Literature DB >> 23974979

Catheter tip tracking for MR-guided interventions using discrete Kalman filter and mean shift localization.

Abubakr Eldirdiri1, Frédéric Courivaud, Rafael Palomar, Per Kristian Hol, Ole Jakob Elle.   

Abstract

PURPOSE: This paper presents and evaluates stochastic computer algorithms used to automatically detect and track marked catheter tip during MR-guided catheterization. The algorithms developed employ extraction and matching of regional features of the catheter tip to perform the localization.
METHOD: To perform the tracking, a probability map that indicates the possible locations of the catheter tip in the MR images is first generated. This map is generated from the similarity to a given marker template. The method to assess the similarity between the marker template image and the different positions on each MR frame is based on speeded-up robust features extracted from the gradient image. The probability map is then used in two different stochastic localization frameworks mean shift (MS) localization and Kalman filter (KF) to track the position of the catheter using pairs of orthogonal projection of 2D MR images. The algorithm developed was tested on catheter tip marked with LC resonant circuit (of size 2 mm x 2 cm) tuned to the Larmor frequency of the MRI scanner and for different image resolutions (1, 3, 5 and 7 mm squared pixel).
RESULTS: The tracking performance was very robust for the two algorithms MS and KF with image resolution as low as 3 mm where the localization error was about 1 mm for KF and 0.9 mm for MS. For the 5-mm resolution images, the error was 2.2 mm for both KF and MS, and for the 7-mm resolution images, the error was 3.6 and 3.7 mm for KF and MS, respectively.
CONCLUSION: Both KF and MS gave comparable results when it comes to accuracy for the different image resolutions. The results showed that the two tracking algorithms tracked the catheter tip with high robustness for image resolution of 3 mm and with acceptable reliability for image resolution as poor as 5 mm with the resonant marker configuration used.

Mesh:

Year:  2013        PMID: 23974979     DOI: 10.1007/s11548-013-0933-4

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  20 in total

1.  Heating around intravascular guidewires by resonating RF waves.

Authors:  M K Konings; L W Bartels; H F Smits; C J Bakker
Journal:  J Magn Reson Imaging       Date:  2000-07       Impact factor: 4.813

2.  Magnetic susceptibility effects and their application in the development of new ferromagnetic catheters for magnetic resonance imaging.

Authors:  D L Rubin; A V Ratner; S W Young
Journal:  Invest Radiol       Date:  1990-12       Impact factor: 6.016

3.  Dephased MRI.

Authors:  Chris J G Bakker; Jan-Henry Seppenwoolde; Koen L Vincken
Journal:  Magn Reson Med       Date:  2006-01       Impact factor: 4.668

4.  Interactive MR imaging and tracking of catheters with multiple tuned fiducial markers.

Authors:  Sanjeet Hegde; Marc E Miquel; Redha Boubertakh; David Gilderdale; Vivek Muthurangu; Stephen F Keevil; Ian Young; Derek L G Hill; Reza S Razavi
Journal:  J Vasc Interv Radiol       Date:  2006-07       Impact factor: 3.464

5.  A catheter tracking method using reverse polarization for MR-guided interventions.

Authors:  Haydar Celik; Asli Ulutürk; Turgut Tali; Ergin Atalar
Journal:  Magn Reson Med       Date:  2007-12       Impact factor: 4.668

6.  Morphing steady-state free precession.

Authors:  O Bieri; S Patil; H H Quick; K Scheffler
Journal:  Magn Reson Med       Date:  2007-12       Impact factor: 4.668

7.  Method for automatic localization of MR-visible markers using morphological image processing and conventional pulse sequences: feasibility for image-guided procedures.

Authors:  Harald Busse; Robert Trampel; Wilfried Gründer; Michael Moche; Thomas Kahn
Journal:  J Magn Reson Imaging       Date:  2007-10       Impact factor: 4.813

8.  Sub-pixel localisation of passive micro-coil fiducial markers in interventional MRI.

Authors:  Marc Rea; Donald McRobbie; Haytham Elhawary; Zion T H Tse; Michael Lamperth; Ian Young
Journal:  MAGMA       Date:  2008-09-19       Impact factor: 2.310

Review 9.  Real-time magnetic resonance imaging guidance for cardiovascular procedures.

Authors:  Keith A Horvath; Ming Li; Dumitru Mazilu; Michael A Guttman; Elliot R McVeigh
Journal:  Semin Thorac Cardiovasc Surg       Date:  2007

10.  Real-time MR imaging-guided passive catheter tracking with use of gadolinium-filled catheters.

Authors:  R A Omary; O Unal; D S Koscielski; R Frayne; F R Korosec; C A Mistretta; C M Strother; T M Grist
Journal:  J Vasc Interv Radiol       Date:  2000-09       Impact factor: 3.464

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.