A comparison of five standard methods for evaluating image intensity uniformity in partially parallel imaging MRI.

Frank L Goerner1, Timothy Duong, R Jason Stafford, Geoffrey D Clarke.   

Abstract

PURPOSE: To investigate the utility of five different standard measurement methods for determining image uniformity for partially parallel imaging (PPI) acquisitions in terms of consistency across a variety of pulse sequences and reconstruction strategies.
METHODS: Images were produced with a phantom using a 12-channel head matrix coil in a 3T MRI system (TIM TRIO, Siemens Medical Solutions, Erlangen, Germany). Images produced using echo-planar, fast spin echo, gradient echo, and balanced steady state free precession pulse sequences were evaluated. Two different PPI reconstruction methods were investigated, generalized autocalibrating partially parallel acquisition algorithm (GRAPPA) and modified sensitivity-encoding (mSENSE) with acceleration factors (R) of 2, 3, and 4. Additionally images were acquired with conventional, two-dimensional Fourier imaging methods (R=1). Five measurement methods of uniformity, recommended by the American College of Radiology (ACR) and the National Electrical Manufacturers Association (NEMA) were considered. The methods investigated were (1) an ACR method and a (2) NEMA method for calculating the peak deviation nonuniformity, (3) a modification of a NEMA method used to produce a gray scale uniformity map, (4) determining the normalized absolute average deviation uniformity, and (5) a NEMA method that focused on 17 areas of the image to measure uniformity. Changes in uniformity as a function of reconstruction method at the same R-value were also investigated. Two-way analysis of variance (ANOVA) was used to determine whether R-value or reconstruction method had a greater influence on signal intensity uniformity measurements for partially parallel MRI.
RESULTS: Two of the methods studied had consistently negative slopes when signal intensity uniformity was plotted against R-value. The results obtained comparing mSENSE against GRAPPA found no consistent difference between GRAPPA and mSENSE with regard to signal intensity uniformity. The results of the two-way ANOVA analysis suggest that R-value and pulse sequence type produce the largest influences on uniformity and PPI reconstruction method had relatively little effect.
CONCLUSIONS: Two of the methods of measuring signal intensity uniformity, described by the (NEMA) MRI standards, consistently indicated a decrease in uniformity with an increase in R-value. Other methods investigated did not demonstrate consistent results for evaluating signal uniformity in MR images obtained by partially parallel methods. However, because the spatial distribution of noise affects uniformity, it is recommended that additional uniformity quality metrics be investigated for partially parallel MR images.

Mesh:

Year:  2013        PMID: 23927345      PMCID: PMC3745492          DOI: 10.1118/1.4816306

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  26 in total

1.  Improved diffusion-weighted single-shot echo-planar imaging (EPI) in stroke using sensitivity encoding (SENSE).

Authors:  R Bammer; S L Keeling; M Augustin; K P Pruessmann; R Wolf; R Stollberger; H P Hartung; F Fazekas
Journal:  Magn Reson Med       Date:  2001-09       Impact factor: 4.668

2.  A wavelet-based regularized reconstruction algorithm for SENSE parallel MRI with applications to neuroimaging.

Authors:  Lotfi Chaâri; Jean-Christophe Pesquet; Amel Benazza-Benyahia; Philippe Ciuciu
Journal:  Med Image Anal       Date:  2010-11-23       Impact factor: 8.545

3.  Intensity non-uniformity correction in MRI: existing methods and their validation.

Authors:  Boubakeur Belaroussi; Julien Milles; Sabin Carme; Yue Min Zhu; Hugues Benoit-Cattin
Journal:  Med Image Anal       Date:  2005-11-22       Impact factor: 8.545

4.  Parallel MR image reconstruction using augmented Lagrangian methods.

Authors:  Sathish Ramani; Jeffrey A Fessler
Journal:  IEEE Trans Med Imaging       Date:  2010-11-18       Impact factor: 10.048

5.  Slotted surface coil with reduced g-factor for SENSE imaging.

Authors:  K Ocegueda; A O Rodriguez
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2006

6.  Evaluation of multicoil breast arrays for parallel imaging.

Authors:  Helen Marshall; Patrick M Devine; Nishanthan Shanmugaratnam; Ryan Fobel; Peter Siegler; Cameron A Piron; Donald B Plewes
Journal:  J Magn Reson Imaging       Date:  2010-02       Impact factor: 4.813

7.  Standing-wave and RF penetration artifacts caused by elliptic geometry: an electrodynamic analysis of MRI.

Authors:  J G Sled; G B Pike
Journal:  IEEE Trans Med Imaging       Date:  1998-08       Impact factor: 10.048

8.  Aliasing artifacts in MR imaging.

Authors:  E Pusey; C Yoon; M L Anselmo; R B Lufkin
Journal:  Comput Med Imaging Graph       Date:  1988 Jul-Aug       Impact factor: 4.790

9.  Performance assessment and quality control in MRI by Eurospin test objects and protocols.

Authors:  R A Lerski; J D de Certaines
Journal:  Magn Reson Imaging       Date:  1993       Impact factor: 2.546

10.  Beyond the g-factor limit in sensitivity encoding using joint histogram entropy.

Authors:  David J Larkman; Philip G Batchelor; David Atkinson; Daniel Rueckert; Jo V Hajnal
Journal:  Magn Reson Med       Date:  2006-01       Impact factor: 4.668

View more
  6 in total

1.  Development of a hybrid magnetic resonance/computed tomography-compatible phantom for magnetic resonance guided radiotherapy.

Authors:  Min-Joo Kim; Seu-Ran Lee; Kyu-Ho Song; Hyeon-Man Baek; Bo-Young Choe; Tae Suk Suh
Journal:  J Radiat Res       Date:  2020-03-23       Impact factor: 2.724

2.  Selecting the reference image for registration of CEST series.

Authors:  Yi Zhang; Hye-Young Heo; Dong-Hoon Lee; Xuna Zhao; Shanshan Jiang; Kai Zhang; Haiyun Li; Jinyuan Zhou
Journal:  J Magn Reson Imaging       Date:  2015-08-13       Impact factor: 4.813

3.  Feature-Based Retinal Image Registration Using D-Saddle Feature.

Authors:  Roziana Ramli; Mohd Yamani Idna Idris; Khairunnisa Hasikin; Noor Khairiah A Karim; Ainuddin Wahid Abdul Wahab; Ismail Ahmedy; Fatimah Ahmedy; Nahrizul Adib Kadri; Hamzah Arof
Journal:  J Healthc Eng       Date:  2017-10-24       Impact factor: 2.682

4.  Virtual monoenergetic micro-CT imaging in mice with artificial intelligence.

Authors:  Brent van der Heyden; Stijn Roden; Rüveyda Dok; Sandra Nuyts; Edmond Sterpin
Journal:  Sci Rep       Date:  2022-02-11       Impact factor: 4.379

5.  Fabrication and evaluation of bilateral Helmholtz radiofrequency coil for thermo-stable breast image with reduced artifacts.

Authors:  Young Han Lee; Kyu-Ho Song; Jaemoon Yang; Won Jun Kang; Keum Sil Lee; Min Jung Kim; Eun-Kyung Kim; Dan Heo; Bo-Young Choe; Jin-Suck Suh
Journal:  J Appl Clin Med Phys       Date:  2021-12-01       Impact factor: 2.102

6.  Quantitative transmission ultrasound tomography: Imaging and performance characteristics.

Authors:  Bilal Malik; Robin Terry; James Wiskin; Mark Lenox
Journal:  Med Phys       Date:  2018-05-28       Impact factor: 4.071

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.