies.

Spiros G Pneumaticos1, Georgios K Triantafyllopoulos, Dimitrios S Evangelopoulos, John A Hipp, Michael H Heggeness.   

Abstract

BACKGROUND CONTEXT: Percutaneous vertebroplasty has been used successfully for many years in the treatment of painful compressive vertebral fractures due to osteoporosis.
PURPOSE: To compare the effect of vertebroplasty on the compressive strength of unfractured vertebral bodies. STUDY
DESIGN: Biomechanical study on cadaveric thoracic vertebrae.
METHODS: Forty vertebral bodies from four cadaveric thoracic spines were used for this experiment. Before testing, each thoracic spine was submitted to bone density testing and radiographic evaluation to rule out any obvious fractures. Under image intensification, 6 mL of a mixture of polymethylmethacrylate (PMMA) with barium (8 g of barium/40 g of PMMA) was injected into every other vertebral body of each spine specimen. After vertebroplasty, all soft tissues were dissected from the spine, and the vertebral bodies were separated and potted for mechanical testing. Testing to failure was performed using a combination of axial compression and anterior flexion moments. Two pneumatic cylinders applied anterior and posterior loads at a distance ratio of 4:3 relative to the anterior vertebral body wall, whereas two additional cylinders applied lateral loads, each at a constant rate of 200 N/s.
RESULTS: The average failure loads for nonvertebroplasty specimens was 6724.02 ± 3291.70 N, whereas the specimens injected with PMMA failed at an average compressive force of 5770.50 ± 2133.72 N. No statistically significant difference in failure loads could be detected between intact specimens and those that had undergone vertebroplasty.
CONCLUSIONS: Under these specific loading conditions, no significant increase in compressive strength of the vertebral bodies could be documented. This suggests that some caution should be applied to the concept of "prophylactic" vertebroplasty in patients at risk for fracture.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Compressive strength; Osteoporosis; Polymethylmethacrylate; Thoracic vertebrae; Vertebroplasty

Mesh:

Substances:

Year:  2013        PMID: 23981817     DOI: 10.1016/j.spinee.2013.06.054

Source DB:  PubMed          Journal:  Spine J        ISSN: 1529-9430            Impact factor:   4.166


  8 in total

1.  The methodology of locating painful responsible vertebrae in osteoporotic vertebral compression fractures.

Authors:  Jun Lin; Hao Liu; Peng Yang; Huilin Yang
Journal:  Int J Clin Exp Med       Date:  2015-03-15

2.  Optimizing bone cement stiffness for vertebroplasty through biomechanical effects analysis based on patient-specific three-dimensional finite element modeling.

Authors:  Yi Peng; Xianping Du; Lihua Huang; Jinsong Li; Ruisen Zhan; Weiguo Wang; Biaoxiang Xu; Song Wu; Cheng Peng; Shijie Chen
Journal:  Med Biol Eng Comput       Date:  2018-05-28       Impact factor: 2.602

3.  Dynamic characteristics of osteoporotic lumbar spine under vertical vibration after cement augmentation.

Authors:  Xinlin Su; Hao Shen; Weidong Shi; Huilin Yang; Feng Lv; Jun Lin
Journal:  Am J Transl Res       Date:  2017-09-15       Impact factor: 4.060

4.  [Biomechanical study of polymethyl methacrylate bone cement and allogeneic bone for strengthening sheep vertebrae].

Authors:  Zhikun Wang; Xiansen Zhang; Zaixue Li; Qingyu Feng; Jianting Chen; Wenwei Xie
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2021-04-15

5.  Repeated adjacent segment diseases and fractures in osteoporotic patients: a case report.

Authors:  Hsin-Yao Chen; Chiu-Liang Chen; Wei-Liang Chen
Journal:  Ther Clin Risk Manag       Date:  2016-07-28       Impact factor: 2.423

6.  Gradual Height Decrease of Augmented Vertebrae after Vertebroplasty at the Thoracolumbar Junction.

Authors:  Han San Oh; Tae Wan Kim; Hyun Gon Kim; Kwan Ho Park
Journal:  Korean J Neurotrauma       Date:  2016-04-30

7.  Effect of unilateral pulsed jet lavage prior to vertebroplasty on the intravertebral pressure and cement distribution.

Authors:  Jun Yan; Qiaohui Liu; Yanping Zheng; Ziqun Liu; Xinyu Liu; Xun Guo; Penghui Liu; Peng Chen; Suomao Yuan; Yonghao Tian; Wanlong Xu
Journal:  J Orthop Surg Res       Date:  2020-07-11       Impact factor: 2.359

8.  How the clinical dosage of bone cement biomechanically affects adjacent vertebrae.

Authors:  Xu-Shi Chen; Jian-Ming Jiang; Pei-Dong Sun; Zhao-Fei Zhang; Hai-Long Ren
Journal:  J Orthop Surg Res       Date:  2020-08-31       Impact factor: 2.359