Abstract
Early detection of occlusal caries in children is challenging for the dentists, because of the morphology of pit and fissures. The aim of this study was to compare in vitro the diagnostic performance of low-powered magnification with light-emitting diode headlight (LPMLED) using ICDAS-II criteria and AC Impedance Spectroscopy (ACIS) device, on occlusal surfaces of primary molars. The occlusal surfaces of 18 extracted primary molars were examined blindly by two examiners. The teeth were sectioned and examined under light microscopy using Downer's histological criteria as gold standard. Good to excellent inter- and intraexaminer reproducibility, higher sensitivity, specificity, and AUC values were achieved by LPMLED at D1 threshold. Also the relationship between histology and LPMLED was statistically significant. In conclusion visual aids have the potential to improve the performance of early caries detection and clinical diagnostics in children. Despite its potential, ACIS device should be considered as an adjunct method in detecting caries on primary teeth.Entities:
Year: 2013 PMID: 23956865 PMCID: PMC3728523 DOI: 10.1155/2013/276070
Source DB: PubMed Journal: ISRN Dent ISSN: 2090-4371
Criteria used in the histological examination [12].
| Score | Criteria used in the Downer histological examination [ |
|---|---|
| 0 | No enamel demineralisation or a narrow surface zone of opacity (edge phenomenon) |
| 1 | Enamel demineralisation limited to the outer 50% of the enamel layer |
| 2 | Demineralisation involving the inner 50% of the enamel, up to the enamel-dentine junction |
| 3 | Demineralisation involving the outer 50% of the dentine |
| 4 | Demineralisation involving the inner 50% of the dentine |
Intra- and interexaminer reproducibility (weighted kappa).
| AC Impedance | Low powered magnification + LED | |
|---|---|---|
| Examiner 1 | 0.6286 | 0.9057 |
| Examiner 2 | 0.6572 | 0.9515 SE |
| Examiner 1 versus Examiner 2 | 0.6473 | 0.8089 SE = 0.1061 |
Area under the ROC curve (standard error), sensitivity, and specificity at D1 diagnostic threshold.
| D1 diagnostic threshold | AC Impedance | Low-powered | ||
|---|---|---|---|---|
| Examiner 1 | Examiner 2 | Examiner 1 | Examiner 2 | |
| AUC (SE) | 0.60 (0.14) | 0.70 (0.13) |
|
|
| Sensitivity | 0.62 | 0.57 | 0.90 | 0.87 |
| Specificity | 0.68 | 0.70 | 0.70 | 0.75 |
AUC: area under the curve; SE: standard error.
*Statistically significant difference (P < 0.05).
Spearman's correlation coefficients using Downer classification systems.
| Examiner 1 | Examiner 2 | |
|---|---|---|
| AC Impedance Spectroscopy versus histology | 0.37 | 0.36 |
| Low powered magnification and LED versus histology |
|
|
*Correlation significant at the 0.05 level.